Skip to content

BD Insyte Autoguard recallZ-2005-2025 · Becton Dickinson Infusion Therapy Systems, Inc. · initiated May 21, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Becton Dickinson Infusion Therapy Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to customer complaints, there is the possibility that catheter needle may be slow to retract or fail to retract.

FDA root cause category: Component design/selection

What the company is doing

On May 21, 2025, Becton Dickinson Infusion Therapy (BD) issued a "Urgent: Medical Device Recall" Notification to affected consignees FedEx and E-Mail. BD asked consignees to take the following actions: 1. Immediately review your inventory for product listed in the above table. Cease use and destroy any unused affected devices. 2. For devices that have deployed and are currently in use or if an impacted device was previously used, no additional follow-up activities are required for patients that have completed therapy. 3. Share and post this Product Recall letter within your facility network and forward to any customers you may have distributed the product to ensure awareness. Ensure the contents of this Product Recall are read and understood by those within your organization. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification.

Recall details

Recall number
Z-2005-2025
Status
Open, Classified
Initiated by the company
May 21, 2025
Posted by the FDA
Jun 23, 2025
Recalling firm
Becton Dickinson Infusion Therapy Systems, Inc.
Quantity
50 units
Product code
FOZ
FDA submissions
K201075
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Belgium, Canada, Chile, Colombia, Guatemala, Japan, New Zealand, Panama, Philippines, Singapore, Vietnam.

Product description

REF: 382544 BD Insyte Autoguard Shielded IV Catheter, 18 GA x 1.16 in (1.3 x 30 mm) 95 mL/min, STERILEEO, RxOnly

Code information (models, lots, serial numbers)

Lot# 4312025 UDI: 00382903825448

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

53 devices cleared under K201075 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96865).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page