Nuvasive, Inc. · MAGEC 2 Rod, 5.0mm 90mm Standard · Class 2 · Implant

All models of MAGEC (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Nuvasive MAGEC 2 growing rod is a magnetically actuated spinal rod that lengthens a child's spine without repeat surgery. The 33 reports mostly describe rods that would not lengthen, corroded or worn surfaces, and locking pin fractures, with several rods removed or revised after failing to distract; 4 reports involve injuries and no deaths. One open recall, started April 2025, covers a lot assembled with a 5.5mm rod instead of the required 5.0mm.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
33
Deaths reported
0
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 00887517834041), received between Nov 5, 2018 and Feb 29, 2024 — classified as 4 injuries, 29 malfunctions.

The most frequently reported specific device problem is “Mechanical Problem” (8 reports).

1 FDA recall record name this device's identifier, 1 still open.

1 other recall cover devices cleared under the same FDA submission (K171791); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 2 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Apr 3, 2025Open, Classified

    Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.

    Z-1657-2025 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported problems

Device problems

  • Mechanical Problem8
  • Corroded6
  • Migration or Expulsion of Device6
  • Migration5
  • Fracture4
  • Defective Device3
  • Mechanical Jam3
  • Failure to Advance2

Patient problems

  • Failure of Implant29
  • No Clinical Signs, Symptoms or Conditions7
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available2
  • Insufficient Information2
  • No Known Impact Or Consequence To Patient1
  • Pain1
  • Unspecified Infection1

Recent adverse event reports

The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 29, 2024Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18807776 ↗
  • MalfunctionFeb 29, 2024Failure to Advance; Failure to Advance

    NO ADDITIONAL INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18807744 ↗
  • InjuryJun 16, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT FORCE TESTING WAS UNABLE TO BE COMPLETED DUE TO THE EXTERNAL REMOTE CONTROLLER (ERC) NOT FUNCTIONING. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION ON THE ROD. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER DESIRED LENGTHENING HAD BEEN ACHIEVED AS WELL AS THE PATIENT EXPERIENCING PAIN. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant; Pain

    MAUDE report 17148894 ↗
  • MalfunctionMay 31, 2023Break; Corroded; Fracture; Mechanical Jam; Break; Corroded; Fracture; Mechanical Jam

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION, THE ROD WOULD NOT DISTRACT AND FORCE TESTING WAS UNABLE TO BE PERFORMED AS THE ELECTRONIC REMOTE CONTROLLER (ERC) WAS NOT FUNCTIONAL. ADDITIONALLY, IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS CORROSION. IT WAS REPORTED THAT THE ROD WAS REMOVED DUE TO A LOCKING PIN FRACTURE. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 17032141 ↗
  • MalfunctionMay 25, 2023Corroded; Mechanical Jam; Corroded; Mechanical Jam

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT THE ROD WAS REMOVED DUE TO THE PATIENT NOT COMPLYING WITH DISTRACTION. FORCE TESTING WAS UNABLE TO BE COMPLETED AS THE ROD WOULD NOT EXTEND. IT WAS NOTED DURING SURFACE DAMAGE GRADING THAT THERE WAS GALLING AND CORROSION ON THE ROD. UPON DISASSEMBLY, THE ROD WAS FOUND TO HAVE SOME METALLIC CORROSION AROUND THE MAGNET AREA. THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17004230 ↗
  • MalfunctionJun 22, 2022Corroded; Naturally Worn; Corroded; Naturally Worn

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT DURING EXAMINATION OF THE ROD, THE RETRIEVAL CENTER FOUND WEAR AND OTHER SURFACE DAMAGE. IT WAS REPORTED THAT THE ROD WAS REMOVED AS PART OF A PLANNED REMOVAL PROCEDURE AFTER THE ROD HAD FULLY EXTENDED; THERE WAS NO ADVERSE PATIENT OR USER IMPACT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 14779886 ↗
  • InjuryJun 8, 2022Mechanical Problem; Naturally Worn; Scratched Material; Mechanical Problem; Naturally Worn; Scratched Material

    AN INVESTIGATION REPORT FROM AN IMPLANT RETRIEVAL CENTER WAS RECEIVED AND REPORTED THAT A REVISION PROCEDURE WAS PERFORMED DUE TO THE ROD FAILING TO DISTRACT. DURING EXAMINATION OF THE ROD, THE RETRIEVAL CENTER FOUND THE ROD UNABLE TO DISTRACT. ADDITIONALLY, DISCOLORATION, WEAR, SCRATCHES, AND OTHER SURFACE DAMAGE WAS NOTED. THERE WAS NO FURTHER REPORTED PATIENT IMPACT AND NO PATIENT OR USER INJURY WAS REPORTED. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 1 OF 2.

