INSPIRA recallZ-2329-2025 · Pentax of America Inc · initiated Jul 16, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Pentax of America Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.

FDA root cause category: Device Design

What the company is doing

On 7/16/2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. Customer Instructions: PENTAX Medical will conduct the software update on the affected devices at your facility. Your local PENTAX Medical representative will contact you to schedule the required updates for your equipment. " Please download the Instructions for Use (IFU) for the EPK-i8020c Video Processor from the PENTAX Medical online IFU library at https://ifu.pentaxmedical.com. " Please complete the enclosed Field Correction Response Form upon receipt of this package, and email to PENTAX Medical at [email protected]. Contact Information: Please indicate through the attached response form that you have received and understood this information, by completing it and returning it no later than July18, 2025 at: [email protected] If you have any questions regarding this action, please feel free to contact us at: " Tel: 1-800-431-5880 (8:30 AM 5:00 PM, Monday Friday, EST) " Fax: (800)-579-5432) " Email: [email protected]

Recall details

Recall number
Z-2329-2025
Status
Open, Classified
Initiated by the company
Jul 16, 2025
Posted by the FDA
Aug 15, 2025
Recalling firm
Pentax of America Inc
Quantity
144 units
Product code
PEA
FDA submissions
K231249, K251127
Distribution
US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.

Product description

Pentax Medical Video Processor; Model Number: EPK-i8020c;

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

Model Number: EPK-i8020c; UDI-DI: 04961333247974; Serial Numbers: B0023Z1657, D0023Z0245, D0023Z0206, D0023Z0125, D0023Z0123, D0023Z0119, D0023Z0118, D0023Z0117, D0023Z0114, D0023Z0112, D0023Z0109, D0023Z0108, D0023Z0011, C0023Z0087, C0023Z0085, C0023Z0074, C0023Z0072, C0023Z0021, B0023Z1712, B0023Z1711, B0023Z1670, B0023Z1665, B0023Z1659, B0023Z1653, B0023Z1628, b0023z1501, B0023Z1195, B0023Z1707, C0023Z0084, B0023Z1732, B0023Z1669, D0023Z0137, D0023Z0141, D0158Z0049, D0158Z0033, E0023Z0022, D0158Z0053, D0158Z0019, E0023Z0026, E0023Z0008, D0023Z0248, D0023Z0241, D0023Z0222, D0023Z0155, B0023Z1276, C0023Z0082, D0023Z0233, B0023Z1461, D0023Z0139, D0023Z0140, D0023Z0122, B0023Z1709, E0023Z0020, E0023Z0009, E0023Z0002, D0023Z0142, E0023Z0024, E0023Z0018, E0023Z0013, E0023Z0004, C0023Z0095, C0023Z0078, E0023Z0007, E0158Z0057, E0158Z0035, B0023Z1735, B0023Z1717, B0023Z1648, B0023Z1639, B0023Z1631, C0158Z0047, C0158Z0042, B0023Z1481, B0023Z1478, B0023Z1451, B0023Z1335, B0023Z1193, B0023Z1288, B0023Z1258, B0023Z1296, B0023Z1290, B0023Z1286, B0023Z1280, B0023Z1279, B0023Z1256, B0023Z1250, B0023Z1245, B0023Z1676, B0023Z1573, B0023Z1565, B0023Z1656, B0023Z1694, D0158Z0055, B0023Z1577, B0023Z1570, B0023Z1682, B0023Z1678, B0023Z1661, B0023Z1637, D0158Z0048, D0158Z0047, E0023Z0016, E0023Z0001, D0023Z0071, D0023Z0150, D0023Z0148, D0023Z0143, B0023Z1277, D0023Z0152, D0023Z0144, D0023Z0106, D0023Z0102, B0023Z1667, B0023Z1652, B0023Z1645, B0023Z1727, B0023Z1723, B0023Z1702, E0023Z0003, D0023Z0244, D0023Z0232, C0023Z0091, D0023Z0208, B0023Z1655, D0023Z0219, D0023Z0017, D0023Z0008, C0023Z0098, C0023Z0097, C0023Z0096, C0023Z0094, C0023Z0092, B0023Z1716, B0023Z1698, B0023Z1697, B0023Z1696, B0023Z1671, B0023Z1255, B0023Z1247, B0023Z1291, B0023Z1289, B0023Z1287, B0023Z1285, B0023Z1243;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K231249, K251127 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page