Carestation 650c A1Recalls & reported problems
Datex-Ohmeda Inc · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Datex-Ohmeda Carestation 650c A1 is an anesthesia gas machine used in operating rooms. The 70 reports mostly describe battery and power failures — including failures during pre-use checkout that could interrupt mechanical ventilation — plus no gas flow, inadequate delivery, and inaccurate output. No deaths and no injuries were reported. Two open recalls name this device: one from November 2025 about unexpected shutdowns when mains power is lost, and one from March 2025 about ineffective ventilation in volume control mode.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 70
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 00840682103954), received between Jul 18, 2019 and Apr 29, 2026 — classified as 70 malfunctions.
The most frequently reported specific device problem is “Failure to Run on Battery” (19 reports).
2 FDA recall records name this device's identifier, 2 still open.
12 other recalls cover devices cleared under the same FDA submission (K151570); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 14, 2025Open, Classified
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Z-0913-2026 · Root cause: Component change control · Recall details · FDA record ↗
- Mar 21, 2025Open, Classified
Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.
Z-1632-2025 · Root cause: Device Design · Recall details · FDA record ↗
Most reported problems
Device problems
- Failure to Run on Battery19
- No Flow14
- Therapeutic or Diagnostic Output Failure12
- Power Problem10
- Insufficient Flow or Under Infusion5
- Electrical /Electronic Property Problem2
- Failure to Power Up2
- Inaccurate Delivery2
Patient problems
- No Clinical Signs, Symptoms or Conditions66
- No Known Impact Or Consequence To Patient2
- No Patient Involvement2
Recent adverse event reports
The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 29, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A BATTERY FAILURE THAT COULD RESULT IN A LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25033222 ↗MalfunctionApr 29, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25028726 ↗MalfunctionApr 29, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25028301 ↗MalfunctionApr 28, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25019073 ↗MalfunctionApr 28, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25015075 ↗MalfunctionApr 22, 2026Failure to Run on Battery; Power Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24957319 ↗MalfunctionApr 9, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING A CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24835884 ↗MalfunctionMar 16, 2026Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24607182 ↗MalfunctionMar 11, 2026Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24567750 ↗MalfunctionFeb 16, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24342682 ↗MalfunctionFeb 11, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24319772 ↗MalfunctionFeb 11, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24313247 ↗MalfunctionFeb 11, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24313205 ↗MalfunctionFeb 6, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24276339 ↗MalfunctionJan 30, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025). THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24219853 ↗MalfunctionJan 29, 2026No Flow
AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025). THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24207186 ↗MalfunctionJan 8, 2026Power Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A POTENTIAL UNIT POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24003856 ↗MalfunctionDec 24, 2025Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING A CASE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23889145 ↗MalfunctionDec 19, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23848792 ↗MalfunctionNov 20, 2025Power Problem; Power Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN POTENTIAL POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23609297 ↗MalfunctionNov 18, 2025Unexpected Therapeutic Results; Unexpected Therapeutic Results
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23583671 ↗MalfunctionNov 13, 2025Failure to Run on Battery; Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23539997 ↗MalfunctionNov 12, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23533750 ↗MalfunctionNov 5, 2025Failure to Run on Battery; Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23476240 ↗MalfunctionOct 21, 2025Failure to Run on Battery; Failure to Run on Battery
IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23351941 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K151570. One submission often covers many models, so check the product and lot details in each record.
- Z-0915-2026 · Nov 14, 2025 · Open, ClassifiedGE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S
- Z-1634-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200. Anesthesia system.
- Z-1624-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systems
- Z-1628-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.
- Z-1631-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.
- Z-1627-2025 · Mar 21, 2025 · Open, ClassifiedGE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.
- Z-0813-2020 · Nov 25, 2019 · TerminatedCarestation 620/650/650c A1 Product Usage: anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric
- Z-2955-2018 · Jul 24, 2018 · TerminatedGE Carestation 30 anesthesia systems
- Z-2954-2018 · Jul 24, 2018 · TerminatedGE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems
- Z-2956-2018 · Jul 24, 2018 · TerminatedGE 9100c NXT anesthesia systems
- Z-0579-2018 · Oct 11, 2017 · Open, ClassifiedCarestation 620 A1, Carestation 650 A1, and Carestation 650c A1
- Z-2703-2017 · May 2, 2017 · TerminatedCarestation 620 A1, Carestation 650 A1, and Carestation 650c A1
Other Gas-machine, anesthesia devices
- Carestation 650 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls3,109 reports
- Carestation 750 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,204 reports
- Aisys CS2Datex-Ohmeda Inc · Gas-machine, anesthesia5 recalls2,048 reports
- Carestation 620 A1Datex-Ohmeda Inc · Gas-machine, anesthesia2 recalls2,015 reports
- Avance CS2Datex-Ohmeda Inc · Gas-machine, anesthesia6 recalls1,540 reports
- Flow-c Anesthesia System 6887700Maquet Critical Care AB · Gas-Machine, Anesthesia6 recalls72 reports
- FLOW-i C20 6677200Maquet Critical Care AB · Gas-Machine, Anesthesia4 recalls99 reports
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page