    Patient problems: Failure of Implant

    MAUDE report 14629745 ↗
  • MalfunctionApr 13, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD WOULD NOT DISTRACT. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14104387 ↗
  • MalfunctionMar 29, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT A REVISION PROCEDURE WAS PERFORMED. AS PER THE REPORTER, THE ROD HAD AN END CAP FAILURE. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13938678 ↗
  • MalfunctionJan 6, 2022Fracture; Mechanical Problem; Fracture; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD DID NOT FULLY ACHIEVE THE DESIRED DISTRACTION INCREMENTS DURING LENGTHENING SESSIONS. X-RAYS REVEALED THE RIGHT ROD HAD A MAGNET FRACTURE AND A BROKEN PIN. THE DEVICE WAS REMOVED FOR FINAL FUSION. ADDITIONALLY, IT WAS REPORTED THAT THERE WAS ORGANIC MATTER BUILDUP AROUND THE ROD AND TISSUE DISCOLORATION. NO PATIENT ADVERSE REACTION OR IMPACT WAS REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant

    MAUDE report 13182298 ↗
  • MalfunctionJan 6, 2022Mechanical Problem; Mechanical Problem

    INFORMATION WAS RECEIVED THAT THE ROD DID NOT FULLY ACHIEVE THE DESIRED DISTRACTION INCREMENTS DURING LENGTHENING SESSIONS. THE DEVICE WAS REMOVED FOR FINAL FUSION. ADDITIONALLY, IT WAS REPORTED THAT THERE WAS ORGANIC MATTER BUILDUP AROUND THE ROD AND TISSUE DISCOLORATION. NO PATIENT ADVERSE REACTION OR IMPACT WAS REPORTED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant

    MAUDE report 13182236 ↗
  • MalfunctionNov 8, 2021Mechanical Problem; Mechanical Problem

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 12775224 ↗
  • MalfunctionNov 3, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    NO ADDITIONAL INFORMATION PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 12745318 ↗
  • MalfunctionOct 6, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    INFORMATION WAS RECEIVED THAT THE ROD'S END CAP DISENGAGED. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12584561 ↗
  • MalfunctionSep 20, 2021Fracture; Fracture

    INFORMATION WAS RECEIVED THAT THE ROD WAS FRACTURED. NO PATIENT ADVERSE EVENT WAS REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 12497740 ↗
  • MalfunctionAug 12, 2021Corroded; Corroded

    NO ADDITIONAL INFORMATION WAS PROVIDED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12315816 ↗
  • MalfunctionAug 12, 2021Corroded; Corroded

    INFORMATION WAS RECEIVED THAT A REVISION WAS PERFORMED FOR A FINAL FUSION. AS PER THE REPORTER, UPON EXPLANT THE ROD WAS NOTED TO HAVE CORROSION AND THERE WAS METAL WEAR DEBRIS IN THE TISSUE. NO PATIENT HARM WAS REPORTED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12315505 ↗
  • MalfunctionApr 5, 2021Migration or Expulsion of Device; Migration; Migration or Expulsion of Device; Migration

    NO NEW INFORMATION PROVIDED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11615120 ↗
  • MalfunctionApr 5, 2021Migration or Expulsion of Device; Migration; Migration or Expulsion of Device; Migration

    NO NEW INFORMATION PROVIDED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11615119 ↗
  • InjuryMar 1, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THE END-CAP WAS SEPARATED.

    Patient problems: Failure of Implant

    MAUDE report 11397121 ↗
  • MalfunctionMar 1, 2021Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020. AS PER THE REPORTER THE ROD WAS NOT LENGTHENING. DURING A REVIEW OF INVESTIGATION FINDINGS FROM LIRC IT WAS IDENTIFIED THAT THERE WAS NO DEVICE PROBLEM.

    Patient problems: Failure of Implant

    MAUDE report 11397093 ↗
  • MalfunctionDec 30, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON (B)(6) 2020 DUE TO AN INFECTION. THERE WAS NO PATIENT INJURY REPORTED. DURING A REVIEW OF INVESTIGATION FINDINGS FROM (B)(6) THE ROD FUNCTIONED AS INTENDED.

    Patient problems: Unspecified Infection

    MAUDE report 11098828 ↗
  • MalfunctionDec 28, 2020Unstable; Unstable

    IT WAS REPORTED THAT A REVISION PROCEDURE WAS PERFORMED ON AN (B)(6) 2019. AS PER THE REPORTER THE ROD PULLED OUT. THERE WAS NO PATIENT INJURY REPORTED. DURING A REVIEW OF INVESTIGATION FINDINGS FROM (B)(6) THE ROD FUNCTIONED AS INTENDED.

    Patient problems: Failure of Implant

    MAUDE report 11085371 ↗
  • MalfunctionNov 20, 2020Migration; Migration

    INFORMATION WAS RECEIVED THAT X-RAY IMAGES REVEALED THAT THE END CAP HAD DISENGAGED. NO REVISION PROCEDURE HAS BEEN SCHEDULED AT THIS TIME. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: Failure of Implant

    MAUDE report 10880798 ↗
  • MalfunctionAug 15, 2020Migration or Expulsion of Device; Migration or Expulsion of Device

    ADDITIONAL INFORMATION PROVIDED: H2, H3, H6, H10.

    Patient problems: Failure of Implant

    MAUDE report 10411801 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K171791. One submission often covers many models, so check the product and lot details in each record.

  • Z-1898-2020 · Feb 13, 2020 · Terminated
    NUVASIVE MAGEC 2 Rod X.X mm XX mm, REF MC2-XXXXS LOT 0000000 AAA - Product Usage: comprised of a sterile single-use spinal rod that is surgically implanted using appropriate NuVa

Other Growing rod system- magnetic actuation devices

All Growing rod system- magnetic actuation devices →