Refer to RES recallZ-1618-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Mar 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.
FDA root cause category: Under Investigation by firm
What the company is doing
An IMMEDIATE ACTION REQUIRED MEDLINE DEVICE RECALL notification letter dated 3/3/25 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Please destroy any affected inventory. Medline will issue credit for the product based upon the quantity and lot numbers identified on the submitted response form The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-25-050-FG Recall Code: RECALL CODE 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers document and destroy any affected product, and include your customer s quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1618-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 3, 2025
- Posted by the FDA
- Apr 17, 2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 11889 units
- Product code
- OJH
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Panama, Virgin Islands.
Product description
Refer to RES
Code information (models, lots, serial numbers)
Show the full code information (10k characters)
1) DYNJ61594, UDI-DI: 10193489586121(each), 40193489586122(case), Lot Number: 24IBF062; 2) DYNJ31798D, UDI-DI: 10195327609078(each), 40195327609079(case), Lot Number: 24IBF219; 3) DYNJ46311F, UDI-DI: 10193489933185(each), 40193489933186(case), Lot Number: 24IBF751; 4) DYNJ49386A, UDI-DI: 10889942292882(each), 40889942292883(case), Lot Number: 24IBF768; 5) DYNJ51781G, UDI-DI: 10198459045974(each), 40198459045975(case), Lot Number: 24IBG614; 6) DYNJ54474C, UDI-DI: 10195327698560(each), 40195327698561(case), Lot Number: 24IBG988; 7) DYNJ46311F, UDI-DI: 10193489933185(each), 40193489933186(case), Lot Number: 24IBH023; 8) DYNJ65348, UDI-DI: 10193489268881(each), 40193489268882(case), Lot Number: 24IBH081; 9) DYNJ37322I, UDI-DI: 10193489961553(each), 40193489961554(case), Lot Number: 24IBH579; 10) DYNJ69926A, UDI-DI: 10195327153601(each), 40195327153602(case), Lot Number: 24IBH589; 11) DYNJ39903, UDI-DI: 10884389842072(each), 40884389842073(case), Lot Number: 24IBH647; 12) DYNJ47816B, UDI-DI: 10193489370256(each), 40193489370257(case), Lot Number: 24IBH806; 13) SYNJ10253B, UDI-DI: 10193489373592(each), 40193489373593(case), Lot Number: 24IBH821; 14) DYNJ31169C, UDI-DI: 10889942373468(each), 40889942373469(case), Lot Number: 24IBK285; 15) DYNJ06622J, UDI-DI: 10889942725649(each), 40889942725640(case), Lot Number: 24IBQ232; 16) DYNJ83390, UDI-DI: 10195327262914(each), 40195327262915(case), Lot Number: 24IBQ640; 17) DYNJ33236, UDI-DI: 10884389260227(each), 40884389260228(case), Lot Number: 24IBR149; 18) DYNDH1510, UDI-DI: 10193489429749(each), 40193489429740(case), Lot Number: 24IBR168; 19) DYNJ52928D, UDI-DI: 10198459022111(each), 40198459022112(case), Lot Number: 24IBR186; 20) DYNJ35021D, UDI-DI: 10195327242343(each), 40195327242344(case), Lot Number: 24IBR323; 21) SYNJ10253B, UDI-DI: 10193489373592(each), 40193489373593(case), Lot Number: 24IBR379; 22) DYNJ69118B, UDI-DI: 10195327695989(each), 40195327695980(case), Lot Number: 24IBR884; 23) DYNJ48542D, UDI-DI: 10889942579327(each), 40889942579328(case), Lot Number: 24IBS260; 24) DYNJ54596B, UDI-DI: 10889942811786(each), 40889942811787(case), Lot Number: 24IBS271; 25) DYNJ82944, UDI-DI: 10195327218850(each), 40195327218851(case), Lot Number: 24IBS296; 26) DYNJ61549C, UDI-DI: 10193489892383(each), 40193489892384(case), Lot Number: 24JBA841; 27) DYNDH1797A, UDI-DI: 10195327264871(each), 40195327264872(case), Lot Number: 24JBF238; 28) DYNJTS07025, UDI-DI: 10889942602230(each), 40889942602231(case), Lot Number: 24JBL653; 29) DYNJ43539, UDI-DI: 10888277061699(each), 40888277061690(case), Lot Number: 24JBM313; 30) DYNJ30650F, UDI-DI: 10193489854701(each), 40193489854702(case), Lot Number: 24JBM728; 31) DYNJ32844, UDI-DI: 10884389233061(each), 40884389233062(case), Lot Number: 24JBM854; 32) DYNJ38083B, UDI-DI: 10889942670529(each), 40889942670520(case), Lot Number: 24JBM857; 33) DYNJ40036A, UDI-DI: 10889942696840(each), 40889942696841(case), Lot Number: 24JBM867; 34) DYNJ40052B, UDI-DI: 10193489982770(each), 40193489982771(case), Lot Number: 24JBM868; 35) DYNJ41851D, UDI-DI: 10195327429041(each), 