Carestation 650c A1Recalls & reported problems

Datex-Ohmeda Inc · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Datex-Ohmeda Carestation 650c A1 is an anesthesia gas machine used in operating rooms. The 70 reports mostly describe battery and power failures — including failures during pre-use checkout that could interrupt mechanical ventilation — plus no gas flow, inadequate delivery, and inaccurate output. No deaths and no injuries were reported. Two open recalls name this device: one from November 2025 about unexpected shutdowns when mains power is lost, and one from March 2025 about ineffective ventilation in volume control mode.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
70
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 70 adverse event reports that list this exact device (DI 00840682103954), received between Jul 18, 2019 and Apr 29, 2026 — classified as 70 malfunctions.

The most frequently reported specific device problem is “Failure to Run on Battery” (19 reports).

2 FDA recall records name this device's identifier, 2 still open.

12 other recalls cover devices cleared under the same FDA submission (K151570); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 3 reportsSep 2024: 3 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 1 reportsMar 2025: 3 reportsApr 2025: 2 reportsMay 2025: 4 reportsJun 2025: 4 reportsJul 2025: 2 reportsAug 2025: 4 reportsSep 2025: 2 reportsOct 2025: 4 reportsNov 2025: 5 reportsDec 2025: 2 reportsJan 2026: 3 reportsFeb 2026: 5 reportsMar 2026: 2 reportsApr 2026: 7 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 14, 2025Open, Classified

    Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.

    Z-0913-2026 · Root cause: Component change control · Recall details · FDA record ↗

  2. Mar 21, 2025Open, Classified

    Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

    Z-1632-2025 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Run on Battery19
  • No Flow14
  • Therapeutic or Diagnostic Output Failure12
  • Power Problem10
  • Insufficient Flow or Under Infusion5
  • Electrical /Electronic Property Problem2
  • Failure to Power Up2
  • Inaccurate Delivery2

Patient problems

  • No Clinical Signs, Symptoms or Conditions66
  • No Known Impact Or Consequence To Patient2
  • No Patient Involvement2

Recent adverse event reports

The 25 most recent of 70 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A BATTERY FAILURE THAT COULD RESULT IN A LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25033222 ↗
  • MalfunctionApr 29, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25028726 ↗
  • MalfunctionApr 29, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25028301 ↗
  • MalfunctionApr 28, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25019073 ↗
  • MalfunctionApr 28, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25015075 ↗
  • MalfunctionApr 22, 2026Failure to Run on Battery; Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24957319 ↗
  • MalfunctionApr 9, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24835884 ↗
  • MalfunctionMar 16, 2026Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24607182 ↗
  • MalfunctionMar 11, 2026Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24567750 ↗
  • MalfunctionFeb 16, 2026No Flow

    AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24342682 ↗
  • MalfunctionFeb 11, 2026No Flow

    AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24319772 ↗
  • MalfunctionFeb 11, 2026No Flow

    AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24313247 ↗
  • MalfunctionFeb 11, 2026No Flow

    AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24313205 ↗
  • MalfunctionFeb 6, 2026No Flow

    AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025) THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24276339 ↗
  • MalfunctionJan 30, 2026No Flow

    AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025). THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24219853 ↗
  • MalfunctionJan 29, 2026No Flow

    AS A RESULT OF AN INSPECTION THAT WAS COMPLETED AS PART OF THE CORRECTION INITIATED BY GE HEALTHCARE (GEHC) ON 02-APR-2025, (RECALL NUMBERS Z-1624-2025, Z-1625-2025, Z-1626-2025, Z-1627-2025, Z-1628-2025, Z-1629-2025, Z-1630-2025, Z-1631-2025, Z-1632-2025, Z-1633-2025, Z-1634-2025, Z-1635-2025, Z-1636-2025, Z-1637-2025, Z-1638-2025). THIS UNIT WAS IDENTIFIED AS HAVING AN ISSUE THAT PREVENTS EFFECTIVE MECHANICAL VENTILATION IN VOLUME CONTROL VENTILATION (VCV) MODE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24207186 ↗
  • MalfunctionJan 8, 2026Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN A POTENTIAL UNIT POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24003856 ↗
  • MalfunctionDec 24, 2025Mechanical Problem; Mechanical Problem

    IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING A CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23889145 ↗
  • MalfunctionDec 19, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT THERE WAS AN ERROR RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23848792 ↗
  • MalfunctionNov 20, 2025Power Problem; Power Problem

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN POTENTIAL POWER FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23609297 ↗
  • MalfunctionNov 18, 2025Unexpected Therapeutic Results; Unexpected Therapeutic Results

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23583671 ↗
  • MalfunctionNov 13, 2025Failure to Run on Battery; Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23539997 ↗
  • MalfunctionNov 12, 2025Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF MECHANICAL VENTILATION DURING PRE-USE CHECKOUT. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533750 ↗
  • MalfunctionNov 5, 2025Failure to Run on Battery; Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23476240 ↗
  • MalfunctionOct 21, 2025Failure to Run on Battery; Failure to Run on Battery

    IT WAS REPORTED THAT THERE WAS A BATTERY FAILURE THAT COULD CAUSE LOSS OF MECHANICAL VENTILATION. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23351941 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K151570. One submission often covers many models, so check the product and lot details in each record.

  • Z-0915-2026 · Nov 14, 2025 · Open, Classified
    GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S
  • Z-1634-2025 · Mar 21, 2025 · Open, Classified
    GE Healthcare Carestation 650c A1, Model/REF Number 1012-9655-200. Anesthesia system.
  • Z-1624-2025 · Mar 21, 2025 · Open, Classified
    GE Healthcare CARESTATION 620 A1, Model/REF Number 1012-9620-000. anesthesia systems
  • Z-1628-2025 · Mar 21, 2025 · Open, Classified
    GE Healthcare CARESTATION 650 A1, Model/REF Number 1012-9650-000. Anesthesia system.
  • Z-1631-2025 · Mar 21, 2025 · Open, Classified
    GE Healthcare Carestation 650 A1, Model/REF Number 1012-9650-200. Anesthesia system.
  • Z-1627-2025 · Mar 21, 2025 · Open, Classified
    GE Healthcare Carestation 620 A1, Model/REF Number 1012-9620-200. Anesthesia system.
  • Z-0813-2020 · Nov 25, 2019 · Terminated
    Carestation 620/650/650c A1 Product Usage: anesthesia systems are intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients (pediatric
  • Z-2955-2018 · Jul 24, 2018 · Terminated
    GE Carestation 30 anesthesia systems
  • Z-2954-2018 · Jul 24, 2018 · Terminated
    GE Carestation 620/650/650c, A1 and A2 (non-US distribution) anesthesia systems
  • Z-2956-2018 · Jul 24, 2018 · Terminated
    GE 9100c NXT anesthesia systems
  • Z-0579-2018 · Oct 11, 2017 · Open, Classified
    Carestation 620 A1, Carestation 650 A1, and Carestation 650c A1
  • Z-2703-2017 · May 2, 2017 · Terminated
    Carestation 620 A1, Carestation 650 A1, and Carestation 650c A1

Other Gas-machine, anesthesia devices

All Gas-machine, anesthesia devices →