Medline Convenience kits used for various procedures recallZ-1597-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Mar 3, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for open/weak seals on purchased breather pouches used for packaging of select sterile convenience kits.

FDA root cause category: Under Investigation by firm

What the company is doing

An IMMEDIATE ACTION REQUIRED MEDLINE DEVICE RECALL notification letter dated 3/3/25 was sent to customers. REQUIRED ACTION: 1.Immediately check your stock for the affected item number and the affected lot numbers which can be located within the recall portal. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the response form. 3.Please destroy any affected inventory. Medline will issue credit for the product based upon the quantity and lot numbers identified on the submitted response form The login for completing the response form is: Website link: https://recalls.medline.com Recall Reference #: R-25-050-FG Recall Code: RECALL CODE 4.If you are a distributor, or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please request that customers document and destroy any affected product, and include your customer s quantities on your response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-1597-2025
Status
Open, Classified
Initiated by the company
Mar 3, 2025
Posted by the FDA
Apr 17, 2025
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
9216 units
Product code
OGR
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Panama, Virgin Islands.

Product description

Medline Convenience kits used for various procedures: 1) T A PACK, Model Number: DYNJ67791A; 2) NASAL PACK, Model Number: DYNJ66514; 3) ENT PACK, Model Number: DYNJ67775A; 4) T/A ENDO PACK, Model Number: DYNJ83158; 5) T & A PACK, Model Number: DYNJ0634878I; 6) PPM-ICD PACK-LF, Model Number: DYNJ15949D; 7) TONSIL PACK, Model Number: DYNJ44525C; 8) ENT TRAY, Model Number: DYNJ44391K; 9) ENT PACK, Model Number: DYNJ49300D; 10) NASAL PACK, Model Number: DYNJ61555B; 11) HEAD AND NECK, Model Number: DYNJ57287A; 12) SINUS ENDOSCOPY-LF, Model Number: DYNJ36720F; 13) ENT PACK, Model Number: DYNJ54035B; 14) ENT PACK, Model Number: DYNJ36516; 15) FESS PACK, Model Number: DYNJ59502B; 16) T AND A PACK, Model Number: DYNJ80790; 17) PK,ENT, Model Number: DYNJ57583; 18) ENT/LAP PACK, Model Number: DYNJ84876; 19) TONSIL PACK, Model Number: DYNJ50109B; 20) TONSIL, Model Number: DYNJ908452C; 21) CSS SINUS PACK, Model Number: DYNJ17005I; 22) ENT, Model Number: DYNJ910514; 23) ADULT ENT - OAKBEND, Model Number: