C8-4v Transducer Probe recallZ-2358-2025 · Philips Ultrasound, Inc · initiated Jul 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Ultrasound, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
FDA root cause category: Device Design
What the company is doing
An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377.
Recall details
- Recall number
- Z-2358-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 3, 2025
- Posted by the FDA
- Aug 22, 2025
- Recalling firm
- Philips Ultrasound, Inc
- Quantity
- 171,322 units
- Product code
- IYN
- FDA submissions
- K030455, K034003, K043535, K081661
- Distribution
- US Nationwide distribution.
Product description
C8-4v Transducer Probe
Code information (models, lots, serial numbers)
Show the full code information (26k characters)
Model No. 989803002685, 989605416131, 989605408153, 989605430232, 989605416141, 989605364943, 989605388111, 989605387261, 989605408151, 989605361973, 989605408141, 989605430242, 989605364941; UDI: Unique Device Identifier (UDI), (01)00884838067745(21)B1YRNV, (01)00884838067981(21)B35BZF, (01)00884838067981(21)B24PBY, (01)00884838067981(21)B27T4Z, (01)00884838067981(21)B35H6L, (01)00884838067981(21)B1RPRM, (01)00884838068032(21)B2DGF2, (01)00884838067981(21)B1MZ1D, (01)00884838067981(21)B1X19N, (01)00884838067981(21)B1MYCZ, (01)00884838067981(21)B1RKT8, (01)00884838067981(21)B3PYXD, (01)00884838067981(21)B2TL3C, (01)00884838067981(21)B20NKJ, (01)00884838068032(21)B1X2X3, (01)00884838068032(21)B2TN85, (01)00884838068032(21)B2TTB8, (01)00884838068032(21)B2ZHBM, (01)00884838068032(21)B2ZHDD, (01)00884838067981(21)B2TKV5, (01)00884838067981(21)B40GX1, (01)00884838068032(21)B1RRJL, (01)00884838067981(21)B3ZCR0, (01)00884838067981(21)F09JPD, (01)00884838067981(21)B20G80, (01)00884838067981(21)B2PKDF, (01)00884838067981(21)B2TL8Y, (01)00884838067981(21)B20NHL, (01)00884838064577(21)B1YRVW, (01)00884838067981(21)B1RQDW, (01)00884838067981(21)B24FWM, (01)00884838067981(21)B2PKT2, (01)00884838067981(21)B35P3D, (01)00884838067981(21)F08W6H, (01)00884838067981(21)B35C6P, (01)00884838067981(21)B1MXNG, (01)00884838067981(21)B20HV9, (01)00884838064577(21)B1F3XP, (01)00884838064577(21)B1F3RL, (01)00884838067981(21)B1MXCF, (01)00884838067981(21)B1RPML, (01)00884838067981(21)B39MMX, (01)00884838067981(21)B39RBC, (01)00884838067981(21)B1RPJP, (01)00884838067981(21)B27FQF, (01)00884838067981(21)B2TL3B, (01)00884838067981(21)B20J83, (01)00884838067981(21)B3LDL6, (01)00884838067981(21)B35P3H, (01)00884838067981(21)B2PKDH, (01)00884838067981(21)B2D91Y, (01)00884838067981(21)B1N2Z7, (01)00884838067981(21)B2TP2K, (01)00884838067981(21)B24P7D, (01)00884838067981(21)B24P7F, (01)00884838067981(21)B2PVXV, (01)00884838067981(21)B2H8BR, (01)00884838067981(21)B39T34, (01)00884838067981(21)B3HMHL, (01)00884838067981(21)B1JYQQ, (01)00884