STA - Fibrinogen 5. Model/Catalog Number: 00674 recallZ-2476-2025 · Diagnostica Stago, Inc. · initiated Jul 21, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.

FDA root cause category: Mixed-up of materials/components

What the company is doing

Diagnostica Stago notified consignees on about 07/21/2025 via emailed letter titled URGENT: MEDICAL DEVICE RECALL (CORRECTION). Consignees were instructed to temporarily systematically run a quality control just before a batch of UFH or LMWH, if you run STA-LIQUID ANTI-Xa in UFH and LMWH applications on your STA SATELLITE analyzer. This request is being asked while Stago works to develop a permanent solution for the analyzer. Consignees were also requested to complete and return the Acknowledgement Form and to notify customers if affected units were transferred or further distributed.

Recall details

Recall number
Z-2476-2025
Status
Open, Classified
Initiated by the company
Jul 21, 2025
Posted by the FDA
Aug 28, 2025
Recalling firm
Diagnostica Stago, Inc.
Quantity
829,800 units (245,200 US)
Product code
KQJ
FDA submissions
K840211
Distribution
Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran, Israel, Jordan, Korea, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Netherlands, Oman, Paraguay, Peru, Republique Tcheque, Romania, Saudi Arabia, Serbia, South Africa, Switzerland, Taiwan, Thailand, Tunisia, Turquie, United Arab Emirates, Uruguay, Vietnam.

Product description

STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.

Code information (models, lots, serial numbers)

Model/Catalog Number: 00674. UDI-DI: 13607450006749

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97287).

Sources and updates