ICU Medical Plum Duo, Item numbers: 1) 400020401, 2)… recallZ-1990-2025 · ICU Medical, Inc. · initiated Apr 22, 2025

Class II Open, Classified All recalls posted in June 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

FDA root cause category: Under Investigation by firm

What the company is doing

ICU Medical issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 05/02/2025 via hand delivery. The notice explained the issue, potential risk, and requested the following: Actions for Users: 1. Identify all affected infusion pumps in your possession and ensure all users or potential users of these pumps are immediately made aware of this notification. 2. If Maximum Dose Limit is not used for a medication, the issue will not occur. If Maximum Dose Limit is used for a medication, users should take the following precautions. If you need to adjust the weight or BSA, first clear the program and then reprogram. For further inquiries: To report adverse events or product complaints: Global Complaint Management - 1-844-654-7780 (M-F, 8:00am 5:00pm CT) or [email protected]. For any questions regarding this action: Device Correction Inquiries https://icumed.custhelp.com/app/market-action. Additional technical information or assistance: Technical Support 1-(800)-241-4002, option 3 or [email protected].

Recall details

Recall number
Z-1990-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Apr 22, 2025
Posted by the FDA
Jun 18, 2025
Recalling firm
ICU Medical, Inc.
Quantity
1701 units
Product code
FRN
FDA submissions
K223607
Distribution
Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

Product description

ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump

Code information (models, lots, serial numbers)

Show the full code information (17k characters)

