Skip to content

MAGEC recallZ-1657-2025 · Globus Medical, Inc. · initiated Apr 3, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Globus Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On April 3, 2025, URGENT: MEDICAL DEVICE NOTIFICATION letter was sent to their customer. Actions to be taken by Customers/Users The customer is instructed to monitor their patient closely for any potential adverse events. Report any adverse events believed to be associated with this notification to Globus Medical. In addition, any adverse events associated with the use of this product and/or quality problems should also be reported to the FDA's MedWatch Program: MedWatch: The FDA Safety Information and Adverse Event Reporting Program; For Questions and Support: Globus Medical Complaint Handling Unit: Email: [email protected] Globus Medical is committed to serving your product needs and our primary objectives are patient safety, exceptional product reliability, and the highest level of customer satisfaction. Thank you for your prompt support on this important matter. We appreciate your cooperation.

Recall details

Recall number
Z-1657-2025
Status
Open, Classified
Initiated by the company
Apr 3, 2025
Posted by the FDA
Apr 23, 2025
Recalling firm
Globus Medical, Inc.
Quantity
1 unit
Product code
PGN
FDA submissions
K171791
Distribution
US Nationwide distribution in the state of AR.

Product description

MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;

Code information (models, lots, serial numbers)

Version/Model: MC2-5090S; UDI-DI (with lot nuber): (01)00887517834041(11)250311(17)300311(10)GB1607ET

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K171791 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.