Brand Name: NIO-A recallZ-2118-2025 · Waismed Ltd. · initiated Jun 5, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Waismed Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.
FDA root cause category: Process change control
What the company is doing
On June 05, 2025, WaisMed issued an "Urgent: Medical Device Recall" notification to affected consignees via E-Mail. WaisMed asked consignees to take the follow actions: 1. Healthcare providers should immediately discontinue use of affected devices and consider alternative intraosseous access products to ensure continuity of patient care without compromising treatment timelines in critical situations. 2. Immediately discontinue, quarantine and remove all NIO A and NIO+ devices from the affected lots from all points of use, including emergency kits, crash carts, ambulances, and clinical storage areas. 3. Verify your inventory against the specific lot numbers listed in this recall notice to identify all affected products. 4. Share and post this Product Recall letter within your facility network. 5. Forward to any customers you may have distributed the product to ensure awareness. 6. Ensure the contents of this Product Recall are read and understood by those within your organization. 7. Please complete the attached Recall Acknowledgement and Receipt Form and return it to: [email protected] or mail to: Waismed Ltd. 10 Amal St. Afek Industrial Park, Rosh Ha'yin 4809234, Israel. Any questions directly associated with this recall should be directed to the: Recall Coordinator or International Director Quality and Regulatory Affairs via email: [email protected], phone: +972-9-9517444 or website: safeguardmedial.com.
Recall details
- Recall number
- Z-2118-2025
- Status
- Open, Classified
- Initiated by the company
- Jun 5, 2025
- Posted by the FDA
- Jul 17, 2025
- Recalling firm
- Waismed Ltd.
- Quantity
- 7056 units (US)
- Product code
- FMI
- FDA submissions
- K142086
- Distribution
- Worldwide distribution: US (nationwide) to state of: NC and OUS (international) countries of: Slovenia, Australia, Brazil, Estonia, South Africa, U. Arab Emirate, Greece, Panama, Singapore, Italy, Germany, Portugal, Spain, Ukraine, Lithuania, Chile, Poland, Germany, New Zealand, Montenegro, Switzerland, Argentina, Poland, Austria, Sweden, United Kingdom, Italy, France, Romania, Netherlands, Croatia, Lithuania, New Zealand, Japan, Cyprus, and Israel.
Product description
Brand Name: NIO-A Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000603 Software Version: N/A Product Description: NIO devices are a spring-based, automatic intra-osseous access device supplied Sterile and single use. Indicated for Intra-osseous access to the Proximal humerus and Proximal Tibia in adult patients, in emergent situations. Component: N/A
Code information (models, lots, serial numbers)
UDI: 0 7290008325059 Lot number: 2440043, 2440045, 2540046, 2540047
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- NIO NIO-AWaisMed Ltd · Needle, Hypodermic, Single Lumen1 recall2 deaths8 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97061).
- Z-2119-2025 · Open, ClassifiedBrand Name: NIO+ Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000908 Software Version: N/A Product Description: NIO devices are a spring-based, automatic in…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page