Skip to content

NIO NIO-ARecalls & reported problems

WaisMed Ltd · The NIO-A device is indicated to provide intraosseous access into the proximal tibia and proximal humeral head, in adult patients older than 12 years. · Class 2

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
8
Deaths reported
2
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 07290008325059), received between Sep 28, 2018 and Apr 2, 2025 — classified as 2 deaths, 6 malfunctions.

The most frequently reported specific device problem is “Firing Problem” (3 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jun 5, 2025Open, Classified

    Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.

    Z-2118-2025 · Root cause: Process change control · FDA record ↗

Most reported problems

Device problems

  • Firing Problem3
  • Mechanical Problem2
  • Adverse Event Without Identified Device or Use Problem1
  • Component Missing1
  • Failure to Fire1

Patient problems

  • Insufficient Information6
  • No Clinical Signs, Symptoms or Conditions1
  • Swelling1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • DeathApr 2, 2025Mechanical Problem; Mechanical Problem

    PATIENT WAS IN CARDIAC ARREST WITH CPR BEING PERFORMED. CREW ARRIVED AND TOOK OVER CPR. IV WAS ATTEMPTED AND FAILED. NIO WAS BROUGHT OUT, SITE LOCATED AND WHEN WE SPUN THE TOP ON THE NIO TO UNLOCK IT, IT WAS TIGHT TO TURN. PRESSURE PLACED WITH DOMINANT HAND, OTHER HAND STABILIZING NIO. NIO DID NOT DEPLOY, SECOND CREWMEMBER ATTEMPTED TO DEPLOY IT AND IT DID NOT WORK. SECOND NIO WAS BROUGHT OUT AND THE SAME THING HAPPENED. TOP WAS TIGHT TO SPIN AND IT DID NOT DEPLOY FOR EITHER THE RN OR AEMT. MANUAL IO WAS PLACED IN THIS PATIENT FIRST ATTEMPT. CPR CONTINUED. PATIENT DIED. ONE OF THE NIO THAT DID NOT DEPLOY WAS BROUGHT BACK TO THE STATION AND IS CURRENTLY BEING SENT BACK WITH THE PROVIDED SHIPPING LABEL.

    Patient problems: Insufficient Information

    MAUDE report 21756448 ↗
  • MalfunctionApr 2, 2025Firing Problem; Firing Problem

    COMPLAINT DESCRIPTION FROM CUSTOMER ON MARCH 30 2025: "PATIENT WAS IN CARDIAC ARREST WITH CPR BEING PERFORMED. CREW ARRIVED, TOOK OVER CPR. IV WAS ATTEMPTED AND FAILED. NIO WAS BROUGHT OUT, SITE LOCATED AND WHEN WE SPUN THE TOP ON THE NIO TO UNLOCK IT, IT WAS TIGHT TO TURN. PRESSURE PLACED WITH DOMINANT HAND, OTHER HAND STABILIZING NIO. NIO DID NOT DEPLOY, SECOND CREWMEMBER ATTEMPTED TO DEPLOY IT AND IT DID NOT WORK. SECOND NIO WAS BROUGHT OUT AND THE SAME THING HAPPENED. TOP WAS TIGHT TO SPIN AND IT DID NOT DEPLOY FOR EITHER THE RN OR EMT. MANUAL IO WAS PLACED IN THIS PATIENT FIRST ATTEMPT. CPR CONTINUED. PATIENT DIED. ONE OF THE NIO THAT DID NOT DEPLOY WAS BROUGHT BACK TO THE STATION AND IS CURRENTLY BEING SENT BACK WITH THE PROVIDED SHIPPING LABEL.

    Patient problems: Insufficient Information

    MAUDE report 21750879 ↗
  • DeathApr 2, 2025Firing Problem; Firing Problem

    COMPLAINT DESCRIPTION FROM CUSTOMER ON (B)(6) 2025: "PATIENT WAS IN CARDIAC ARREST WITH CPR BEING PERFORMED. CREW ARRIVED, TOOK OVER CPR. IV WAS ATTEMPTED AND FAILED. NIO WAS BROUGHT OUT, SITE LOCATED AND WHEN WE SPUN THE TOP ON THE NIO TO UNLOCK IT, IT WAS TIGHT TO TURN. PRESSURE PLACED WITH DOMINANT HAND, OTHER HAND STABILIZING NIO. NIO DID NOT DEPLOY, SECOND CREWMEMBER ATTEMPTED TO DEPLOY IT AND IT DID NOT WORK. SECOND NIO WAS BROUGHT OUT AND THE SAME THING HAPPENED. TOP WAS TIGHT TO SPIN AND IT DID NOT DEPLOY FOR EITHER THE RN OR EMT. MANUAL IO WAS PLACED IN THIS PATIENT FIRST ATTEMPT. CPR CONTINUED. PATIENT DIED. ONE OF THE NIO THAT DID NOT DEPLOY WAS BROUGHT BACK TO THE STATION AND IS CURRENTLY BEING SENT BACK WITH THE PROVIDED SHIPPING LABEL.

