interventional fluoroscopic x-ray system recallZ-2020-2025 · Siemens Medical Solutions USA, Inc · initiated May 15, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

In very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose.

FDA root cause category: Radiation Control for Health and Safety Act

What the company is doing

Siemens Healthineers sent customers a Customer Advisory Notice dated May 15, 2025 informing customers of a potential issue with ARTIS pheno and ARTIS icono systems with the following model numbers: ARTIS pheno 10849000, ARTIS icono biplane 11327600, ARTIS icono ceiling 11328100, ARTIS icono floor 11327700; due to in very rare situations, communication between the sensor measuring dose applied to the area and the corresponding software fails. As a result of the failure, the system may incorrectly report an application of a high dose and appropriate software update will be implemented, based on the current software version of the affected system.

Recall details

Recall number
Z-2020-2025
Status
Open, Classified
Initiated by the company
May 15, 2025
Posted by the FDA
Jun 27, 2025
Recalling firm
Siemens Medical Solutions USA, Inc
Product code
OWB
FDA submissions
K241572
Distribution
Worldwide.

Product description

interventional fluoroscopic x-ray system

Code information (models, lots, serial numbers)

4056869063317

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K241572 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97135).