Medline Convenience kits used for various procedures recallZ-1592-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Mar 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.
FDA root cause category: Under Investigation by firm
What the company is doing
An IMMEDIATE ACTION REQUIRED MEDLINE DEVICE RECALL notification letter dated 3/3/25 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Please destroy any affected inventory. Medline will issue credit for the product based upon the quantity and lot numbers identified on the submitted response form The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-25-050-FG Recall Code: RECALL CODE 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers document and destroy any affected product, and include your customer s quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1592-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 3, 2025
- Posted by the FDA
- Apr 17, 2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 3205 units
- Product code
- OEZ
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Panama, Virgin Islands.
Product description
Medline Convenience kits used for various procedures: 1) DR. KHAN CABG ADDS PACK, Model Number: DYNJ59310D; 2) CARDIAC STERNAL PACK, Model Number: DYNJ32436A; 3) VEIN CLOSURE PACK, Model Number: DYNJ35586G; 4) ABLATION PACK-LF, Model Number: DYNJ55343B; 5) ABLATION PACK-LF, Model Number: DYNJ55343B; 6) CARDIAC PACK, Model Number: DYNJ65027B; 7) VASCULAR PACK-LF, Model Number: DYNJ48196C; 8) IMPLANT PACK, Model Number: DYNJ40480; 9) OPEN HEART, Model Number: DYNJ900523B; 10) PACEMAKER PACK-LF, Model Number: DYNJ19758I; 11) CHOG VP SHUNT PACK, Model Number: SYNJ10280C; 12) VASCULAR PACK-LF, Model Number: DYNJ48196C; 13) EP PACK-LF, Model Number: DYNJ27949F; 14) PACEMAKER PACK WRO-LF, Model Number: DYNJ41874C; 15) CARDIAC PACK OPTION 1, Model Number: DYNJ83208A; 16) EP DRAPE PACK, Model Number: DYNJ56729; 17) CSTM EP PACEMAKER DRAPE PACK, Model Number: DYNJCD0326; 18) MINOR INTERVENTIONAL TRAY, Model Number: DYNJ80084; 19) CT PACK, Model Number: DYNJ39111A; 20) VASCULAR PACK-LF, Model Number: DYNJ25102I; 21) VASCULAR ACCESS PROCEDURE PACK, Model Number: DYNJ62339; 22) ANGIO TRAY, Model Number: DYNJ43278C; 23) PACEMAKER PACK-LF, Model Number: DYNJ19758I; 24) E.P. BASIC PACK-LF, Model Number: DYNJ24735F; 25) PORT A CATH PACEMAKER BJC-LF, Model Number: DYNJ41683B; 26) PACEMAKER PACK, Model Number: DYNJ39379C; 27) VEIN PACK, Model Number: DYNJ64016B; 28) EP PACK, Model Number: DYNJ68019A; 29) ENDOVENOUS BASIC PACK, Model Number: DYNJ69108A; 30) PACEMAKER PACK-LF, Model Number: DYNJHTCLP101A; 31) EP PACK, Model Number: DYNJ38147C; 32) HEART PRCMNT CHOICE, Model Number: DYNDA1841; 33) PACEMAKER PACK SMGH-LF, Model Number: DYNJ41925I; 34) EP PACK, Model Number: DYNJ81951B; 35) HEALTH ALL VASCULAR PACK-LF, Model Number: DYNJ19538F; 36) OPEN HEART, Model Number: CDS982026V; 37) OPEN HEART, Model Number: CDS982026V
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
1) DYNJ59310D, UDI-DI: 10195327402396(each), 40195327402397(case), Lot Number: 24IBB643; 2) DYNJ32436A, UDI-DI: 10884389686614(each), 40884389686615(case), Lot Number: 24IBG295; 3) DYNJ35586G, UDI-DI: 10193489663235(each), 40193489663236(case), Lot Number: 24IBG343; 4) DYNJ55343B, UDI-DI: 10195327478056(each), 40195327478057(case), Lot Number: 24IBG535; 5) DYNJ55343B, UDI-DI: 10195327478056(each), 40195327478057(case), Lot Number: 24IBH248; 6) DYNJ65027B, UDI-DI: 10193489471991(each), 40193489471992(case), Lot Number: 24IBH653; 7) DYNJ48196C, UDI-DI: 10193489712469(each), 40193489712460(case), Lot Number: 24IBL194; 8) DYNJ40480, UDI-DI: 10884389861196(each), 40884389861197(case), Lot Number: 24IBO308; 9) DYNJ900523B, UDI-DI: 10195327628703(each), 40195327628704(case), Lot Number: 24IBT180; 10) DYNJ19758I, UDI-DI: 10193489992328(each), 40193489992329(case), Lot Number: 24IBV182; 11) SYNJ10280C, UDI-DI: 10195327373597(each), 40195327373598(case), Lot Number: 24JBL521; 12) DYNJ48196C, UDI-DI: 10193489712469(each), 40193489712460(case), Lot Number: 24JBM305; 13) DYNJ27949F, UDI-DI: 10195327016104(each), 40195327016105(case), Lot Number: 24JBM849; 14) DYNJ41874C, UDI-DI: 10195327428914(each), 40195327428915(case), Lot Number: 24JBM893; 15) DYNJ83208A, UDI-DI: 