STA-Liquid Anti-Xa 8. Model/Catalog Number: 00322US recallZ-2475-2025 · Diagnostica Stago, Inc. · initiated Jul 21, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
FDA root cause category: Mixed-up of materials/components
What the company is doing
Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed.
Recall details
- Recall number
- Z-2475-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 21, 2025
- Posted by the FDA
- Aug 28, 2025
- Recalling firm
- Diagnostica Stago, Inc.
- Quantity
- 126,200 units (125,900 US)
- Product code
- KFF
- FDA submissions
- K111822
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay, Peru, Republique Tcheque, Romania, Saudi Arabia, Serbia, South Africa, Switzerland, Taiwan, Thailand, Tunisia, Turquie, United Arab Emirates, Uruguay, Vietnam.
Product description
STA-Liquid Anti-Xa 8. Model/Catalog Number: 00322US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative determination of the plasma levels of unfractionated (UFH) and low molecular weight (LMWH) heparins by measuring their anti-Xa activity on antithrombin in a competitive assay using a synthetic chromogenic substrate.
Code information (models, lots, serial numbers)
Model/Catalog Number: 00322US. UDI-DI: 03607450011791
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- STA® - Liquid Anti-Xa 8 00322USDiagnostica Stago · ASSAY, HEPARIN1 recall0 reports
Other devices under the same FDA submission
1 device cleared under K111822 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- STA® - Liquid Anti-Xa 4 00311USDiagnostica Stago · ASSAY, HEPARIN1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97287).
- Z-2473-2025 · Open, ClassifiedSTA Satellite. Product Code: All references. All software versions. The STA Satellite is a fully automated benchtop analyzer capable of simultaneously performing clotting, chromog…
- Z-2474-2025 · Open, ClassifiedSTA-Liquid Anti-Xa 4. Model/Catalog Number: 00311US. The STA - Liquid Anti-Xa kits are intended for use with STA-R, STA Compact and STA Satellite analyzers, for the quantitative d…
- Z-2476-2025 · Open, ClassifiedSTA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the qu…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page