Skip to content

Ivenix, Inc. recallZ-1869-2025 · Fresenius Kabi USA, LLC · initiated May 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Fresenius Kabi USA, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.

FDA root cause category: Under Investigation by firm

What the company is doing

Consignees received an "URGENT - VOLUNTARY RECALL" notification via email dated 5/11/25. This notification instructs consignees to discontinue use and distribution of affected devices immediately. Consignees are to check their inventory for affected devices and place them in quarantine until product can be returned for replacement. The provided recall notification is to be distributed to users of the affected devices. Consignees are instructed to return the completed Customer Reply Form via fax or email to facilitate return of affected devices for replacement. Consignees with any questions can contact Fresenius Kabi by phone at 855-354-6387x1 or by email at [email protected].

Recall details

Recall number
Z-1869-2025
Status
Open, Classified
Initiated by the company
May 12, 2025
Posted by the FDA
Jun 6, 2025
Recalling firm
Fresenius Kabi USA, LLC
Quantity
14,280 units (724 cases)
Product code
FPA
FDA submissions
K183311
Distribution
US distribution: CA, CO, DE, GA, IL, MD, MI, MN, MS, NC, NJ, NV, OK, SC, TX, VA, WA & WI.

Product description

IVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.

Code information (models, lots, serial numbers)

Model No. SET-0014-20; UDI 20811505030034; Lot No. FA24K05015.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

4 devices cleared under K183311 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page