Skip to content

Filter SafeStar 55 Plus MP05790Recalls & reported problems

Drägerwerk AG & Co. KGaA · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 04048675558736), received between Mar 12, 2025 and May 26, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Decreased Sensitivity” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

3 other recalls cover devices cleared under the same FDA submission (K221836); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 16, 2025Open, Classified

    Reports of slow increase of CO2-curve when gas sample is taken from the gas sampling port of filters.

    Z-2159-2025 · Root cause: Process control · FDA record ↗

Most reported problems

Device problems

  • Decreased Sensitivity1
  • Imprecision1
  • Mechanical Problem1
  • Obstruction of Flow1

Patient problems

  • Bronchospasm1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2025Imprecision; Decreased Sensitivity; Imprecision; Decreased Sensitivity

    IT WAS REPORTED THAT THE ETCO2-CURVE WAS DISPLAYED FLATTENED. DURING FURTHER INTERNAL REVIEW, IT WAS FOUND THAT THE EVENT WAS TO BE ASSESSED REPORTABLE BECAUSE THE CHANGED CURVE COULD LEAD TO MISINTERPRETATION BY THE USER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22084501 ↗
  • InjuryMar 12, 2025Mechanical Problem; Obstruction of Flow; Mechanical Problem; Obstruction of Flow

    THE FOLLOWING WAS REPORTED ¿SUDDEN EMERGENCY SITUATION: CLINICAL SITUATION APPEARED AS A MASSIVE BRONCHOSPASM DUE TO AN ANAPHYLACTIC REACTION TO THE ANTIBIOTIC CEFUROXIME, WITH VENTILATION PRESSURES OF 40 MBAR ONLY A VT OF 20 - 120 ML, LATER PARTLY UP TO 220 ML COULD BE ACHIEVED, THERE WAS AN IMMEDIATE VITAL THREAT!!! CORRECT TUBE POSITION (LARYNGOSCOPIC AND BROCHOSCOPIC POSITION CHECK, EXCLUSION OF A CUFF HERNIA, NO INDICATION OF ASPIRATION, MEDICATION WAS GIVEN ACCORDING TO THE GUIDELINES, AFTER SWITCHING TO THE EVITA 4 INTENSIVE CARE VENTILATOR (NEW PATIENT FILTER!) THE PATIENT COULD BE VENTILATED IMMEDIATELY WITH NORMAL PRESSURES, THE CHECK OF THE ANESTHESIA MACHINE REVEALED NO DEFECTS, BUT THE ANESTHESIA FILTER USED (VISUALLY UNREMARKABLE) IS ONLY PASSABLE FOR AIR AT HIGH PRESSURE.¿ ACCORDING TO ADDITIONAL INFORMATION RECEIVED, THE OPERATION WAS CANCELED DUE TO THE SYMPTOMS DESCRIBED ABOVE.

    Patient problems: Bronchospasm

    MAUDE report 21580868 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K221836. One submission often covers many models, so check the product and lot details in each record.

  • Z-2158-2025 · Jun 16, 2025 · Open, Classified
    SafeStar 90 Plus Filter. Bidirectionally breathing system filter.
  • Z-2160-2025 · Jun 16, 2025 · Open, Classified
    HME TwinStar HEPA Plus Filter. Bidirectionally breathing system filter.
  • Z-2161-2025 · Jun 16, 2025 · Open, Classified
    SafeStar 60A Plus Filter. Bidirectionally breathing system filter.

Other Filter, Bacterial, Breathing-Circuit devices

All Filter, Bacterial, Breathing-Circuit devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.