1) 400020401: UDI/DI 400020401, Serial/Lot Numbers: 50000184; UDI/DI +M335400021, Serial/Lot Numbers: 50002748, 50002774, 50001647, 50001654, 50001601, 50001615, 50001651, 50002547, 50002577, 50002678, 50002689, 50002750, 50002751, 50002755, 50002773, 50002788, 50002798, 50002799, 50002801, 50002807, 50002808, 50002818, 50002819, 50002820, 50002821, 50002822, 50002823, 50002824, 50002825, 50002828, 50002832, 50002835, 50002837, 50002839, 50002840, 50002844, 50002845, 50002847, 50002851, 50002853, 50002858, 50002892, 50003000, 50003002, 50003009, 50003023, 50003026, 50003029, 50003033, 50003034, 50003035, 50003036, 50003039, 50003043, 50003045, 50003051, 50003054, 50003058, 50003059, 50003060, 50003062, 50003063, 50003065, 50003087, 50003092, 50003126, 50003143, 50003157, 50003167, 50003201, 50003232, 50003241, 50003244, 50003246, 50003257, 50003259, 50003260, 50003261, 50003262, 50003266, 50003267, 50003268, 50003269, 50003272, 50003274, 50003277, 50003278, 50003279, 50003280, 50003281, 50003282, 50003285, 50003286, 50003287, 50003288, 50003290, 50003291, 50003292, 50003293, 50003298, 50003299, 50003300, 50003301, 50003302, 50003305, 50003307, 50003308, 50003309, 50003311, 50003312, 50003316, 50003317, 50003319, 50003329, 50003331, 50003334, 50003335, 50003339, 50003340, 50003343, 50003345, 50003352, 50003360, 50003403, 50003415, 50003431, 50003433, 50003456, 50003460, 50003485, 50003492, 50003497, 50003498, 50003500, 50003514, 50003524, 50003548, 50003575, 50003603, 50003604, 50003615, 50003617, 50003689, 50003690, 50003702, 50003716, 50003736, 50003754, 50003757, 50003779, 50003826, 50003830, 50003835, 50003862, 50003867, 50003873, 50003874, 50003883, 50003937, 50003950, 50003958, 50003975, 50003982, 50003983, 50003988, 50003990, 50003992, 50003995, 50004006, 50004023, 50004074, 50004077, 50004099, 50004106, 50004121, 50004125, 50004136, 50000124, 50000125, 50000126, 50000127, 50000128, 50000129, 50000130, 50000131, 50000132, 50000133, 50000134, 50000135, 50000136, 50000137, 50000138, 50000139, 50000140, 50000141, 50000142, 50000143, 50000144, 50000145, 50000146, 50000147, 50000148, 50000149, 50000150, 50000151, 50000152, 50000153, 50000154, 50000155, 50000156, 50000157, 50000158, 50000159, 50000160, 50000161, 50000162, 50000163, 50000164, 50000165, 50000166, 50000167, 50000168, 50000169, 50000170, 50000171, 50000172, 50000173, 50000174, 50000175, 50000176, 50000177, 50000178, 50000179, 50000180, 50000181, 50000182, 50000183, 50000185, 50000186, 50000187, 50000188, 50000189, 50000190, 50000191, 50000192, 50000193, 50000194, 50000195, 50000196, 50000197, 50000198, 50000199, 50000200, 50000201, 50000202, 50000203, 50000204, 50000205, 50000206, 50000207, 50000208, 50000209, 50000210, 50000211, 50000212, 50000213, 50000214, 50000215, 50000216, 50000217, 50000218, 50000219, 50000220, 50000222, 50000223, 50000224, 50000225, 50000226, 50000227, 50000228, 50000229, 50000230, 50000231, 50000232, 50000233, 50000234, 50000235, 50000236, 50000237, 50000240, 50000264, 50000268, 50000270, 50000271, 50002115, 50002176, 50002178, 50002184, 50002211, 50002224, 50002239, 50002251, 50002280, 50002281, 50002290, 50002292, 50002293, 50002294, 50002296, 50002297, 50002298, 50002301, 50002302, 50002311, 50001567, 50001612, 50001658, 50002836, 50002873, 50002938, 50002974, 50002975, 50002992, 50002995, 50002997, 50003001, 50003006, 50003007, 50003011, 50003012, 50003016, 50003017, 50003019, 50003020, 50003021, 50003022, 50003024, 