    Patient problems: Insufficient Information

    MAUDE report 21750369 ↗
  • MalfunctionNov 10, 2024Firing Problem; Firing Problem

    COMPLAINT DESCRPTION FROM CUSTOMER ON OCTOBER 15, 2024: "WE RECENTLY HAD AN NIO ADULT DEVICE MALFUNCTION DURING A CODE EVENT. WE STOCK THESE IN OUR CODE CARTS AND WHEN MY TEAM MEMBER ATTEMPTED TO INSERT THE DEVICE, FOLLOWING THE IFU, IT DEPLOYED, WITH DELAY, AND ACTUALLY FELL APART. FOLLOWS IS THE COMMUNICATION I RECEIVED FROM MY COLLEAGUE: I WENT TO PLACE AN NIO DURING A CODE AND IT WAS VERY DIFFICULT TO DEPLOY THE NIO. I AM COMFORTABLE PLACING THE NIO AND USING THE NIO DEVICE. THE DEVICE I USED WAS FROM THE CRASH CART AND THERE ARE NO EXTRAS IN THE CRASH CART WHEN SOMETHING LIKE THIS OCCURS. HERE WAS THE SEQUENCE: I HAD UNLOCKED IT, UNCAPPED IT, WAS STABILIZING IT AND DID THE SQUEEZE AND FINALLY, AFTER WHAT SEEMED LIKE AN ETERNITY IN A BAD SITUATION, IT DID DEPLOY BUT AT THE SAME TIME, THE ENTIRE NIO APPARATUS EXPLODED. LITERALLY. I DON'T KNOW WHAT TO SAY. I DID SAVE IT FOR YOU IF YOU WANT TO TAKE A LOOK, BUT THAT IS SERIOUSLY WHAT HAPPENED. I CAME BACK TO THE OFFICE AFTER I WAS ORGANIZED AND GRABBED ANOTHER NIO JUST TO SEE IF IT WAS ME AND NOPE. IT DEPLOYED EASY JUST LIKE IT SHOULD. NO ISSUES AND NO EXPLOSION. NO HARM WAS DONE TO THE PATIENT OTHER THAN A DELAY IN VASCULAR ACCESS

    Patient problems: Insufficient Information

    MAUDE report 20648712 ↗
  • MalfunctionJun 26, 2024Mechanical Problem; Mechanical Problem

    DESCRIPTION PROVIDED BY THE CUSTOMER: "PATIENT WAS S/P CARDIAC ARREST. CODE TEAM MEMBER ARRIVED AND WAS ASKED TO PLACE AN IO IN THE PATIENT'S LEG. PATIENT WAS PREPPED AND SCRUBBED IN PROXIMAL LEFT TIBIA. IO END CAP WAS REMOVED, AND STYLET WAS TURNED TO ACTIVATE THE NEEDLE INTO POSITION. UPON PLACING PRESSURE INTO LEG AND PULLING TRIGGER, IO FAILED TO DEPLOY FULLY INTO THE BONE. CANNULA WAS REMOVED. SECOND ATTEMPT WAS MADE IN THE PROXIMAL RIGHT TIBIA IN SIMILAR FASHION. SECOND IO WAS PULLED FROM CODE TEAM BAG AND WAS A SEPARATE BUT SIMILAR MODEL. UPON DEPLOYING INTO THE BONE, INNER CANNULA WAS UNABLE TO BE REMOVED FROM THE OUTER IO CANNULA AND THUS WAS UNABLE TO BE USED APPROPRIATELY. AS PATIENT HAD CENTRAL LINE ACCESS AT THIS TIME, ATTEMPTS TO PLACE IO WERE HALTED AND WOUNDS DRESSED WITH TEGADERM."

    Patient problems: Insufficient Information

    MAUDE report 19611823 ↗
  • MalfunctionOct 18, 2023Failure to Fire; Failure to Fire

    COMPLAINT DESCRIPTION FROM CUSTOMER ON (B)(6) 2023. "VAT NURSE OPENED PACKAGE, UNLOCKED DEVICE BY TWISTING TOP, SQUEEZED DOWN AND THE NEEDLE DID NOT DEPLOY. SECOND NURSE STEPPED IN AND ATTEMPTED TO DEPLOY NEEDLE AND GOT A NEGATIVE RESPONSE. THEY OPENED ANOTHER UNIT, FOLLOWED THE SAME STEPS, AND SUCCESSFULLY DEPLOYED NEEDLE INTO PATIENT. THERE WAS A ONE-TO-TWO-MINUTE DELAY BETWEEN START AND DEPLOYMENT BECAUSE OF THIS INCIDENT. BOTH NURSES HAVE BEEN TRAINED ON THE DEVICE AND HAVE SUCCESSFULLY USED IT IN THE PAST".

    Patient problems: Insufficient Information

    MAUDE report 17955292 ↗
  • MalfunctionJun 1, 2022Component Missing; Component Missing

    THE FOLLOWING INFORMATION WAS PROVIDED BY THE USER ON (B)(6) 2022: DESCRIBE THE EVENT OR PROBLEM: IO (INTRAOSSEOUS NEEDLE) WAS PLACED TO DELIVER LIFESAVING MEDICATIONS. THERE WAS NO CATHETER WITH THE NEEDLE. THIS CAUSED A DELAY WITH PATIENT GETTING MEDICATIONS. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? TO DELIVER PRESSORS AND RSI MEDICATIONS. WHAT PROBLEM DID THE USER HAVE (CHECK ALL THAT APPLY): DEVICE MALFUNCTION - THAT IS, THE DEVICE DID NOT DO WHAT IT WAS SUPPOSED TO DO.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14565969 ↗
  • MalfunctionSep 28, 2018Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE NIO DEPLOYED IN THE LEFT TIBIA OF THE PATIENT. AFTER FLUSHING, IT WAS REPORTED THAT BONE MARROW WAS PRESENT AND NEEDLE PLACEMENT IN THE MARROW CAVITY WAS CONFIRMED. SEVERAL MINUTES AFTER DEPLOYMENT, PATIENT'S LEG APPEARED SWOLLEN IN COMPARISON TO HIS RIGHT LEG. THE PATIENT BECAME RESPONSIVE TO TREATMENTS AFTER ACCESS VIA DIFFERENT IO NEEDLE WAS PLACED.

    Patient problems: Swelling

    MAUDE report 7917513 ↗

Other Needle, Hypodermic, Single Lumen devices

All Needle, Hypodermic, Single Lumen devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.