Skip to content

MAGEC MC2-5070RRecalls & reported problems

Nuvasive, Inc. · MAGEC 2 Rod, 5.0mm 70mm Offset · Class 2 · Implant

All models of MAGEC (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00887517834010), received between Nov 11, 2019 and May 26, 2023 — classified as 4 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K171791); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem3
  • Break2
  • Mechanical Problem2
  • Corroded1
  • Defective Device1
  • Mechanical Jam1
  • Unstable1

Patient problems

  • Failure of Implant5
  • Metal Related Pathology2
  • Unspecified Infection2
  • Swelling1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2023Break; Corroded; Mechanical Jam; Break; Corroded; Mechanical Jam

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, FORCE TESTING WAS UNABLE TO BE PERFORMED. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION. HIGH ENERGY X-RAY EVALUATION SHOWED EVIDENCE OF INTERNAL SCREW FRACTURE. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 17015004 ↗
  • InjuryMar 9, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED THAT DURING REMOVAL FOR DEFINITIVE FUSION AFTER TREATMENT GOALS WERE ACHIEVED, IT WAS NOTED THERE WAS "METALLOSIS" PRESENT.

    Patient problems: Metal Related Pathology

    MAUDE report 16513658 ↗
  • InjuryAug 19, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED THAT THERE WERE SIGNS OF METALLOSIS. THE ROD HAS BEEN EXPLANTED AND REPLACED WITH A TRADITIONAL GROWING ROD. REPORT 2 OF 2.

    Patient problems: Metal Related Pathology

    MAUDE report 15259556 ↗
  • MalfunctionJul 13, 2021Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD WAS NOT LENGTHENING. NO PATIENT HARM WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12161111 ↗
  • MalfunctionJun 29, 2021Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT A ROD FAILED TO DISTRACT SUBSEQUENT TO THE INITIAL DISTRACTION SESSION. NO PATIENT IMPACT WAS REPORTED BUT A REVISION SURGERY IS ANTICIPATED. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 12087612 ↗
  • MalfunctionMar 1, 2021Unstable; Unstable

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020 DUE TO PULLOUT. DURING A REVIEW OF THE INVESTIGATION FINDINGS FROM (B)(4) THERE WAS NO PROBLEM FOUND.

    Patient problems: Failure of Implant

    MAUDE report 11397112 ↗
  • MalfunctionDec 9, 2020Break; Break

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED. AS PER THE REPORTER, THE ROD WAS BROKEN. REPORTEDLY, WHEN THE ROD WAS EXPLANTED IT WAS NOTICEABLY BROKEN AT THE PART WHERE THE ROD COMES OUT WHEN DISTRACTING. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10979237 ↗
  • InjuryMar 13, 2020Defective Device; Defective Device

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2019. AS PER THE REPORTER, THE ROD WAS REMOVED DUE TO INFECTION.

    Patient problems: Unspecified Infection

    MAUDE report 9831128 ↗
  • InjuryNov 11, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2019. AS PER REPORTER THE PATIENT WAS EXPERIENCING POSTOPERATIVE INFECTION, FEVER, AND SWELLING.

    Patient problems: Swelling; Unspecified Infection

    MAUDE report 9307530 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171791. One submission often covers many models, so check the product and lot details in each record.

  • Z-1657-2025 · Apr 3, 2025 · Open, Classified
    MAGEC 2 Rod, 5.0mm 90mm Standard; Version/Model: MC2-5090S;
  • Z-1898-2020 · Feb 13, 2020 · Terminated
    NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spinal rod that is surgically implanted using appropriate NuVa

Other Growing rod system- magnetic actuation devices

All Growing rod system- magnetic actuation devices →

Sources and updates