Custom Waste Management Kit recallZ-2296-2025 · Merit Medical Systems, Inc. · initiated Jun 5, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Merit Medical Systems, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.
FDA root cause category: Process control
What the company is doing
On 5/5/2025, recall notices were emailed to customers who were asked to do the following: 1) Identify affected devices and quarantine, discontinue use and distribution. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them and note the quantity distributed on the customer response form. 4. Return all affected devices to the firm. 5. Complete, and return the customer response form via email to [email protected] If you have questions, contact the firm's customer service at [email protected] or via phone at +1 800 356 3748, Hours: 6 am to 6 pm MST, Mon-Fri
Recall details
- Recall number
- Z-2296-2025
- Status
- Open, Classified
- Initiated by the company
- Jun 5, 2025
- Posted by the FDA
- Aug 12, 2025
- Recalling firm
- Merit Medical Systems, Inc.
- Quantity
- 858
- Product code
- PPD
- Distribution
- US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.
Product description
Custom Waste Management Kit: REF: K10-00353K, K10-01603C, K10-01694Q, K10-05159B, K10-05743, K10-05825AP
Code information (models, lots, serial numbers)
REF/UDI-DI/Lot(Expiration): K10-00353K/00884450179246/H3127803(1/4/2028), H3166061(2/24/2028), H3172299(3/14/2028); K10-01603C/00884450593639/H3163984(3/2/2028), H3176777(3/23/2028); K10-01694Q/00884450239308/H3116890(10/31/2027), H3152493(2/3/2028), H3177622(3/14/2028); K10-05159B/00884450153598/H3116922(12/23/2027), H3119571(1/4/2028), H3127907(1/11/2028), H3152490(2/10/2028), H3152491(2/10/2028), H3152492(2/10/2028); K10-05743/00884450456767/H3122012(1/6/2028), H3128429(1/13/2028), H3177446(3/16/2028); K10-05825AP/00884450808108/H3112585(11/27/2027)
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Merit Medical® 00884450153598Merit Medical Systems, Inc. · Universal drainage tray1 recall0 reports
- Merit Medical® 00884450179246Merit Medical Systems, Inc. · Universal drainage tray1 recall0 reports
- Merit Medical® 00884450239308Merit Medical Systems, Inc. · Universal drainage tray1 recall0 reports
- Merit Medical® 00884450456767Merit Medical Systems, Inc. · Universal drainage tray1 recall0 reports
- Merit Medical® 00884450593639Merit Medical Systems, Inc. · Universal drainage tray1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97151).
- Z-2293-2025 · Open, ClassifiedHigh Pressure Tubing: REF: 502211002, HP9720E/C, HP9721E/C
- Z-2294-2025 · Open, ClassifiedCustom Tubing Kit: REF: K04-00417, K04-00963, K04-01029
- Z-2295-2025 · Open, ClassifiedCustom Manifold Kit: REF: K09-02239C, K09-02249UP, K09-09110B, K09-10294A, K09-11227A, K09-11456B, K09-12123B, K09-13342AP, K09-13391BP, K09-13548, K09-14000A, K09T-12020D, K09T-12…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page