838067981(21)B1X1LF, (01)00884838067981(21)B3FGP7, (01)00884838067981(21)B20J4G, (01)00884838067981(21)B2D91Z, (01)00884838067981(21)B35BZJ, (01)00884838067981(21)B1MZW1, (01)00884838067981(21)B2D6LW, (01)00884838067981(21)B24NLZ, (01)00884838067981(21)B2PNJB, (01)00884838067981(21)B2ZQDK, (01)00884838067981(21)B24M85, (01)00884838067981(21)B2H07F, (01)00884838067981(21)B35P3C, (01)00884838067981(21)B2TTFV, (01)00884838067981(21)B3HM3G, (01)00884838067981(21)B27G10, (01)00884838067981(21)B2TTFX, (01)00884838067981(21)B35HLV, (01)00884838067981(21)B3FG7D, (01)00884838067981(21)B3FGP8, (01)00884838067981(21)B3HM3J, (01)00884838067981(21)B3LD7N, (01)00884838067981(21)B3LDL3, (01)00884838067981(21)B3ZCM0, (01)00884838067981(21)B2PKGD, (01)00884838067981(21)B1MYB6, (01)00884838067981(21)B1RQ18, (01)00884838067981(21)B39RBD, (01)00884838068032(21)B2TP71, (01)00884838067981(21)B1N2GT, (01)00884838067981(21)B20P5V, (01)00884838067981(21)B24NZN, (01)00884838067981(21)B1N2PB, (01)00884838067981(21)B3FGGM, (01)00884838067981(21)B3HLRQ, (01)00884838067981(21)B3HMHT, (01)00884838067981(21)B2D6M2, (01)00884838067981(21)B3FGGT, (01)00884838067981(21)B2ZPCX, (01)00884838067981(21)B20J4P, (01)00884838067981(21)B2H02V, (01)00884838067981(21)B2H052, (01)00884838067981(21)B2H3B2, (01)00884838064577(21)B1F3WX, (01)00884838067981(21)B1MZ1H, (01)00884838068032(21)B2TN88, (01)00884838068032(21)B32T3N, (01)00884838067981(21)B2D923, (01)00884838067981(21)B35C6R, (01)00884838067981(21)B3FHHK, (01)00884838067981(21)B20PRK, (01)00884838067981(21)B35DBL, (01)00884838067981(21)B2ZH7P, (01)00884838067981(21)B1MYYK, (01)00884838067981(21)B1X1LC, (01)00884838067981(21)B24HZ1, (01)00884838067981(21)B3LFN4, (01)00884838068032(21)B2ZPYN, (01)00884838067981(21)B1X1LB, (01)00884838067981(21)B2TNT0, (01)00884838067981(21)B39RB9, (01)00884838067981(21)B3L442, (01)00884838067981(21)B1MXCD, (01)00884838067981(21)B1MZQG, (01)00884838068032(21)B1RRJM, (01)00884838067981(21)B20HV6, (01)00884838068032(21)B1RM25, (01)00884838068032(21)B35D8Z, (01)00884838067981(21)B24P7P, (01)00884838067981(21)B2PP8B, (01)00884838067981(21)B27PKV, (01)00884838067981(21)B3TFFT, (01)00884838067981(21)B27PKW, (01)00884838067981(21)B2TKV2, (01)00884838067981(21)B2TP2M, (01)00884838067981(21)B27P92, (01)00884838067981(21)B3ZCR8, (01)00884838067981(21)B2H7TD, (01)00884838067981(21)B2ZQDT, (01)00884838067981(21)B2PWGJ, (01)00884838067981(21)B2PWWZ, (01)00884838067981(21)B2ZPCT, (01)00884838067981(21)B1JZKQ, (01)00884838067981(21)B20PQD, (01)00884838067981(21)B2ZHDC, (01)00884838067981(21)B2H07H, (01)00884838067981(21)B2PWX1, (01)00884838068032(21)F07DCT, (01)00884838067981(21)B1MYPP, (01)00884838067981(21)B1JZH5, (01)00884838067981(21)B1JZZG, (01)00884838067981(21)B1N2P8, (01)00884838067981(21)B1RQF1, (01)00884838067981(21)B1WTD3, (01)00884838067981