SafeStar 55 Plus Filter. Bidirectionally breathing system… recallZ-2159-2025 · Draeger, Inc. · initiated Jun 16, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
FDA root cause category: Process control
What the company is doing
An initial Urgent Medical Device Recall notification dated June 2025 was mailed to consignees. The notification instructs consignees to not make therapeutic decisions based solely on individual measured values or parameters, regardless of the measuring point and the ventilator or anesthesia machine used. Consignees are to not connect the sample line to the filter sampling port; if an alternative sample port is not feasible, the firm requests that consignees use an alternative filter. All users of affected devices are to be notified of this recall notice and the notice forwarded to customers to which product was further distributed. Consignees are asked to complete and return the provided Response Acknowledgement and Receipt Form. Questions can be directed to Michael Kelhart at 267-664-1131 from 8:00 AM EST to 4:30 PM EST, or by email at [email protected]. A second Urgent Medical Device Recall notification dated July 2025 was mailed to consignees. This notification recommends that any unused filters affected by this recall not be used and any unused stock should be removed. Consignees are to contact their local Draeger Representative for information regarding alternatives. Draeger Customer Success can be contacted at 1-800-437-2437 (option 2, option 1) or [email protected] to coordinate return/replacement. This recall notification should be shared with all users and forwarded to where devices were further distributed.
Recall details
- Recall number
- Z-2159-2025
- Status
- Open, Classified
- Initiated by the company
- Jun 16, 2025
- Posted by the FDA
- Jul 31, 2025
- Recalling firm
- Draeger, Inc.
- Quantity
- 9,885,139 units
- Product code
- CAH
- FDA submissions
- K221836
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AZ, CA, CO, CT, FL, ID, IL, IN, KAS, LA, MA, ME, MI, MN, MO, NC, NE, NM, NV, NY, OH, PA, SD, TN, TX, UT, WA, WI, WY and the countries of Angola, Argentina, Australia, Austria, Belarus, Belgium, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Burkina-Faso, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dominican Rep., El Salvador, Faeroe, France, French Guiana, Gabon, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Libya, Liechtenstein, Malaysia, Mali, Mexico, Moldavia, Mongolia, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad, Tobago, Turkey, Ukraine, United Kingdom, United Arab Emirates, Uzbekistan, Vietnam.
Product description
SafeStar 55 Plus Filter. Bidirectionally breathing system filter.
Code information (models, lots, serial numbers)
Mode No. MP05790; UDI: 04048675558736; All Lots.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Filter SafeStar 55 Plus MP05790Drägerwerk AG & Co. KGaA · Filter, Bacterial, Breathing-Circuit1 recall2 reports
Other devices under the same FDA submission
3 devices cleared under K221836 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Filter SafeStar 90 Plus MP05785Drägerwerk AG & Co. KGaA · Filter, Bacterial, Breathing-Circuit1 recall0 reports
- Filter SafeStar 60A Plus MP05795Drägerwerk AG & Co. KGaA · Filter, Bacterial, Breathing-Circuit1 recall0 reports
- Filter/HME TwinStar HEPA Plus MP05801Drägerwerk AG & Co. KGaA · Filter, Bacterial, Breathing-Circuit1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97026).
- Z-2158-2025 · Open, ClassifiedSafeStar 90 Plus Filter. Bidirectionally breathing system filter.
- Z-2160-2025 · Open, ClassifiedHME TwinStar HEPA Plus Filter. Bidirectionally breathing system filter.
- Z-2161-2025 · Open, ClassifiedSafeStar 60A Plus Filter. Bidirectionally breathing system filter.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page