MAGEC MC2-5070RRecalls & reported problems
Nuvasive, Inc. · MAGEC 2 Rod, 5.0mm 70mm Offset · Class 2 · Implant
All models of MAGEC (4)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00887517834010), received between Nov 11, 2019 and May 26, 2023 — classified as 4 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K171791); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Break2
- Mechanical Problem2
- Corroded1
- Defective Device1
- Mechanical Jam1
- Unstable1
Patient problems
- Failure of Implant5
- Metal Related Pathology2
- Unspecified Infection2
- Swelling1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 26, 2023Break; Corroded; Mechanical Jam; Break; Corroded; Mechanical Jam
AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION. HIGH ENERGY X-RAY EVALUATION SHOWED EVIDENCE OF INTERNAL SCREW FRACTURE. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 17015004 ↗InjuryMar 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT DURING REMOVAL FOR DEFINITIVE FUSION AFTER TREATMENT GOALS WERE ACHIEVED, IT WAS NOTED THERE WAS "METALLOSIS" PRESENT.
Patient problems: Metal Related Pathology
MAUDE report 16513658 ↗InjuryAug 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT THERE WERE SIGNS OF METALLOSIS. THE ROD HAS BEEN EXPLANTED AND REPLACED WITH A TRADITIONAL GROWING ROD. REPORT 2 OF 2.
Patient problems: Metal Related Pathology
MAUDE report 15259556 ↗MalfunctionJul 13, 2021Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED THAT THE ROD WAS NOT LENGTHENING. NO PATIENT HARM WAS REPORTED.
Patient problems: Failure of Implant
MAUDE report 12161111 ↗MalfunctionJun 29, 2021Mechanical Problem; Mechanical Problem
INFORMATION WAS RECEIVED THAT A ROD FAILED TO DISTRACT SUBSEQUENT TO THE INITIAL DISTRACTION SESSION. NO PATIENT IMPACT WAS REPORTED BUT A REVISION SURGERY IS ANTICIPATED. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Failure of Implant
MAUDE report 12087612 ↗MalfunctionMar 1, 2021Unstable; Unstable
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020 DUE TO PULLOUT. DURING A REVIEW OF THE INVESTIGATION FINDINGS FROM (B)(4) THERE WAS NO PROBLEM FOUND.
Patient problems: Failure of Implant
MAUDE report 11397112 ↗MalfunctionDec 9, 2020Break; Break
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED. AS PER THE REPORTER, THE ROD WAS BROKEN. REPORTEDLY, WHEN THE ROD WAS EXPLANTED IT WAS NOTICEABLY BROKEN AT THE PART WHERE THE ROD COMES OUT WHEN DISTRACTING. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: Failure of Implant
MAUDE report 10979237 ↗InjuryMar 13, 2020Defective Device; Defective Device
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2019. AS PER THE REPORTER, THE ROD WAS REMOVED DUE TO INFECTION.
Patient problems: Unspecified Infection
MAUDE report 9831128 ↗InjuryNov 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2019. AS PER REPORTER THE PATIENT WAS EXPERIENCING POSTOPERATIVE INFECTION, FEVER, AND SWELLING.
Patient problems: Swelling; Unspecified Infection
MAUDE report 9307530 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K171791. One submission often covers many models, so check the product and lot details in each record.
- Z-1657-2025 · Apr 3, 2025 · Open, ClassifiedMAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;
- Z-1898-2020 · Feb 13, 2020 · TerminatedNUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spinal rod that is surgically implanted using appropriate NuVa
Other Growing rod system- magnetic actuation devices
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation112 reports
- MAGEC MC2-5090SNuvasive, Inc. · Growing rod system- magnetic actuation1 recall33 reports
- MAGEC® Spinal Bracing and Distraction System RA002-4545SLNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation82 reports
- MAGEC® Spinal Bracing and Distraction System RA002-5555SLRNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation81 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall26 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4590SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall25 reports
- MAGEC® Spinal Bracing and Distraction System MC2-4570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
- MAGEC® Spinal Bracing and Distraction System MC2-5570SNuvasive Specialized Orthopedics, Inc. · Growing Rod System- Magnetic Actuation1 recall14 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page