BD Texium recallZ-1971-2025 · Bd Switzerland Sarl · initiated May 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bd Switzerland Sarl or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterile, single use closed system drug transfer devices, standalone or on infusion sets or bonded syringes, may leak at the interface between the closed male luer component and the needle-free connector, which may cause procedure delay, treatment interruptions, exposure to hazardous drugs, and/or underdosing of medication.
FDA root cause category: Process control
What the company is doing
On 5/12/2025, removal notices were mailed or emailed to Medical Director, Risk Manager, Medical Device Safety Officer, Nurse Managers, and Distributors who were asked to do the following: 1. Immediately review your inventory for affected product. Destroy all unused affected products subject to the recall following your institution s process for destruction. 2. Share this recall notification with all users within your facility network of the product to ensure they are also aware of this recall. 3. Complete the Customer Response Form via email to [email protected] 4. In addition, distributors were asked to identify all customers that purchased any affected product and to provide a copy of the customer letter to all customers and to advise them of the removal notification. Report any complaints experienced with the use of this product to the firm's North American Regional Complaint Center: Phone: 1-844-823-5433 Mon-Fri 8:00am and 5:00pm CT, Email: [email protected]
Recall details
- Recall number
- Z-1971-2025
- Status
- Open, Classified
- Initiated by the company
- May 12, 2025
- Posted by the FDA
- Jun 12, 2025
- Recalling firm
- Bd Switzerland Sarl
- Quantity
- 1,265,400
- Product code
- FMF
- FDA submissions
- K071108
- Distribution
- Worldwide - US Nationwide distribution including in the states of MA, CA, IL, AK, AL, MO, FL, VA, ME, NC, MD, IA, WI, AZ, CO, KY, IN, TX, AR, NJ, NY, MT, NE, TN, PA, GA, WA, KS, MN, ID, SC, MI, SD, OH, WV, LA, ND, VT, GU, NV, UT, OK, OR, DE, NM, MS, WY and the countries of CA, AU, NZ, CN.
Product description
BD Texium Needle-Free Syringe: 3 mL, REF: MY8003; 5 mL, REF: MY8005; 10 mL, REF: MY8010; 20 mL, REF: MY8020; 30 mL, REF: MY8030; 60 mL, REF: MY8060
Code information (models, lots, serial numbers)
Show the full code information (3k characters)
UDI-DI/Lot(Expiration): MY8003/10885403221651/23125144(December 5, 2026), 24015468(January 20, 2027), 24025611(February 5, 2027), 24026390(February 27, 2027), 24036107(March 25, 2027), 24045647(April 24, 2027), 24055859(May 27, 2027), 24095513(September 24, 2027), 92217803(June 27, 2025), 92225801(September 15, 2025), 92231303(November 9, 2025), 92300201(January 2, 2026), 92308702(March 28, 2026), 92309501(April 5, 2026), 92313601(May 16, 2026), 92315701(June 6, 2026) MY8005/10885403221668/23125145(December 5, 2026), 24015467(January 20, 2027), 24026387(February 27, 2027), 24045646(April 24, 2027), 24055858(May 27, 2027), 24075164(July 26, 2027), 24095512(September 24, 2027), 92215201(June 1, 2025), 92220802(July 27, 2025), 92225502(September 12, 2025), 92315102(May 31, 2026); MY8010/10885403221675/24015312(January 15, 2027), 24025610(February 5, 2027), 24045645(April 24, 2027), 24055857(May 27, 2027), 24066953(June 25, 2027), 24075165(July 26, 2027), 24085486(August 26, 2027), 24095511(September 24, 2027), 24105167(October 25, 2027), 92218601(July 5, 2025), 92222902(August 17, 2025), 92223601(August 24, 2025), 92229201(October 19, 2025), 92230701(November 3, 2025), 92231301(November 9, 2025), 92231302(November 9, 2025), 92307402(March 15, 2026), 92311601(April 26, 2026), 92315101(May 31, 2026); MY8020/10885403221682/23115069(November 3, 2026), 24015311(January 15, 2027), 24025935(February 7, 2027), 24035566(March 5, 2027), 24036106(March 25, 2027), 24045644(April 24, 2027), 24055856(May 27, 2027), 24066952(June 25, 2027), 24075163(July 26, 2027), 24085485(August 26, 2027), 24095083(September 9, 2027), 24105196(October 30, 2027), 92216401(June 13, 2025), 92217801(June 27, 2025), 92218001(June 29, 2025), 92219202(July 11, 2025), 92231901(November 15, 2025), 92300205(January 2, 2026), 92300902(January 9, 2026), 92309503(April 5, 2026), 92310101(April 11, 2026), 92313002(May 10, 2026), 92313602(May 16, 2026) MY8030/10885403230523/23115070(November 3, 2026), 23125143(December 5, 2026), 24015264(January 12, 2027), 24015644(January 27, 2027), 24026389(February 27, 2027), 24036105(March 25, 2027), 24045643(April 24, 2027), 24055855(May 27, 2027), 24075162(July 26, 2027), 24085484(August 26, 2027), 24095509(September 24, 2027), 24115110(November 5, 2027), 92225501(September 12, 2025), 92233602(December 2, 2025), 92307403(March 15, 2026), 92308701(March 28, 2026), 92309502(April 5, 2026), 92315702(June 6, 2026), 92316402(June 13, 2026), 24066951(June 25, 2027); MY8060/10885403221699/23115496(November 28, 2026), 23125393(December 18, 2026), 24015102(January 8, 2027), 24026388(February 27, 2027), 24036104(March 25, 2027), 24045366(April 23, 2027), 24055854(May 27, 2027), 24075166(July 26, 2027), 24095596(September 24, 2027), 24115111(November 14, 2027), 92216402(June 13, 2025), 92222002(August 8, 2025), 92222901(August 17, 2025), 92224303(August 31, 2025), 92224301(August 31, 2025), 92226502(September 22, 2025), 92228501(October 12, 2025), 92230702(November 3, 2025), 92230703(November 3, 2025), 92230704(November 3, 2025), 92300202(January 2, 2026), 92300203(January 2, 2026), 92306703(March 8, 2026), 92306701(March 8, 2026), 92306702(March 8, 2026), 92311602(April 26, 2026), 92312302(May 3, 2026), 92313001(May 10, 2026), 92316601(June 15, 2026), 92316602(June 15, 2026)
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- BD Texium MY8060BD Switzerland Sàrl · Syringe, piston1 recall94 reports
- BD Texium MY8020BD Switzerland Sàrl · Syringe, piston1 recall31 reports
- BD Texium MY8030BD Switzerland Sàrl · Syringe, piston1 recall30 reports
- BD Texium MY8010BD Switzerland Sàrl · Syringe, piston1 recall7 reports
- BD Texium MY8003BD Switzerland Sàrl · Syringe, piston1 recall5 reports
- BD Texium MY8005BD Switzerland Sàrl · Syringe, piston1 recall1 report
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 96737).
- Z-1970-2025 · Open, ClassifiedAlaris Pump Infusion Set: SmartSite Bag Access Non-Vented Bonded Texium With Priming Cap Back Check Valve 2 SmartSite Y-Sites, REF: 10013361T; SmartSite Bag Access Non-Vented Bon…
- Z-1973-2025 · Open, ClassifiedBD TEXIUM CLOSED MALE LUER WITH FEMALE CAP, REF: 10012241-0500
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page