(21)B2TMFY, (01)00884838067981(21)B35DKD, (01)00884838067981(21)B35HLZ, (01)00884838067981(21)B2TTG1, (01)00884838067981(21)B3HXWF, (01)00884838067981(21)B2ZQDQ, (01)00884838067981(21)B2PVXL, (01)00884838067981(21)B1MXR6, (01)00884838067981(21)B1MXRC, (01)00884838067981(21)B2D9CW, (01)00884838067981(21)B32MV8, (01)00884838067981(21)B3FT1R, (01)00884838067981(21)F07L4R, (01)00884838067981(21)F09CVW, (01)00884838068032(21)B1X31N, (01)00884838068032(21)B24M5T, (01)00884838067981(21)B3ZCR1, (01)00884838067981(21)B2PNJD, (01)00884838067981(21)B1MYB1, (01)00884838067981(21)B24NZB, (01)00884838067981(21)B1JYNL, (01)00884838068032(21)B2TNFG, (01)00884838067981(21)B39T37, (01)00884838067981(21)B3FHHM, (01)00884838067981(21)B20G9L, (01)00884838067981(21)B2PW0Q, (01)00884838067981(21)B2PWGG, (01)00884838067981(21)B2PWWX, (01)00884838067981(21)B2PWX3, (01)00884838067981(21)B2TL91, (01)00884838067981(21)B32N43, (01)00884838068032(21)B24HTL, (01)00884838068032(21)B32T3L, (01)00884838068032(21)B39M3C, (01)00884838068032(21)B3FGZQ, (01)00884838068032(21)B3HWG3, (01)00884838068032(21)B3HWG9, (01)00884838067981(21)B2H4TF, (01)00884838067981(21)B20G9J, (01)00884838068032(21)B1MY6N, (01)00884838067981(21)B2TNK2, (01)00884838068032(21)B24NG3, (01)00884838067981(21)B20P1X, (01)00884838068032(21)B2ZHH4, (01)00884838068032(21)B35JQJ, (01)00884838068032(21)B39J3K, (01)00884838067981(21)B2TL90, (01)00884838068032(21)B2PR5B, (01)00884838067981(21)B20J9L, (01)00884838067981(21)B20HGT, (01)00884838067981(21)B20NKL, (01)00884838067981(21)B2TL8W, (01)00884838067981(21)B3HM3L, (01)00884838064577(21)B1F3N1, (01)00884838064577(21)B1F3N8, (01)00884838067981(21)B1X19V, (01)00884838067981(21)B1X2YV, (01)00884838067981(21)B27GFD, (01)00884838067981(21)B35DBH, (01)00884838067981(21)B3HXWJ, (01)00884838067981(21)B20HQH, (01)00884838067981(21)B2TNT4, (01)00884838067981(21)B1MZBP, (01)00884838067981(21)B24FWT, (01)00884838067981(21)B27FKK, (01)00884838067981(21)B32M94, (01)00884838067981(21)B39RBH, (01)00884838067981(21)B3HY3G, (01)00884838067981(21)B2PNJ9, (01)00884838067981(21)B24P8W, (01)00884838067981(21)B2D8ZF, (01)00884838067981(21)B1WTCY, (01)00884838067981(21)B20J2H, (01)00884838068032(21)B2H0BX, (01)00884838067981(21)B32MYV, (01)00884838067981(21)B1N2GL, (01)00884838067981(21)B35C6H, (01)00884838067981(21)B32M99, (01)00884838067981(21)B32MYX, (01)00884838067981(21)B2H077, (01)00884838067981(21)B3FG77, (01)00884838067745(21)B2YV0R, (01)00884838067981(21)B3HM3H, (01)00884838067981(21)B3L440, (01)00884838067981(21)B20J9K, (01)00884838067981(21)B3FHHX, (01)00884838067981(21)B3HLRM, (01)00884838067981(21)B1X2YD, (01)00884838067981(21)B20P4T, (01)00884838067981(21)B2H57H, (01)00884838067981(21)B3FGPD, (01)00884838067981(21)B1MXCJ, (01)00884838067981(21)B24NX6, (01)00884838067981(21)B2H090, (01)00884838067981(21)B20GCQ, (01)00884838067981(21)B2GZXG, (01)00884838067981(21)B2PKDG, (01)00884838067745(21)B288RJ, (01)00884838067981(21)B1JZBN, (01)00884838068032(21)B3FTTH, (01)00884838067981(21)B2D8Z8, (01)00884838067981(21)B20NHK, (01)00884838067981(21)B20PQF, (01)00884838067981(21)B24HYW, (01)00884838067981(21)B24NL5, (01)00884838067981(21)B3LFMY, (01)00884838067981(21)B3PYKZ, (01)00884838067981(21)B20J4T, (01)00884838067981(21)B2TNT6, (01)00884838067981(21)B3FGP9, (01)00884838067981(21)B1N2Z3, (01)00884838067981(21)B2H4TM, (01)00884838067981(21)B1MZ1L, (01)00884838067981(21)B1N23C, (01)00884838067981(21)B2PQNL, (01)00884838067981(21)B1N2GK, (01)00884838067981(21)B1X2YL, (01)00884838067981(21)B1X2YY, (01)00884838067981(21)B2D922, (01)00884838067981(21)B27FKG, (01)00884838067745(21)B29XQ2, (01)00884838067981(21)B1MZ33, (01)00884838068032(21)B3FTTG, (01)00884838067981(21)B2ZQDL, (01)00884838067981(21)B1MYT4, (01)00884838067981(21)B1X1Q1, (01)00884838067981(21)B1MYPT, (01)00884838067981(21)B2PW0R, (01)00884838067981(21)B1JZZT, (01)00884838067981(21)B1N21X, (01)00884838068032(21)B2PQT2, (01)00884838068032(21)B1RHYC, (01)00884838067981(21)B1WQNB, (01)00884838067981(21)B2TL93, (01)00884838067981(21)F09JPB, (01)00884838068032(21)B1RQ88, (01)00884838068032(21)B39MB2, (01)00884838068032(21)B1RPTV, (01)00884838067981(21)B3HY3K, (01)00884838067981(21)B27PKQ, (01)00884838067981(21)B2TP62, (01)00884838068032(21)B3XVYJ, (01)00884838067745(21)B287WR, (01)00884838068032(21)B1WTC8, (01)00884838068032(21)F054X7, (01)00884838067981(21)B20J9F, (01)00884838064577(21)B1F3XR, (01)00884838068032(21)B1RQ6D, (01)00884838067981(21)B2TTDC, (01)00884838067981(21)B3HXWL, (01)00884838067981(21)B1X1LK, (01)00884838067981(21)B3FGP5, (01)00884838067981(21)B3FT1N, (01)00884838067981(21)B1MXN4, (01)00884838067981(21)B35DK8, (01)00884838067981(21)B2TKV1, (01)00884838067981(21)B24NZ7, (01)00884838067981(21)B27MZ2, (01)00884838067981(21)B2PKDJ, (01)00884838067981(21)B2D6LZ, (01)00884838067981(21)B1N2RF, (01)00884838067981(21)B1X19W, (01)00884838068032(21)B1WRWG, (01)00884838068032(21)B1WT3B, (01)00884838068032(21)B24HVR, (01)00884838067981(21)B2TNK1, (01)00884838067981(21)F09CVV, (01)00884838067981(21)B35H6G, (01)00884838067981(21)B2ZHD3, (01)00884838068032(21)B3Y5T7, (01)00884838064577(21)B1F3QV, (01)00884838064577(21)B287QM, (01)00884838067981(21)B24NLV, (01)00884838067981(21)B20P4P, (01)00884838067745(21)B287YQ, (01)00884838064577(21)B1YRZH, (01)00884838067981(21)B1N10P, (01)00884838067981(21)B2PVXM, (01)00884838067981(21)B3FGPG, (01)00884838067981(21)B1JZDT, (01)00884838067981(21)B1N2HF, (01)00884838067981(21)B32N48, (01)00884838067745(21)B1F3NB, (01)00884838068032(21)B24HP5, (01)00884838068032(21)B20FJ2, (01)00884838067981(21)B27FNP, (01)00884838067745(21)B288VF, (01)00884838067745(21)B2P4CL, (01)00884838067981(21)B2H7X3, (01)00884