40195327429042(case), Lot Number: 24JBM889; 36) DYNJ53406A, UDI-DI: 10889942573844(each), 40889942573845(case), Lot Number: 24JBM917; 37) DYNJ38075B, UDI-DI: 10195327001209(each), 40195327001200(case), Lot Number: 24JBN075; 38) DYNJ49386A, UDI-DI: 10889942292882(each), 40889942292883(case), Lot Number: 24JBN268; 39) DYNJS0815, UDI-DI: 10080196597543(each), 40080196597544(case), Lot Number: 24JBN668; 40) DYNJ63114C, UDI-DI: 10193489870206(each), 40193489870207(case), Lot Number: 24JBN803; 41) DYNJ40456F, UDI-DI: 10195327174798(each), 40195327174799(case), Lot Number: 24JBS603; 42) DYNJ61990A, UDI-DI: 10193489942385(each), 40193489942386(case), Lot Number: 24JBT522; 43) DYNJ52746D, UDI-DI: 10195327469726(each), 40195327469727(case), Lot Number: 24JBT734; 44) DYNJ68243, UDI-DI: 10193489498622(each), 40193489498623(case), Lot Number: 24JBT999; 45) SYNJ10324A, UDI-DI: 10195327017835(each), 40195327017836(case), Lot Number: 24JBU045; 46) DYNJ64867A, UDI-DI: 10195327329204(each), 40195327329205(case), Lot Number: 24JBU398; 47) DYNJ65348, UDI-DI: 10193489268881(each), 40193489268882(case), Lot Number: 24JBU434; 48) DYNJ51781G, UDI-DI: 10198459045974(each), 40198459045975(case), Lot Number: 24JBV555; 49) DYNJ63773A, UDI-DI: 10193489322699(each), 40193489322690(case), Lot Number: 24JBV852; 50) DYNJ69821, UDI-DI: 10193489981452(each), 40193489981453(case), Lot Number: 24JBV870; 51) DYNJ50403C, UDI-DI: 10193489489767(each), 40193489489768(case), Lot Number: 24JBW140; 52) DYNJ81916, UDI-DI: 10195327155339(each), 40195327155330(case), Lot Number: 24JBW146; 53) DYNJHS0614J, UDI-DI: 10193489436211(each), 40193489436212(case), Lot Number: 24JBW161; 54) DYNJ50467C, UDI-DI: 10193489733327(each), 40193489733328(case), Lot Number: 24JBW190; 55) DYNJ47874F, UDI-DI: 10195327579623(each), 40195327579624(case), Lot Number: 24JBW236; 56) DYNDH1905A, UDI-DI: 10198459157615(each), 40198459157616(case), Lot Number: 24JBX271; 57) DYNJ48574, UDI-DI: 10888277913905(each), 40888277913906(case), Lot Number: 24JBX316; 58) DYNJ37641B, UDI-DI: 10889942572977(each), 40889942572978(case), Lot Number: 24JBX693; 59) DYNJCD0138, UDI-DI: 10888277089884(each), 40888277089885(case), Lot Number: 24JBX854; 60) OP8082B, UDI-DI: 10888277589254(each), 40888277589255(case), Lot Number: 24JBX873; 61) DYNJ23138M, UDI-DI: 10193489707007(each), 40193489707008(case), Lot Number: 24JBY245; 62) DYNJS0801, UDI-DI: 10080196799107(each), 40080196799108(case), Lot Number: 24KBA048; 63) DYNJS0803, UDI-DI: 10080196986897(each), 40080196986898(case), Lot Number: 24KBA085; 64) DYNJ35191D, UDI-DI: 10888277786554(each), 40888277786555(case), Lot Number: 24KBA591; 65) DYNDH1836, UDI-DI: 10195327325251(each), 40195327325252(case), Lot Number: 24KBA901; 66) DYNJ48888A, UDI-DI: 10193489772333(each), 40193489772334(case), Lot Number: 24KBB200; 67) DYNJ46724B, UDI-DI: 10889942747412(each), 40889942747413(case), Lot Number: 24KBB420; 68) DYNJ64526B, UDI-DI: 10195327518417(each), 40195327518418(case), Lot Number: 24KBB435; 69) DYNJ67835A, UDI-DI: 10195327486686(each), 40195327486687(case), Lot Number: 24KBB774; 70) DYNJ82944, UDI-DI: 10195327218850(each), 40195327218851(case), Lot Number: 24KBB853; 71) DYNDH1510, UDI-DI: 10193489429749(each), 40193489429740(case), Lot Number: 24KBB886; 72) DYNJ47541F, UDI-DI: 10195327456467(each), 40195327456468(case), Lot Number: 24KBC169; 73) DYNJ40025B, UDI-DI: 10889942896455(each), 40889942896456(case), Lot Number: 24KBC506; 74) DYNJ41797D, UDI-DI: 10195327325817(each), 40195327325818(case), Lot Number: 24KBC508; 75) DYNJ40693F, UDI-DI: 10193489958256(each), 40193489958257(case), Lot Number: 24KBC602; 76) DYNJ57800C, UDI-DI: 10195327493882(each), 40195327493883(case), Lot Number: 24KBC676; 77) DYNJ59158D, UDI-DI: 10195327043872(each), 40195327043873(case), Lot Number: 24KBC686; 78) DYNJ61199A, UDI-DI: 10195327604257(each), 40195327604258(case), Lot Number: 24KBC700; 79) DYNJ61202A, UDI-DI: 10195327604295(each), 40195327604296(case), Lot Number: 24KBC702; 80) DYNJ63838I, UDI-DI: 10195327497606(each), 40195327497607(case), Lot Number: 24KBC726; 81) DYNJ81780D, UDI-DI: 10195327675660(each), 40195327675661(case), Lot Number: 24KBC801; 82) DYNJ83389A, UDI-DI: 10195327366124(each), 40195327366125(case), Lot Number: 24KBD302; 