DYNJ905762; 24) EENT SPLIT PACK-LF, Model Number: LLSCAENT01; 25) ENT PACK, Model Number: DYNJ31453D; 26) ENT PACK, Model Number: DYNJ59447A; 27) TONSIL PACK, Model Number: DYNJ26441R; 28) ORAL ENT PACK, Model Number: DYNJ58408A; 29) ENT PACK, Model Number: DYNJ51085; 30) DR LANO NASAL CHRISTUS, Model Number: DYNJ61196A; 31) ENT PACK, Model Number: DYNJ66178D; 32) BWNBORO EENT PACK-LF, Model Number: DYNJ37551A; 33) HEAD AND NECK PACK, Model Number: DYNJ68872F; 34) EENT PACK, Model Number: DYNJ50301C; 35) ENT PACK, Model Number: DYNJ40498B; 36) SHN-C-THYROIDECTOMY-LF, Model Number: DYNJ46869A; 37) MINOR CYSTO PACK, Model Number: DYNJ58727D; 38) TONSILS AND ADENOIDS, Model Number: DYNJ45171C; 39) T&A PACK, Model Number: DYNJ56543J; 40) ENT PACK, Model Number: DYNJ54035B; 41) TONSIL PACK, Model Number: DYNJ50109B; 42) ENT PACK, Model Number: DYNJ40139D; 43) NASAL/SEPTO PACK, Model Number: DYNJ32099A; 44) EENT SPLIT PACK-LF, Model Number: LLSCAENT01; 45) ENT PACK, Model Number: DYNJ53573A; 46) CHS ENT PACK II, Model Number: DYNJCH908; 47) SIMPLE NECK DISSECTION, Model Number: DYNJ47481F; 48) TONSIL PACK, Model Number: DYNJ81479A; 49) E.N.T. TRAY, Model Number: DYNJ22357K; 50) NASAL PACK, Model Number: DYNJ39317B; 51) HEAD AND NECK PACK, Model Number: DYNJ54321D; 52) T A PACK, Model Number: DYNJ67791A; 53) ENT PACK, Model Number: DYNJ42451B; 54) HEAD AND NECK PACK-FOX, Model Number: DYNJ53348C; 55) HEAD AND NECK PACK, Model Number: DYNJ58596B; 56) SINUS PACK, Model Number: DYNJ56262; 57) ENT PACK, Model Number: DYNJ51085; 58) TONSIL T AND A PACK SHD, Model Number: DYNJ68417C; 59) ENT PACK, Model Number: DYNJ48595; 60) T AND A PACK-LF, Model Number: DYNJ52611; 61) HEAD AND NECK PACK, Model Number: DYNJ81897B; 62) ENSEMBLE ORL THYROIDE-LF, Model Number: DYNJ43390A; 63) EENT PACK-LF, Model Number: DYNJ34728D; 64) ENT, Model Number: DYNJ910514; 65) TONSIL ADENOID CKHA-LF, Model Number: DYNJ41771C; 66) ENT PACK, Model Number: DYNJ68679B; 67) PACK,T & A ASC, Model Number: DYNJ906914; 68) EENT SPLIT PACK-LF, Model Number: LLSCAENT01; 69) ENT PACK, Model Number: DYNJ83603; 70) ENT PACK, Model Number: DYNJ32247; 71) PPM-ICD PACK-LF, Model Number: DYNJ15949D; 72) ENT PACK, Model Number: DYNJ31453D; 73) SINOSCOPY PACK-LF, Model Number: DYNJ25600L; 74) T & A PACK, Model Number: DYNJ38978C; 75) CSC-N T&A, Model Number: DYNJ50870D; 76) ENT I-LF, Model Number: DYNJS0702; 77) ENT KIT, Model Number: DYNJ902836F; 78) ENT PACK, Model Number: DYNJ66039B; 79) TONSIL CDS, Model Number: DYNJ907844D; 80) PACK,T & A ASC, Model Number: DYNJ906914