50003025, 50003027, 50003038, 50003040, 50003053, 50003064, 50003067, 50003068, 50003074, 50003078, 50002736, 50002990, 50003037, 50003103, 50003131, 50003136, 50003145, 50003146, 50003152, 50003153, 50003155, 50003156, 50003158, 50003162, 50003163, 50003165, 50003171, 50003173, 50003176, 50003177, 50003178, 50003188, 50003189, 50003190, 50003192, 50003193, 50003194, 50003195, 50003197, 50003198, 50003199, 50003203, 50003206, 50003208, 50003209, 50003214, 50003217, 50003218, 50003423, 50003440, 50003455, 50003509, 50003518, 50003534, 50003545, 50003585, 50003605, 50003625, 50003631, 50003645, 50003648, 50003796, 50002876, 50002897, 50003018, 50003028, 50003030, 50003032, 50003044, 50003048, 50003057, 50003070, 50003071, 50003076, 50003077, 50003079, 50003080, 50003083, 50003084, 50003085, 50003088, 50003093, 50003094, 50003095, 50003098, 50003100, 50003101, 50003102, 50003104, 50003105, 50003106, 50003107, 50003108, 50003109, 50003110, 50003111, 50003112, 50003114, 50003115, 50003116, 50003117, 50003118, 50003119, 50003120, 50003121, 50003122, 50003123, 50003124, 50003125, 50003127, 50003128, 50003129, 50003130, 50003132, 50003133, 50003134, 50003135, 50003137, 50003138, 50003139, 50003140, 50003141, 50003142, 50003144, 50003148, 50003149, 50003151, 50003154, 50003159, 50003161, 50003166, 50003168, 50003170, 50003172, 50003175, 50003179, 50003181, 50003182, 50003183, 50003184, 50003185, 50003186, 50003187, 50003191, 50003196, 50003202, 50003205, 50003207, 50003211, 50003212, 50003215, 50003216, 50003219, 50003220, 50003221, 50003222, 50003223, 50003224, 50003225, 50003226, 50003227, 50003228, 50003229, 50003230, 50003231, 50003233, 50003234, 50003235, 50003236, 50003238, 50003239, 50003240, 50003242, 50003243, 50003245, 50003247, 50003248, 50003249, 50003250, 50003251, 50003252, 50003253, 50003254, 50003255, 50003256, 50003258, 50003263, 50003264, 50003265, 50003270, 50003271, 50003273, 50003276, 50003283, 50003284, 50003295, 50003296, 50003297, 50003303, 50003304, 50003306, 50003310, 50003313, 50003314, 50003315, 50003318, 50003320, 50003321, 50003322, 50003323, 50003324, 50003325, 50003326, 50003328, 50003330, 50003332, 50003333, 50003336, 50003337, 50003342, 50003346, 50003349, 50003350, 50003351, 50003353, 50003354, 50003355, 50003356, 50003357, 50003359, 50003361, 50003365, 50003366, 50003367, 50003368, 50003369, 50003370, 50003371, 50003372, 50003373, 50003374, 50003375, 50003377, 50003378, 50003379, 50003380, 50003381, 50003383, 50003385, 50003386, 50003387, 50003388, 50003389, 50003390, 50003392, 50003394, 50003395, 50003396, 50003397, 50003398, 50003399, 50003401, 50003402, 50003404, 50003405, 50003406, 50003407, 50003409, 50003410, 50003411, 50003412, 50003413, 50003414, 50003418, 50003419, 50003420, 50003421, 50003422, 50003427, 50003429, 50003430, 50003435, 50003436, 50003441, 50003449, 50003465, 50003467, 50003473, 50003475, 50003476, 50003477, 50003480, 50003482, 50003484, 50003486, 50003489, 50003490, 50003493, 50003496, 50003499, 50003504, 50003506, 50003507, 50003510, 50003511, 50003516, 50003517, 50003520, 50003527, 50003531, 50003541, 50003553, 50003554, 50003556, 50003559, 50003565, 50003569, 50003574, 50003576, 50003588, 50003591, 50003593, 50003596, 50003597, 50003609, 50003614, 50003619, 50003621, 50003622, 50003626, 50003635, 50003640, 50003642, 50003646, 50003647, 50003649, 50003651, 50003652, 50003653, 50003662, 50003663, 50003667, 50003668, 50003670, 50003672, 50003674, 50003675, 50003678, 