Filter SafeStar 55 Plus MP05790Recalls & reported problems
Drägerwerk AG & Co. KGaA · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048675558736), received between Mar 12, 2025 and May 26, 2025 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Decreased Sensitivity” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
3 other recalls cover devices cleared under the same FDA submission (K221836); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Jun 16, 2025Class IOpen, Classified
Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.
Z-2159-2025 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Decreased Sensitivity1
- Imprecision1
- Mechanical Problem1
- Obstruction of Flow1
Patient problems
- Bronchospasm1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 26, 2025Imprecision; Decreased Sensitivity; Imprecision; Decreased Sensitivity
IT WAS REPORTED THAT THE ETCO2-CURVE WAS DISPLAYED FLATTENED. DURING FURTHER INTERNAL REVIEW, IT WAS FOUND THAT THE EVENT WAS TO BE ASSESSED REPORTABLE BECAUSE THE CHANGED CURVE COULD LEAD TO MISINTERPRETATION BY THE USER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22084501 ↗InjuryMar 12, 2025Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow
THE FOLLOWING WAS REPORTED ¿SUDDEN EMERGENCY SITUATION: CLINICAL SITUATION APPEARED AS A MASSIVE BRONCHOSPASM DUE TO AN ANAPHYLACTIC REACTION TO THE ANTIBIOTIC CEFUROXIME, WITH VENTILATION PRESSURES OF 40 MBAR ONLY A VT OF 20 - 120 ML, LATER PARTLY UP TO 220 ML COULD BE ACHIEVED, THERE WAS AN IMMEDIATE VITAL THREAT!!! CORRECT TUBE POSITION (LARYNGOSCOPIC AND BROCHOSCOPIC POSITION CHECK, EXCLUSION OF A CUFF HERNIA, NO INDICATION OF ASPIRATION, MEDICATION WAS GIVEN ACCORDING TO THE GUIDELINES, AFTER SWITCHING TO THE EVITA 4 INTENSIVE CARE VENTILATOR (NEW PATIENT FILTER!) THE PATIENT COULD BE VENTILATED IMMEDIATELY WITH NORMAL PRESSURES, THE CHECK OF THE ANESTHESIA MACHINE REVEALED NO DEFECTS, BUT THE ANESTHESIA FILTER USED (VISUALLY UNREMARKABLE) IS ONLY PASSABLE FOR AIR AT HIGH PRESSURE.¿ ACCORDING TO ADDITIONAL INFORMATION RECEIVED, THE OPERATION WAS CANCELED DUE TO THE SYMPTOMS DESCRIBED ABOVE.
Patient problems: Bronchospasm
MAUDE report 21580868 ↗
FDA clearance and classification
Cleared through 510(k) K221836 on Dec 7, 2022 (decision: Substantially Equivalent). Submitted by Drägerwerk AG & Co. KGaA as “Filter CareStar Plus, Filter SafeStar Plus, Filter/HME TwinStar Plus”. The FDA review took 167 days. FDA record ↗
The FDA regulates product code CAH (filter, bacterial, breathing-circuit) under 21 CFR 868.5260 as a Class II device. 21 CFR 868.5260 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K221836. One submission often covers many models, so check the product and lot details in each record.
- Z-2161-2025 · Jun 16, 2025 · Open, ClassifiedSafeStar 60A Plus Filter. Bidirectionally breathing system filter.
- Z-2160-2025 · Jun 16, 2025 · Open, ClassifiedHME TwinStar HEPA Plus Filter. Bidirectionally breathing system filter.
- Z-2158-2025 · Jun 16, 2025 · Open, ClassifiedSafeStar 90 Plus Filter. Bidirectionally breathing system filter.
Common questions about the Filter SafeStar 55 Plus MP05790
Has the Filter SafeStar 55 Plus MP05790 been recalled?
Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-2159-2025 (Class I), began on Jun 16, 2025: Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters. If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Filter SafeStar 55 Plus MP05790?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Mar 12, 2025 and May 26, 2025. In those reports, the most common specific device problems are “Decreased Sensitivity” (1), “Imprecision” (1) and “Mechanical Problem” (1); the most common patient problem is “Bronchospasm” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Filter SafeStar 55 Plus MP05790?
None of the reports are classified as deaths, 1 as an injury and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Filter SafeStar 55 Plus MP05790 FDA approved?
It was cleared through 510(k) K221836 on Dec 7, 2022. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the Filter SafeStar 55 Plus MP05790 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the Filter SafeStar 55 Plus MP05790, and is it still sold?
The FDA's device database lists Drägerwerk AG & Co. KGaA as the company responsible for it (brand Filter SafeStar 55 Plus, model MP05790). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Filter, Bacterial, Breathing-Circuit devices
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls570 reports
- VOCSN V+Pro Emergency, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall557 reports
- VOCSN V+O+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit4 recalls371 reports
- VOCSN V+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit3 recalls2 deaths330 reports
- VOCSN V+C+S+N+Pro, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall185 reports
- VOCSN VOCSN, EnglishVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit1 recall1 death97 reports
- VOCSN V+C+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls34 reports
- VOCSN V+O+C+S+N+Pro, JapaneseVentec Life Systems, Inc. · Filter, Bacterial, Breathing-Circuit2 recalls32 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page