CELL-DYN recallZ-1654-2025 · Abbott Laboratories · initiated Mar 18, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
FDA root cause category: Software change control
What the company is doing
On March 18, 2025, the firm notified customers through "Urgent Medical Device Correction" letters delivered via FedEx. The letters described the product, problem and actions to be taken. The Customers were instructed to not use expired reagents as this may compromise test accuracy. Always use reagents within their valid expiration dates for all testing procedures. The operator must confirm the container labeling matches the data on the reagent log entry screen dialog box. Quality Controls should be run after any reagent lot number change, maintenance, component replacement, field service action, software change, or calibration. This should be done in accordance with your laboratory s quality control program and regulatory requirements. Complete and return the Customer Reply Form via email: [email protected] or fax: 1-800-777-0051. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction, provide to them a copy of this letter and retain this letter for your laboratory records. If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT.
Recall details
- Recall number
- Z-1654-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 18, 2025
- Posted by the FDA
- Apr 22, 2025
- Recalling firm
- Abbott Laboratories
- Quantity
- 19
- Product code
- GKZ
- FDA submissions
- K061667
- Distribution
- Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.
Product description
Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-10 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
Code information (models, lots, serial numbers)
UDI-DI 00380740017170 SNs: 35405BG 36034BG 36259BG 36326BG 36327BG 36328BG 36372BG 36545BG 36556BG 36583BG 36588BG 36589BG 54539BG 55361BG 55454BG 55520BG 55814BG 55828BG 72229BG
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CELL-DYN 08H67-01Abbott Laboratories · Counter, differential cell6 recalls269 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96567).
- Z-1652-2025 · Open, ClassifiedBrand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-01 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
- Z-1653-2025 · Open, ClassifiedBrand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-03 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
- Z-1655-2025 · Open, ClassifiedBrand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-13 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No
- Z-1656-2025 · Open, ClassifiedBrand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page