Code information (models, lots, serial numbers)

Show the full code information (7k characters)

1) DYNJ67791A, UDI-DI: 10193489941715(each), 40193489941716(case), Lot Number: 24IBE758; 2) DYNJ66514, UDI-DI: 10193489359282(each), 40193489359283(case), Lot Number: 24IBE817; 3) DYNJ67775A, UDI-DI: 10193489996258(each), 40193489996259(case), Lot Number: 24IBG346; 4) DYNJ83158, UDI-DI: 10195327236885(each), 40195327236886(case), Lot Number: 24IBG449; 5) DYNJ0634878I, UDI-DI: 10195327694500(each), 40195327694501(case), Lot Number: 24IBG548; 6) DYNJ15949D, UDI-DI: 10195327573805(each), 40195327573806(case), Lot Number: 24IBG570; 7) DYNJ44525C, UDI-DI: 10193489232042(each), 40193489232043(case), Lot Number: 24IBG608; 8) DYNJ44391K, UDI-DI: 10195327457648(each), 40195327457649(case), Lot Number: 24IBG848; 9) DYNJ49300D, UDI-DI: 10193489297287(each), 40193489297288(case), Lot Number: 24IBH734; 10) DYNJ61555B, UDI-DI: 10195327608101(each), 40195327608102(case), Lot Number: 24IBH769; 11) DYNJ57287A, UDI-DI: 10193489854152(each), 40193489854153(case), Lot Number: 24IBH959; 12) DYNJ36720F, UDI-DI: 10198459079306(each), 40198459079307(case), Lot Number: 24IBQ081; 13) DYNJ54035B, UDI-DI: 10195327680541(each), 40195327680542(case), Lot Number: 24IBQ933; 14) DYNJ36516, UDI-DI: 10884389544112(each), 40884389544113(case), Lot Number: 24IBR105; 15) DYNJ59502B, UDI-DI: 10193489372939(each), 40193489372930(case), Lot Number: 24IBR121; 16) DYNJ80790, UDI-DI: 10195327079482(each), 40195327079483(case), Lot Number: 24IBR337; 17) DYNJ57583, UDI-DI: 10889942755332(each), 40889942755333(case), Lot Number: 24IBR873; 18) DYNJ84876, UDI-DI: 10195327431778(each), 40195327431779(case), Lot Number: 24IBR895; 19) DYNJ50109B, UDI-DI: 10198459115776(each), 40198459115777(case), Lot Number: 24IBS352; 20) DYNJ908452C, UDI-DI: 10195327610937(each), 40195327610938(case), Lot Number: 24JBD799; 21) DYNJ17005I, UDI-DI: 10193489241860(each), 40193489241861(case), Lot Number: 24JBF130; 22) DYNJ910514, UDI-DI: 10198459021411(each), 40198459021412(case), Lot Number: 24JBK654; 23) DYNJ905762, UDI-DI: 10193489507966(each), 40193489507967(case), Lot Number: 24JBK682; 24) LLSCAENT01, UDI-DI: 10080196120802(each), 40080196120803(case), Lot Number: 24JBK897; 25) DYNJ31453D, UDI-DI: 10195327286224(each), 40195327286225(case), Lot Number: 24JBM367; 26) DYNJ59447A, UDI-DI: 10193489707663(each), 40193489707664(case), Lot Number: 24JBM926; 27) DYNJ26441R, UDI-DI: 10198459085888(each), 40198459085889(case), Lot Number: 24JBN335; 28) DYNJ58408A, UDI-DI: 10889942969029(each), 40889942969020(case), Lot Number: 24JBN611; 29) DYNJ51085, UDI-DI: 10889942117284(each), 40889942117285(case), Lot Number: 24JBN781; 30) DYNJ61196A, UDI-DI: 10195327604240(each), 40195327604241(case), Lot Number: 24JBN797; 31) DYNJ66178D, UDI-DI: 10198459100307(each), 40198459100308(case), Lot Number: 24JBN811; 32) DYNJ37551A, UDI-DI: 10888277585720(each), 40888277585721(case), Lot Number: 24JBT022; 33) DYNJ68872F, UDI-DI: 10198459079269(each), 40198459079260(case), Lot Number: 24JBT065; 34) DYNJ50301C, UDI-DI: 10198459052330(each), 40198459052331(case), Lot Number: 24JBT456; 35) DYNJ40498B, UDI-DI: 10193489806779(each), 40193489806770(case), Lot Number: 24JBV199; 36) DYNJ46869A, UDI-DI: 10198459069949(each), 40198459069940(case), Lot Number: 24JBV309; 37) DYNJ58727D, UDI-DI: 10198459048425(each), 40198459048426(case), Lot Number: 24JBV608; 38) DYNJ45171C, UDI-DI: 10889942485260(each), 40889942485261(case), Lot Number: 24JBV814; 39) DYNJ56543J, UDI-DI: 10195327453602(each), 40195327453603(case), Lot Number: 24JBW129; 40) DYNJ54035B, UDI-DI: 10195327680541(each), 40195327680542(case), Lot Number: 24JBW183; 