GE Healthcare Carestation 650c A1, Model/REF Number… recallZ-1634-2025 · GE Medical Systems China Co., Ltd. · initiated Mar 21, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from GE Medical Systems China Co., Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
FDA root cause category: Device Design
What the company is doing
GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice on 03/21/2025. The notice explained the safety issue with the device, potential risk, and requested the following actions: The anesthesia system may continue to be used following these instructions: Perform the Ventilation Screening Test specified in Appendix 1 for each affected Carestation system. If the Carestation system passes the Ventilation Screening Test, you can continue to use the device. If the Carestation system fails the Ventilation Screening Test and must be used prior to the system being corrected by GE HealthCare, follow the instructions below: 1. Use the device in accordance with the instructions in the URM with these changes: i. Use only Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) mode to mechanically ventilate a patient. ii. Do not use Volume Control Ventilation (VCV) mode to mechanically ventilate a patient. NOTE: Manual mode of the anesthesia system can be used to provide manual ventilation or allow spontaneous ventilation of the patient. 2. Users are not to use VCV mode on the device until the device has been corrected by GE HealthCare. Please ensure all potential staff in your facility are made aware of this safety notification and the recommended actions and retain this document for your record. Questions: contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative. Notice sent on 04/21/25 with updated instructions. Change: Attach a 2 Liter reservoir bag (original) to Attach a reservoir bag (updated)in step 5 of the Ventilation Screening Test in Appendix 1 of the letter. Notice sent on 5/18/26 added additional affected devices for GE HealthCare Urgent Medical Device Correction notification, Field Replaceable Unit (FRU) parts: DRIVER GAS CHECK VALVE ASSY, M1807901-S; Valve Assembly, 3.5cm H2O Drive Gas Check Valve BCG, 2071125-001-S; A180 Ventilator Engine Assembly, 2071117-001-S.
Recall details
- Recall number
- Z-1634-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 21, 2025
- Posted by the FDA
- Apr 29, 2025
- Recalling firm
- GE Medical Systems China Co., Ltd.
- Quantity
- 19 units
- Product code
- BSZ
- FDA submissions
- K151570
- Distribution
- Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T¿rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
Product description
GE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200. Anesthesia system.
Code information (models, lots, serial numbers)
GTIN 00195278439543, Serial Numbers: SM824450008WD, SM824450009WD, SM824450010WD, SM824450011WD, SM824450012WD, SM824450013WD, SM824450014WD, SM824450015WD, SM823300001MD, SM824360002WD, SM824360004WD, SM824360005WD, SM824360006WD, SM824360007WD, SM825020002MD, SM825020003MD, SM825020004MD, SM825020005MD, SM825020006MD.
Other devices under the same FDA submission
3 devices cleared under K151570 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,109 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,015 reports
- Carestation 650c A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls70 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96634).
- Z-1624-2025 · Open, ClassifiedGE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systems
- Z-1627-2025 · Open, ClassifiedGE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.
- Z-1628-2025 · Open, ClassifiedGE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.
- Z-1631-2025 · Open, ClassifiedGE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.
- Z-1632-2025 · Open, ClassifiedGE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.
- Z-1635-2025 · Open, ClassifiedGE Healthcare CARESTATION 750 A1, Model/REF Number 1012-9750-000. Anesthesia system.
- Z-1637-2025 · Open, ClassifiedGE Healthcare CARESTATION 750c A1, Model/REF Number 1012-9755-000. Anesthesia system.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page