50003679, 50003681, 50003683, 50003685, 50003687, 50003692, 50003695, 50003696, 50003698, 50003699, 50003706, 50003712, 50003713, 50003718, 50003719, 50003722, 50003723, 50003726, 50003727, 50003728, 50003734, 50003737, 50003738, 50003740, 50003741, 50003742, 50003745, 50003748, 50003749, 50003751, 50003752, 50003756, 50003761, 50003762, 50003764, 50003765, 50003767, 50003768, 50003775, 50003776, 50003785, 50003786, 50003787, 50003789, 50003793, 50003794, 50003797, 50003798, 50003801, 50003804, 50003807, 50003812, 50003816, 50003817, 50003818, 50003819, 50003820, 50003822, 50003823, 50003824, 50003829, 50003831, 50003832, 50003833, 50003834, 50003836, 50003838, 50003839, 50003840, 50003842, 50003847, 50003848, 50003849, 50003852, 50003853, 50003854, 50003855, 50003856, 50003857, 50003860, 50003863, 50003864, 50003866, 50003868, 50003869, 50003871, 50003875, 50003877, 50003878, 50003880, 50003884, 50003885, 50003886, 50003887, 50003888, 50003889, 50003891, 50003892, 50003893, 50003894, 50003895, 50003896, 50003897, 50003898, 50003899, 50003900, 50003901, 50003902, 50003903, 50003905, 50003906, 50003908, 50003911, 50003912, 50003913, 50003914, 50003916, 50003917, 50003918, 50003919, 50003920, 50003922, 50003923, 50003925, 50003926, 50003927, 50003929, 50003931, 50003934, 50003936, 50003938, 50003940, 50003942, 50003944, 50003945, 50003946, 50003947, 50003948, 50003951, 50003952, 50003953, 50003954, 50003955, 50003956, 50003957, 50003959, 50003960, 50003961, 50003962, 50003963, 50003964, 50003965, 50003966, 50003968, 50003970, 50003971, 50003972, 50003976, 50003977, 50003981, 50003984, 50003985, 50003986, 50003987, 50003989, 50003991, 50003993, 50003994, 50003996, 50003999, 50004001, 50004002, 50004003, 50004004, 50004005, 50004007, 50004008, 50004009, 50004010, 50004011, 50004012, 50004013, 50004014, 50004017, 50004018, 50004019, 50004021, 50004022, 50004024, 50004026, 50004028, 50004029, 50004030, 50004031, 50004032, 50004033, 50004034, 50004035, 50004036, 50004037, 50004038, 50004039, 50004041, 50004042, 50004043, 50004044, 50004045, 50004047, 50004049, 50004050, 50004051, 50004052, 50004053, 50004054, 50004055, 50004056, 50004057, 50004058, 50004059, 50004061, 50004062, 50004063, 50004064, 50004065, 50004067, 50004068, 50004069, 50004071, 50004073, 50004075, 50004076, 50004080, 50004081, 50004082, 50004084, 50004085, 50004086, 50004087, 50004088, 50004089, 50004090, 50004091, 50004092, 50004093, 50004095, 50004096, 50004097, 50004098, 50004101, 50004102, 50004103, 50004104, 50004109, 50004110, 50004112, 50004113, 50004114, 50004115, 50004116, 50004117, 50004118, 50004119, 50004122, 50004123, 50004126, 50004127, 50004129, 50004130, 50004131, 50004134, 50004135, 50004137, 50004138, 50004139, 50004140, 50004142, 50004143, 50004145, 50004147, 50004151, 50004152, 50004153, 50004155, 50004159, 50004162, 50004164, 50004165, 50004169, 50004170, 50004171, 50004172, 50004173, 50004174, 50004175, 50004176, 50004180, 50004186, 50004187, 50004190, 50004192, 50004195, 50004196, 50004199, 50004215, 50004221, 50004231, 50004233, 50004238, 50004239, 50001392, 50001415, 50001427, 50001439, 50001440, 50001448, 50001458, 50001461, 50001474, 50001482, 50001501, 50001505, 50001510, 50001536, 50001544, 50001558, 50001559, 50001562, 50001563, 50001565, 50001574, 50001576, 50001593, 50001600, 50001631, 50002514, 50002575, 50002623, 50002629, 50002646, 50002677, 50002692, 50002722, 50002726, 50002735, 50002742, 50002764, 50002817, 