41) DYNJ50109B, UDI-DI: 10198459115776(each), 40198459115777(case), Lot Number: 24JBW623; 42) DYNJ40139D, UDI-DI: 10195327663506(each), 40195327663507(case), Lot Number: 24JBX313; 43) DYNJ32099A, UDI-DI: 10195327610890(each), 40195327610891(case), Lot Number: 24JBY276; 44) LLSCAENT01, UDI-DI: 10080196120802(each), 40080196120803(case), Lot Number: 24JBY278; 45) DYNJ53573A, UDI-DI: 10193489201635(each), 40193489201636(case), Lot Number: 24JBY404; 46) DYNJCH908, UDI-DI: 10193489393347(each), 40193489393348(case), Lot Number: 24KBA907; 47) DYNJ47481F, UDI-DI: 10195327067335(each), 40195327067336(case), Lot Number: 24KBB421; 48) DYNJ81479A, UDI-DI: 10195327378370(each), 40195327378371(case), Lot Number: 24KBB455; 49) DYNJ22357K, UDI-DI: 10193489926736(each), 40193489926737(case), Lot Number: 24KBB793; 50) DYNJ39317B, UDI-DI: 10889942775729(each), 40889942775720(case), Lot Number: 24KBC590; 51) DYNJ54321D, UDI-DI: 10195327679323(each), 40195327679324(case), Lot Number: 24KBC663; 52) DYNJ67791A, UDI-DI: 10193489941715(each), 40193489941716(case), Lot Number: 24KBC754; 53) DYNJ42451B, UDI-DI: 10193489740318(each), 40193489740319(case), Lot Number: 24KBC910; 54) DYNJ53348C, UDI-DI: 10195327453565(each), 40195327453566(case), Lot Number: 24KBD155; 55) DYNJ58596B, UDI-DI: 10195327635947(each), 40195327635948(case), Lot Number: 24KBD243; 56) DYNJ56262, UDI-DI: 10889942624331(each), 40889942624332(case), Lot Number: 24KBD661; 57) DYNJ51085, UDI-DI: 10889942117284(each), 40889942117285(case), Lot Number: 24KBE334; 58) DYNJ68417C, UDI-DI: 10195327341633(each), 40195327341634(case), Lot Number: 24KBE396; 59) DYNJ48595, UDI-DI: 10888277915183(each), 40888277915184(case), Lot Number: 24KBE590; 60) DYNJ52611, UDI-DI: 10889942300853(each), 40889942300854(case), Lot Number: 24KBF355; 61) DYNJ81897B, UDI-DI: 10198459155710(each), 40198459155711(case), Lot Number: 24KBF577; 62) DYNJ43390A, UDI-DI: 10889942670321(each), 40889942670322(case), Lot Number: 24KBG019; 63) DYNJ34728D, UDI-DI: 10198459061851(each), 40198459061852(case), Lot Number: 24KBG035; 64) DYNJ910514, UDI-DI: 10198459021411(each), 40198459021412(case), Lot Number: 24KBG072; 65) DYNJ41771C, UDI-DI: 10889942599189(each), 40889942599180(case), Lot Number: 24KBG178; 66) DYNJ68679B, UDI-DI: 10195327129095(each), 40195327129096(case), Lot Number: 24KBG186; 67) DYNJ906914, UDI-DI: 10193489280173(each), 40193489280174(case), Lot Number: 24KBG371; 68) LLSCAENT01, UDI-DI: 10080196120802(each), 40080196120803(case), Lot Number: 24KBG399; 69) DYNJ83603, UDI-DI: 10195327284886(each), 40195327284887(case), Lot Number: 24KBG953; 70) DYNJ32247, UDI-DI: 10884389203873(each), 40884389203874(case), Lot Number: 24KBG965; 71) DYNJ15949D, UDI-DI: 10195327573805(each), 40195327573806(case), Lot Number: 24KBG993; 72) DYNJ31453D, UDI-DI: 10195327286224(each), 40195327286225(case), Lot Number: 24KBG996; 73) DYNJ25600L, UDI-DI: 10889942686247(each), 40889942686248(case), Lot Number: 24KBH226; 74) DYNJ38978C, UDI-DI: 10195327583590(each), 40195327583591(case), Lot Number: 24KBH242; 75) DYNJ50870D, UDI-DI: 10195327165666(each), 40195327165667(case), Lot Number: 24KBH280; 76) DYNJS0702, UDI-DI: 10080196798728(each), 40080196798729(case), Lot Number: 24KBI176; 77) DYNJ902836F, UDI-DI: 10198459002816(each), 40198459002817(case), Lot Number: 24KBI638; 78) DYNJ66039B, UDI-DI: 10195327151997(each), 40195327151998(case), Lot Number: 24KBL783; 79) DYNJ907844D, UDI-DI: 10195327387334(each), 40195327387335(case), Lot Number: 24KBP945; 80) DYNJ906914, UDI-DI: 10193489280173(each), 40193489280174(case), Lot Number: 24LBA080

Devices named in this recall

71 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 96510).