Skip to content

High Pressure Tubing recallZ-2293-2025 · Merit Medical Systems, Inc. · initiated Jun 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.

FDA root cause category: Process control

What the company is doing

On 5/5/2025, recall notices were emailed to customers who were asked to do the following: 1) Identify affected devices and quarantine, discontinue use and distribution. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4. Return all affected devices to the firm. 5. Complete, and return the customer response form via email to [email protected] If you have questions, contact the firm's customer service at [email protected] or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri

Recall details

Recall number
Z-2293-2025
Status
Open, Classified
Initiated by the company
Jun 5, 2025
Posted by the FDA
Aug 12, 2025
Recalling firm
Merit Medical Systems, Inc.
Quantity
8040
Product code
FPK
FDA submissions
K883718
Distribution
US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.

Product description

High Pressure Tubing: REF: 502211002, HP9720E/C, HP9721E/C

Code information (models, lots, serial numbers)

REF/UDI-DI/Lot(Expiration): 502211002/I3116954(8/16/2028); HP9720E/C/00884450859889/I3116059(12/10/2027), I3127321(12/30/2027), I3137220(1/28/2028); HP9721E/C/00884450859896/I3116060(12/10/2027)

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 97151).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page