Skip to content

CELL-DYN recallZ-1655-2025 · Abbott Laboratories · initiated Mar 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Laboratories or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

FDA root cause category: Software change control

What the company is doing

On March 18, 2025, the firm notified customers through "Urgent Medical Device Correction" letters delivered via FedEx. The letters described the product, problem and actions to be taken. The Customers were instructed to not use expired reagents as this may compromise test accuracy. Always use reagents within their valid expiration dates for all testing procedures. The operator must confirm the container labeling matches the data on the reagent log entry screen dialog box. Quality Controls should be run after any reagent lot number change, maintenance, component replacement, field service action, software change, or calibration. This should be done in accordance with your laboratory s quality control program and regulatory requirements. Complete and return the Customer Reply Form via email: [email protected] or fax: 1-800-777-0051. If you have forwarded the product listed above to other laboratories, please inform them of this Product Correction, provide to them a copy of this letter and retain this letter for your laboratory records. If you or any of the health care providers you serve have questions regarding this information, please contact Customer Service at 1-877-4ABBOTT.

Recall details

Recall number
Z-1655-2025
Status
Open, Classified
Initiated by the company
Mar 18, 2025
Posted by the FDA
Apr 22, 2025
Recalling firm
Abbott Laboratories
Quantity
1
Product code
GKZ
FDA submissions
K061667
Distribution
Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

Product description

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 08H67-13 Software Version: All software versions Product Description: CELL-DYN Ruby Component: No

Code information (models, lots, serial numbers)

UDI-DI 00380740017170 SN: 34036BG96

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 96567).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page