83) DYNJ51930A, UDI-DI: 10889942319350(each), 40889942319351(case), Lot Number: 24KBD436; 84) DYNDH1838, UDI-DI: 10195327323530(each), 40195327323531(case), Lot Number: 24KBD478; 85) CDS983609N, UDI-DI: 10195327209759(each), 40195327209750(case), Lot Number: 24KBD672; 86) DYNJ17529N, UDI-DI: 10195327257743(each), 40195327257744(case), Lot Number: 24KBD810; 87) DYNJ49824I, UDI-DI: 10198459164965(each), 40198459164966(case), Lot Number: 24KBD869; 88) DYNJ39310C, UDI-DI: 10193489982930(each), 40193489982931(case), Lot Number: 24KBE302; 89) DYNJ82946, UDI-DI: 10195327218874(each), 40195327218875(case), Lot Number: 24KBE442; 90) DYNJS0821, UDI-DI: 10080196986651(each), 40080196986652(case), Lot Number: 24KBE489; 91) DYNJS0843, UDI-DI: 10080196597581(each), 40080196597582(case), Lot Number: 24KBE498; 92) DYNJ54596C, UDI-DI: 10198459169519(each), 40198459169510(case), Lot Number: 24KBE672; 93) DYNJ69826, UDI-DI: 10193489982107(each), 40193489982108(case), Lot Number: 24KBE700; 94) DYNJ59157D, UDI-DI: 10198459165580(each), 40198459165581(case), Lot Number: 24KBF080; 95) DYNDH1797A, UDI-DI: 10195327264871(each), 40195327264872(case), Lot Number: 24KBF205; 96) DYNJ40036A, UDI-DI: 10889942696840(each), 40889942696841(case), Lot Number: 24KBF329; 97) DYNJ53406A, UDI-DI: 10889942573844(each), 40889942573845(case), Lot Number: 24KBF357; 98) DYNJ64918A, UDI-DI: 10193489822328(each), 40193489822329(case), Lot Number: 24KBF386; 99) M101413A, UDI-DI: 10193489265019(each), 40193489265010(case), Lot Number: 24KBF904; 100) DYNJ85057A, UDI-DI: 10198459186295(each), 40198459186296(case), Lot Number: 24KBF953; 101) DYNJ40043A, UDI-DI: 10889942739226(each), 40889942739227(case), Lot Number: 24KBG144; 102) DYNJ81328A, UDI-DI: 10195327270506(each), 40195327270507(case), Lot Number: 24KBG157; 103) DYNJ68092, UDI-DI: 10193489484595(each), 40193489484596(case), Lot Number: 24KBG331; 104) DYNDH1209B, UDI-DI: 10889942924080(each), 40889942924081(case), Lot Number: 24KBG698; 105) DYNJ30523B, UDI-DI: 10193489773842(each), 40193489773843(case), Lot Number: 24KBH230; 106) DYNJ69968C, UDI-DI: 10198459113130(each), 40198459113131(case), Lot Number: 24KBH378; 107) DYNJ902287F, UDI-DI: 10195327130367(each), 40195327130368(case), Lot Number: 24LBO456
Devices named in this recall
97 devices in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ69926AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJ67835AMedline Industries, Inc. · Orthopedic tray3 recalls0 reports
- Medline Industries, Inc. DYNJS0801Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJS0803Medline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. SYNJ10253BMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ50403CMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ50467CMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ63114CMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ61990AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ37322IMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ39310CMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ59158DMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. CDS983609NMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ51930AMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJ54596BMedline Industries, Inc. · Orthopedic tray2 recalls0 reports
- Medline Industries, Inc. DYNJS0821Medline Industries, Inc. · Orthopedic tray1 recall1 report
- Medline Industries, Inc. DYNJ54474CMedline Industries, Inc. · Orthopedic tray1 recall1 report
- Medline Industries, Inc. DYNJ33236Medline Industries, Inc. · Orthopedic tray1 recall1 report
- Medline Industries, Inc. DYNJ40025BMedline Industries, Inc. · Orthopedic tray1 recall1 report