50003004, 50000294, 50000295, 50000296, 50000298, 50000299, 50000303, 50000304, 50000309, 50001694, 50001714, 50001723, 50001739, 50001788, 50001795, 50001812, 50001813, 50001815, 50001818, 50001819, 50001831, 50001840, 50001842, 50001849, 50001859, 50001866, 50001869, 50001871, 50001874, 50001875, 50001876, 50001878, 50001879, 50001884, 50001885, 50001889, 50001892, 50001894, 50001895, 50001899, 50001900, 50001902, 50001903, 50001904, 50001905, 50001906, 50001907, 50001911, 50001920, 50001923, 50001924, 50001925, 50001928, 50001929, 50001930, 50001934, 50001935, 50001938, 50001942, 50001945, 50001950, 50001955, 50001961, 50001963, 50001965, 50001968, 50001971, 50001975, 50001977, 50001979, 50001987, 50001992, 50001996, 50002004, 50002027, 50002034, 50002047, 50002052, 50002071, 50002085, 50002087, 50002091, 50002107, 50002114, 50002120, 50002123, 50002126, 50002128, 50002143, 50002144, 50002148, 50002149, 50002157, 50002166, 50002174, 50002175, 50002177, 50002179, 50002183, 50002186, 50002195, 50002196, 50002198, 50002201, 50002207, 50002209, 50002214, 50002220, 50002222, 50002225, 50002226, 50002227, 50002230, 50002234, 50002236, 50002243, 50002247, 50002249, 50002253, 50002256, 50002258, 50002259, 50002260, 50002262, 50002271, 50002273, 50002295, 50002314, 50002339, 50002351, 50002355, 50002365, 50002375, 50002393, 50002406, 50002415, 50002424, 50002428, 50002431, 50002435, 50002436, 50002438, 50002440, 50002445, 50002448, 50002451, 50002454, 50002458, 50002460, 50002462, 50002463, 50002464, 50002465, 50002466, 50002467, 50002470, 50002472, 50002475, 50002476, 50002477, 50002478, 50002480, 50002482, 50002483, 50002488, 50002489, 50002490, 50002493, 50002494, 50002496, 50002497, 50002498, 50002499, 50002500, 50002501, 50002505, 50002506, 50002507, 50002508, 50002509, 50002510, 50002511, 50002512, 50002513, 50002515, 50002516, 50002517, 50002520, 50002522, 50002523, 50002524, 50002527, 50002528, 50002529, 50002531, 50002532, 50002533, 50002535, 50002536, 50002538, 50002539, 50002540, 50002542, 50002543, 50002545, 50002548, 50002550, 50002551, 50002552, 50002553, 50002554, 50002557, 50002559, 50002560, 50002564, 50002565, 50002566, 50002567, 50002568, 50002570, 50002571, 50002572, 50002573, 50002578, 50002580, 50002583, 50002584, 50002585, 50002586, 50002587, 50002588, 50002589, 50002591, 50002592, 50002594, 50002596, 50002597, 50002600, 50002602, 50002606, 50002608, 50002610, 50002611, 50002613, 50002616, 50002617, 50002621, 50002622, 50002624, 50002628, 50002633, 50002638, 50002640, 50002642, 50002643, 50002645, 50002647, 50002648, 50002653, 50002654, 50002655, 50002656, 50002658, 50002659, 50002660, 50002662, 50002664, 50002665, 50002666, 50002667, 50002668, 50002671, 50002685, 50002686, 50002694, 50002697, 50002699, 50002700, 50002701, 50002703, 50002704, 50002705, 50002706, 50002709, 50002713, 50002714, 50002715, 50002718, 50002719, 50002721, 50002723, 50002724, 50002727, 50002728, 50002729, 50002730, 50002732, 50002733, 50002738, 50002740, 50002741, 50002747, 50002749, 50002752, 50002754, 50002759, 50002760, 50002763, 50002766, 50002767, 50002768, 50002769, 50002771, 50002772, 50002775, 50002776, 50002777, 50002778, 50002779, 50002782, 50002783, 50002785, 50002786, 50002787, 50002789, 50002790, 50002791, 50002793, 50002794, 50002795, 50002796, 50002797, 50002800, 50002802, 50002806, 50002812, 50002813, 50002815, 50002826, 50002829, 50002830, 50002833, 50002838, 50002841, 50002842, 50002849, 50002850, 