Skip to content

Signa Pet/Mr recallZ-2133-2025 · GE Medical Systems, LLC · initiated May 23, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

FDA root cause category: Device Design

What the company is doing

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 05/23/2025 delivered using traceable means. The notice explained the issue, risk, and requested the following: Actions to be taken by Customer/User: You can continue to use your MR system by following the instructions below: Ensure hearing protection with a Noise Reduction Rating (NRR) of at least 33dB is used during exams on these systems. 200 pairs are provided with this letter. Please post this letter in your facility on or near the MR operator console. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected].

Recall details

Recall number
Z-2133-2025
Status
Open, Classified
Initiated by the company
May 23, 2025
Posted by the FDA
Jul 18, 2025
Recalling firm
GE Medical Systems, LLC
Product code
LNH
Distribution
Worldwide distribution - US Nationwide and the country of China.

Product description

SIGNA PET/MR

Code information (models, lots, serial numbers)

GTIN: 00840682105378

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97030).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page