- Medline Industries, Inc. DYNJS0815Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJS0843Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ65348Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ63773AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ47816BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNDH1510Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJHS0614JMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ68092Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ68243Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ61594Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ23138MMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ48888AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ30523BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ64918AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ30650FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ61549CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ46311FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ40693FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ69821Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ69826Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ40052BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ44385GMedline Industries, Inc. · Cardiac catheterization kit1 recall0 reports
- Medline Industries, Inc. DYNJ38075BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. SYNJ10324AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ902287FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ81916Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ40456FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ82944Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ82946Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ35021DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ17529NMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ83390Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNDH1797AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ81328AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNDH1838Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNDH1836Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ41797DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ64867AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ83389AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ41851DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ47541FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ52746DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ57800CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ63838IMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ64526BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ47874FMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ61199AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ61202AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ31798DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ81780DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ69118BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ52928DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ51781GMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ69968CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNDH1905AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ49824IMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ59157DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ54596CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ85057AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ32844Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ39903Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ43539Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJCD0138Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. OP8082BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ35191DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ48574Medline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ49386AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ31169CMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ37641BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ53406AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ48542DMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- MEDLINE DYNJTS07025Medline Industries, Inc. · Lumbar puncture tray (adult & pediatric)1 recall0 reports
- Medline Industries, Inc. DYNJ38083BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ40036AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ06622JMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ40043AMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNJ46724BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
- Medline Industries, Inc. DYNDH1209BMedline Industries, Inc. · Orthopedic tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96510).