50002852, 50002854, 50002856, 50002857, 50002859, 50002860, 50002862, 50002863, 50002865, 50002866, 50002867, 50002868, 50002869, 50002870, 50002871, 50002872, 50002875, 50002877, 50002879, 50002880, 50002881, 50002882, 50002883, 50002884, 50002885, 50002886, 50002887, 50002888, 50002890, 50002891, 50002893, 50002894, 50002895, 50002898, 50002899, 50002900, 50002901, 50002902, 50002903, 50002904, 50002905, 50002906, 50002907, 50002908, 50002909, 50002910, 50002911, 50002913, 50002914, 50002915, 50002916, 50002917, 50002918, 50002919, 50002921, 50002923, 50002924, 50002925, 50002926, 50002927, 50002928, 50002929, 50002930, 50002932, 50002933, 50002934, 50002935, 50002936, 50002937, 50002939, 50002940, 50002942, 50002944, 50002947, 50002948, 50002949, 50002950, 50002951, 50002956, 50002958, 50002960, 50002961, 50002962, 50002963, 50002964, 50002965, 50002966, 50002967, 50002968, 50002969, 50002970, 50002971, 50002972, 50002976, 50002977, 50002978, 50002979, 50002980, 50002981, 50002982, 50002986, 50002987, 50002988, 50002989, 50002991, 50002993, 50002994, 50002996, 50002998, 50003003, 50003005, 50003008, 50003010, 50003013, 50003014, 50003073, 50003200; 2) 400021001: UDI/DI 400021001, Serial/Lot Numbers: 50000070; UDI/DI +M335400021, Serial/Lot Numbers: 50000004, 50000005, 50000006, 50000007, 50000008, 50000009, 50000010, 50000011, 50000012, 50000013, 50000014, 50000015, 50000016, 50000017, 50000018, 50000019, 50000020, 50000021, 50000022, 50000023, 50000024, 50000025, 50000026, 50000027, 50000028, 50000029, 50000030, 50000031, 50000032, 50000033, 50000034, 50000035, 50000036, 50000037, 50000038, 50000039, 50000040, 50000041, 50000042, 50000043, 50000044, 50000045, 50000046, 50000047, 50000048, 50000049, 50000050, 50000051, 50000052, 50000053, 50000054, 50000055, 50000056, 50000057, 50000058, 50000059, 50000060, 50000061, 50000062, 50000063, 50000064, 50000065, 50000066, 50000067, 50000068, 50000069, 50000071, 50000072, 50000073, 50000074, 50000075, 50000076, 50000077, 50000078, 50000079, 50000080, 50000081, 50000082, 50000083, 50000084, 50000086, 50000087, 50000088, 50000089, 50000090, 50000091, 50000092, 50000093, 50000094, 50000095, 50000096, 50000097, 50000098, 50000099, 50000100, 50000101, 50000102, 50000103, 50000104, 50000105, 50000106, 50000107, 50000108, 50000109, 50000110, 50000111, 50000112, 50000113, 50000114, 50000115, 50000116, 50000117; 3) 400021003: UDI/DI +M335400021, Serial/Lot Numbers: 80000052, 80000053, 80000054, 80000055, 80000056, 80000057, 80000058, 80000059, 80000060, 80000061, 80000062, 80000063, 80000064, 80000065, 80000066, 80000067, 80000068, 80000069, 80000071, 80000072, 80000073, 80000074, 80000075, 80000076, 80000078, 80000079, 80000080, 80000081, 80000082, 80000083, 80000084, 80000085, 80000086, 80000087, 80000089, 80000090, 80000091, 80000092, 80000093, 80000094, 80000095, 80000096, 80000097, 80000098, 80000099, 80000100, 80000101, 80000102, 80000103, 80000104, 80000024, 80000027, 80000028, 80000030, 80000031, 80000032, 80000034, 80000035, 80000037, 80000038, 80000039, 80000041, 80000042, 80000043, 80000044, 80000045, 80000046, 80000047, 80000048, 80000049, 80000051, 80000001, 80000002, 80000003, 80000006, 80000007, 80000008, 80000009, 80000010, 80000011, 80000012, 80000013, 80000014, 80000015, 80000017, 80000021, 80000022, 80000023, 80000025, 80000026.

Other devices under the same FDA submission

1 device cleared under K223607 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.