10198459097065(each), 40198459097066(case), Lot Number: 24JBM959; 16) DYNJ56729, UDI-DI: 10889942676026(each), 40889942676027(case), Lot Number: 24JBT045; 17) DYNJCD0326, UDI-DI: 10193489203509(each), 40193489203500(case), Lot Number: 24JBU078; 18) DYNJ80084, UDI-DI: 10195327012014(each), 40195327012015(case), Lot Number: 24JBU113; 19) DYNJ39111A, UDI-DI: 10195327429744(each), 40195327429745(case), Lot Number: 24JBU159; 20) DYNJ25102I, UDI-DI: 10195327573812(each), 40195327573813(case), Lot Number: 24JBU436; 21) DYNJ62339, UDI-DI: 10193489646924(each), 40193489646925(case), Lot Number: 24JBV115; 22) DYNJ43278C, UDI-DI: 10193489957358(each), 40193489957359(case), Lot Number: 24JBV655; 23) DYNJ19758I, UDI-DI: 10193489992328(each), 40193489992329(case), Lot Number: 24JBW212; 24) DYNJ24735F, UDI-DI: 10198459014277(each), 40198459014278(case), Lot Number: 24JBW301; 25) DYNJ41683B, UDI-DI: 10198459124112(each), 40198459124113(case), Lot Number: 24JBW466; 26) DYNJ39379C, UDI-DI: 10193489611526(each), 40193489611527(case), Lot Number: 24JBX574; 27) DYNJ64016B, UDI-DI: 10193489258363(each), 40193489258364(case), Lot Number: 24KBA541; 28) DYNJ68019A, UDI-DI: 10198459163937(each), 40198459163938(case), Lot Number: 24KBC477; 29) DYNJ69108A, UDI-DI: 10195327025717(each), 40195327025718(case), Lot Number: 24KBC478; 30) DYNJHTCLP101A, UDI-DI: 10080196980154(each), 40080196980155(case), Lot Number: 24KBC947; 31) DYNJ38147C, UDI-DI: 10193489854862(each), 40193489854863(case), Lot Number: 24KBF070; 32) DYNDA1841, UDI-DI: 10889942522286(each), 40889942522287(case), Lot Number: 24KBF305; 33) DYNJ41925I, UDI-DI: 10198459160516(each), 40198459160517(case), Lot Number: 24KBG148; 34) DYNJ81951B, UDI-DI: 10198459003615(each), 40198459003616(case), Lot Number: 24KBG626; 35) DYNJ19538F, UDI-DI: 10195327658137(each), 40195327658138(case), Lot Number: 24KBG632; 36) CDS982026V, UDI-DI: 10195327657635(each), 40195327657636(case), Lot Number: 24KBP804; 37) CDS982026V, UDI-DI: 10195327657635(each), 40195327657636(case), Lot Number: 24LBI154
Devices named in this recall
33 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ65027BMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ48196CMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ43278CMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ69108AMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ55343BMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJ19538FMedline Industries, Inc. · Cardiovascular procedure kit2 recalls0 reports
- Medline Industries, Inc. DYNJHTCLP101AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJCD0326Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ64016BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ39379CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ62339Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ35586GMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ38147CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ19758IMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ80084Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ27949FMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. SYNJ10280CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ59310DMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ41874CMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ39111AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ25102IMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ900523BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. CDS982026VMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ81951BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ24735FMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ83208AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ41683BMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ41925IMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ68019AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ32436AMedline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ40480Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNDA1841Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
- Medline Industries, Inc. DYNJ56729Medline Industries, Inc. · Cardiovascular procedure kit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96510).
