Why it was recalled

Potential breach in pouch packaging which could lead to loss of sterility.

FDA root cause category: Packaging process control

What the company is doing

Medline distributed an IMMEDIATE ACTION REQUIRED MEDICAL DEVICE RECALL notification, dated 2/13/25, to consignees who purchased individually packaged Proxima sterile surgical drapes and kits that contained Proxima sterile surgical drapes. Consignees that purchased kits containing recalled product are to immediately check stock for affected kits and place them in quarantine. Consignees are to complete the provided response form to receive over-labels to be applied to quarantined kits with additional instructions to remove the affected components prior to use of the kit. Consignees that purchased individual units of Proxima sterile surgical drapes are to immediately check stock for affected devices and immediately destroy identified product. Consignees are to complete the provided response form to obtain credit for the destroyed product. If a consignee is a distributor or have further distributed affected devices, they are instructed to forward the provided recall notification. Questions about this recall can be directed to Medline via phone at 866-359-1704 or by email to [email protected].

Recall details

Recall number
Z-1368-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Feb 13, 2025
Posted by the FDA
Mar 14, 2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
2,422 units
Product code
LRO
FDA submissions
K213481
Distribution
US Nationwide distribution.

Product description

Proxima Drape labeled as HEAD AND NECK OSIS, WILLIS SHOULDER PACK, EXTREMITY PACK-LF, TOTAL HIP PACK-LF, SHOULDER W/POUCH-LF, VEIN PACK. Sterile surgical drape.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Item No. DYNJ35520B, DYNJ38529F, DYNJ43406F, DYNJ43421D, DYNJ69185A, DYNJ69340B, DYNJ88923; UDI (case): 40198459106904, 40195327693870, 40198459074005, 40198459074784, 40195327582839, 40198459026127, 40198459216856; UDI (each): 10198459106903, 10195327693879, 10198459074004, 10198459074783, 10195327582838, 10198459026126, 10198459216855; Lot No. (Shelf Life) 24KMB843(05/31/2027), 24IMD400(01/31/2027), 24IMB784(01/31/2027), 24LBK460(02/28/2027), 24JBY288(12/31/2026), 24LBJ155(01/31/2027), 24JBU154(12/31/2026), 24HBP361(11/30/2026), 24HBA919(10/31/2026), 24JBX700(06/30/2026), 24JBC370(04/30/2026), 24IBI711(04/30/2026), 24HBO453(04/30/2026), 24HBA954(01/31/2026), 24JBK365(04/30/2026), 24HBR810(04/30/2026), 24GBX503(03/31/2026), 24LBO297(06/30/2026), 24JBV432(06/30/2026), 24IBU974(05/31/2026), 24HBR072(04/30/2026), 25ABB569(02/10/2026).

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

321 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.