Skip to content

Plum Duo Infusion System 1.1Recalls & reported problems

Icu Medical, Inc. · Plum Duo Dual Channel Infusion System · Class 2

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M335400021), received between Dec 5, 2024 and Oct 28, 2025 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Device Markings/Labelling Problem” (2 reports).

6 other recalls cover devices cleared under the same FDA submission (K223607); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Markings/Labelling Problem2
  • Inaccurate Delivery1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2025Inaccurate Delivery; Inaccurate Delivery

    IT WAS REPORTED, VIA EMAIL, THAT A PLUM DUO, ALLEGING THAT AFTER THEY PROGRAMMED A BOLUS AMOUNT, THE PUMP CONTINUED TO BOLUS AT THE BOLUS RATE. THEY WERE TOLD THAT OUR DRUG LIBRARY DID NOT INCLUDE SPECIFICS ABOUT WHAT THE PUMP SHOULD DO WHEN THE BOLUS INFUSION WAS COMPLETED, SO IT CONTINUES AT THAT RATE. THEY SET UP A CHANGE TO THE LIBRARY TO FIX THIS FOR OBSTETRIC (OB) RELATED DRUGS, BUT THE UPDATE COULD NOT BE PUSHED OUT DUE TO SERVER ISSUES WITH ICU MEDICAL. THERE WAS NO PATIENT HARM REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23407629 ↗
  • MalfunctionDec 6, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    THE EVENT INVOLVED A PLUM DUO INFUSION SYSTEM WHERE IT WAS REPORTED THAT THE DEVICE SERIAL NUMBER SHOW ON SCREEN DOES NOT MATCH SERIAL NUMBER ON BARCODE LABEL DURING IMPLEMENTATION/CONFIGURATION OF DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20860692 ↗
  • MalfunctionDec 5, 2024Device Markings/Labelling Problem; Device Markings/Labelling Problem

    THE EVENT INVOLVED A PLUM DUO INFUSION SYSTEM WHERE IT WAS REPORTED THAT THE DEVICE SERIAL NUMBER SHOW ON SCREEN DOES NOT MATCH SERIAL NUMBER ON BARCODE LABEL DURING IMPLEMENTATION/CONFIGURATION OF DEVICE. THERE WAS NO PATIENT INVOLVEMENT AND NO PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20858850 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K223607. One submission often covers many models, so check the product and lot details in each record.

  • Z-2723-2026 · Jun 10, 2026 · Open, Classified
    ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump
  • Z-2312-2026 · Apr 29, 2026 · Open, Classified
    Plum Duo Infusion Pump, 40002-0401
  • Z-1118-2026 · Dec 15, 2025 · Open, Classified
    ICU Medical Plum Duo Infusion Pump, List Numbers: 400020401 (US), 400021001 (Philippines)
  • Z-1094-2026 · Dec 15, 2025 · Open, Classified
    Plum Duo Infusion System, List Number 40002-04-01
  • Z-2129-2025 · Jul 11, 2025 · Open, Classified
    Plum Duo Infusion System, List Number: 400020401
  • Z-1990-2025 · Apr 22, 2025 · Open, Classified
    ICU Medical Plum Duo, Item numbers: 1) 400020401, 2) 400021001 , and 3) 400021003; infusion pump

Other Pump, infusion devices

All Pump, infusion devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.