- Z-1586-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) INTUBATION TRAY, Model Number: MNS10395;
- Z-1587-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) RHINOPLASTY PACK, Model Number: DYNJ84116A
- Z-1588-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) PAIN PACK, Model Number: DYNJ67577A; 2) PAIN TRIAL PACK, Model Number: DYNJ67576A;
- Z-1589-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) CUSTOM BLOCK KEMP PACK-LF, Model Number: DYNJ0371999G;
- Z-1590-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) ANGIO CONSOLIDATED PACK, Model Number: DYNJHTCAD1; 2) ANGIOGRAPHY TRAY-LF, Model Number: DYNJ05350V; 3) ANGIOGRAPHY PACK,…
- Z-1591-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) ANGIOGRAPHY DRAPE PACK, Model Number: HCP075G; 2) ENSEMBLE CORONAROGRAPHIE-LF, Model Number: DYNJ64354A; 3) RADIOL ANGIO …
- Z-1593-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) LINE INSERTION PACK, Model Number: DYNJ44949C; 2) PICC LINE INSERTION BJC-LF, Model Number: DYNJ41682C; 3) SUPERFICIAL VE…
- Z-1594-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) LARGE BIOPSY PACK NTX, Model Number: DYNJ68393A; 2) SOFT TISSUE BIOPSY PACK, Model Number: DYNJ81441B; 3) BILAT TESTICULA…
- Z-1595-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) DENTAL PACK-LF, Model Number: DYNJ84880; 2) DENTAL PACK, Model Number: DYNJ56229A; 3) DENTAL PACK-LF, Model Number: PHS80…
- Z-1596-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) RHINOPLASTY PACK, Model Number: DYNJ27333F
- Z-1597-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) T A PACK, Model Number: DYNJ67791A; 2) NASAL PACK, Model Number: DYNJ66514; 3) ENT PACK, Model Number: DYNJ67775A; 4) …
- Z-1598-2025 · Open, ClassifiedRefer to RES
- Z-1599-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) T&A ENDO PACK, Model Number: DYNJ59620A; 2) TRAY ADD A CATH FOL UM MF 10ML, Model Number: DYND150200S
- Z-1600-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) OCULOPLASTIC PICK PACK, Model Number: DYNJ43914B; 2) PLASTIC PACK, Model Number: DYNJ65109; 3) PLASTIC PROCEDURE PACK, Mo…
- Z-1601-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) MINOR BASIN -SMH, Model Number: DYNJ907733B
- Z-1602-2025 · Open, ClassifiedRefer to RES
- Z-1603-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) PRECIP TRAY, Model Number: MNS12590
- Z-1604-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) IV PACK SOUTH, Model Number: DYNJ88422
- Z-1605-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) NAIL PACK, Model Number: SUT15830A;
- Z-1606-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) SKIN LESION SET, Model Number: MNS6270
- Z-1608-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) UNIVERSAL DRAPE PACK, Model Number: DYNJCD0239
- Z-1610-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) NEURO PACK WRO-LF, Model Number: DYNJ41856C; 2) STIM IMPLANT PACK, Model Number: DYNJ53032C; 3) WMC VP SHUNT PACK-LF, Mod…
- Z-1611-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) LABOR AND DELIVERY PACK-LF, Model Number: DYNJ26971L; 2) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ0378319P; 3) VUWC OB…
- Z-1612-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) CW PERI GYN, Model Number: DYNJ80807B
- Z-1613-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) STANDARD D&C, Model Number: DYNJ69733; 2) NHS - LAP BOWEL GNG, Model Number: DYNJ39904A; 3) LAPAROSCOPY GYN SURGERY PA, M…
- Z-1614-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) ED CRASH C-SECTION TRAY, Model Number: MNS13635A; 2) C-SECTION PCH-LF, Model Number: DYNJ81281; 3) C-SECTION PACK-LF, Mod…
- Z-1615-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) PERI GYN PACK, Model Number: DYNJ80786; 2) D & C HYSTEROSCOPY PACK, Model Number: DYNJ21937J; 3) VAGINAL HYSTERECTOMY WRH…
- Z-1616-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) UTHET TYLER OB PACK, Model Number: DYNJ68673; 2) D AND C PACK, Model Number: DYNJ66290A; 3) D & C PACK-LF, Model Number: …
- Z-1617-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) BASIC CATARACT, Model Number: DYNJ41097A; 2) PK EYE, Model Number: DYNJ60631B; 3) CATARACT PACK, Model Number: DYNJ80316A…
- Z-1618-2025 · Open, ClassifiedRefer to RES
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page