- Z-1586-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) INTUBATION TRAY, Model Number: MNS10395;
- Z-1587-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) RHINOPLASTY PACK, Model Number: DYNJ84116A
- Z-1588-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) PAIN PACK, Model Number: DYNJ67577A; 2) PAIN TRIAL PACK, Model Number: DYNJ67576A;
- Z-1589-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) CUSTOM BLOCK KEMP PACK-LF, Model Number: DYNJ0371999G;
- Z-1590-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) ANGIO CONSOLIDATED PACK, Model Number: DYNJHTCAD1; 2) ANGIOGRAPHY TRAY-LF, Model Number: DYNJ05350V; 3) ANGIOGRAPHY PACK,…
- Z-1591-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) ANGIOGRAPHY DRAPE PACK, Model Number: HCP075G; 2) ENSEMBLE CORONAROGRAPHIE-LF, Model Number: DYNJ64354A; 3) RADIOL ANGIO …
- Z-1592-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) DR. KHAN CABG ADDS PACK, Model Number: DYNJ59310D; 2) CARDIAC STERNAL PACK, Model Number: DYNJ32436A; 3) VEIN CLOSURE PAC…
- Z-1593-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) LINE INSERTION PACK, Model Number: DYNJ44949C; 2) PICC LINE INSERTION BJC-LF, Model Number: DYNJ41682C; 3) SUPERFICIAL VE…
- Z-1594-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) LARGE BIOPSY PACK NTX, Model Number: DYNJ68393A; 2) SOFT TISSUE BIOPSY PACK, Model Number: DYNJ81441B; 3) BILAT TESTICULA…
- Z-1595-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) DENTAL PACK-LF, Model Number: DYNJ84880; 2) DENTAL PACK, Model Number: DYNJ56229A; 3) DENTAL PACK-LF, Model Number: PHS80…
- Z-1596-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) RHINOPLASTY PACK, Model Number: DYNJ27333F
- Z-1597-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) T A PACK, Model Number: DYNJ67791A; 2) NASAL PACK, Model Number: DYNJ66514; 3) ENT PACK, Model Number: DYNJ67775A; 4) …
- Z-1598-2025 · Open, ClassifiedRefer to RES
- Z-1599-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) T&A ENDO PACK, Model Number: DYNJ59620A; 2) TRAY ADD A CATH FOL UM MF 10ML, Model Number: DYND150200S
- Z-1600-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) OCULOPLASTIC PICK PACK, Model Number: DYNJ43914B; 2) PLASTIC PACK, Model Number: DYNJ65109; 3) PLASTIC PROCEDURE PACK, Mo…
- Z-1601-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) MINOR BASIN -SMH, Model Number: DYNJ907733B
- Z-1602-2025 · Open, ClassifiedRefer to RES
- Z-1603-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) PRECIP TRAY, Model Number: MNS12590
- Z-1604-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) IV PACK SOUTH, Model Number: DYNJ88422
- Z-1605-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) NAIL PACK, Model Number: SUT15830A;
- Z-1606-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) SKIN LESION SET, Model Number: MNS6270
- Z-1608-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) UNIVERSAL DRAPE PACK, Model Number: DYNJCD0239
- Z-1610-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) NEURO PACK WRO-LF, Model Number: DYNJ41856C; 2) STIM IMPLANT PACK, Model Number: DYNJ53032C; 3) WMC VP SHUNT PACK-LF, Mod…
- Z-1611-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) LABOR AND DELIVERY PACK-LF, Model Number: DYNJ26971L; 2) VAGINAL DELIVERY PACK-LF, Model Number: DYNJ0378319P; 3) VUWC OB…
- Z-1612-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) CW PERI GYN, Model Number: DYNJ80807B
- Z-1613-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) STANDARD D&C, Model Number: DYNJ69733; 2) NHS - LAP BOWEL GNG, Model Number: DYNJ39904A; 3) LAPAROSCOPY GYN SURGERY PA, M…
- Z-1614-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) ED CRASH C-SECTION TRAY, Model Number: MNS13635A; 2) C-SECTION PCH-LF, Model Number: DYNJ81281; 3) C-SECTION PACK-LF, Mod…
- Z-1615-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) PERI GYN PACK, Model Number: DYNJ80786; 2) D & C HYSTEROSCOPY PACK, Model Number: DYNJ21937J; 3) VAGINAL HYSTERECTOMY WRH…
- Z-1616-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) UTHET TYLER OB PACK, Model Number: DYNJ68673; 2) D AND C PACK, Model Number: DYNJ66290A; 3) D & C PACK-LF, Model Number: …
- Z-1617-2025 · Open, ClassifiedMedline Convenience kits used for various procedures: 1) BASIC CATARACT, Model Number: DYNJ41097A; 2) PK EYE, Model Number: DYNJ60631B; 3) CATARACT PACK, Model Number: DYNJ80316A…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page