838067981(21)B1MZ1J, (01)00884838067981(21)B1N1Z6, (01)00884838067981(21)B1WT5N, (01)00884838067981(21)B20NHQ, (01)00884838067981(21)B2H051, (01)00884838064577(21)B1YT29, (01)00884838067981(21)B3FT1M, (01)00884838067981(21)B20G9M, (01)00884838067981(21)B20NKN, (01)00884838067981(21)B35C6M, (01)00884838067981(21)B1X1N6, (01)00884838067981(21)B2TNT7, (01)00884838064577(21)B1F404, (01)00884838068032(21)F07RGD, (01)00884838067981(21)B3HXWP, (01)00884838068032(21)B2GZZ7, (01)00884838067981(21)B3L43R, (01)00884838067981(21)B2D6KR, (01)00884838068032(21)B2TMRW, (01)00884838068032(21)B2ZHGY, (01)00884838067981(21)B1MZW2, (01)00884838067981(21)B1N21Z, (01)00884838067981(21)B2ZJ42, (01)00884838067981(21)B2PVXQ, (01)00884838068032(21)B1X1JW, (01)00884838067981(21)B1N2GM, (01)00884838068032(21)B20FLB, (01)00884838067981(21)B2PWGM, (01)00884838068032(21)B1MZ71, (01)00884838067981(21)B2TTD9, (01)00884838067981(21)B1MXN8, (01)00884838067981(21)B1MY9V, (01)00884838067981(21)B2PP8C, (01)00884838067981(21)B24NZ6, (01)00884838067981(21)B1JZ5K, (01)00884838067981(21)B2TMQM, (01)00884838067981(21)B32N45, (01)00884838067981(21)B3HLRH, (01)00884838067981(21)B1MYT3, (01)00884838067981(21)B32MV7, (01)00884838068032(21)B24HR8, (01)00884838067981(21)B3FHHN, (01)00884838067981(21)B1MXQ6, (01)00884838067981(21)B1N224, (01)00884838067981(21)B1N242, (01)00884838067981(21)B20HV5, (01)00884838067981(21)B20P20, (01)00884838067745(21)B338JL, (01)00884838068032(21)B2TN62, (01)00884838064577(21)B1YRQ6, (01)00884838068032(21)B20JQL, (01)00884838067981(21)B2TTD2, (01)00884838068032(21)B1RKN7, (01)00884838067981(21)B1MZQ9, (01)00884838068032(21)B24M3C, (01)00884838068032(21)B27NRL, (01)00884838068032(21)F05LWR, (01)00884838067981(21)B20J9J, (01)00884838067981(21)B1MXQ5, (01)00884838067981(21)B2PW0W, (01)00884838067981(21)B2ZQDP, (01)00884838067981(21)B1N244, (01)00884838067745(21)B28MCK, (01)00884838067981(21)B20P1Y, (01)00884838067981(21)B2ZJ41, (01)00884838068032(21)B3FGZT, (01)00884838068032(21)B1RQD0, (01)00884838068032(21)B39HXV, (01)00884838067981(21)B20J2F, (01)00884838067981(21)B2TKTZ, (01)00884838068032(21)F07RGG, (01)00884838068032(21)B24NR2, (01)00884838067981(21)B3HXWK, (01)00884838068032(21)B2PR8R, (01)00884838068032(21)B2ZHH0, (01)00884838068032(21)B3ZBG2, (01)00884838067981(21)B24F21, (01)00884838067981(21)B35D19, (01)00884838067981(21)B1MYYJ, (01)00884838068032(21)B35D2L, (01)00884838068032(21)B2TT7B, (01)00884838068032(21)B24HB7, (01)00884838067745(21)B2FXHG, (01)00884838067745(21)B2P4GY, (01)00884838068032(21)B1MZG8, (01)00884838067981(21)B3HM3F, (01)00884838067981(21)B1MZQF, (01)00884838067981(21)B2D6LX, (01)00884838068032(21)B1RKNC, (01)00884838068032(21)B20JHZ, (01)00884838068032(21)B32T8M, (01)00884838067981(21)B1WRXY, (01)00884838067981(21)B1MZQB, (01)00884838067981(21)B2PWWY, (01)00884838068032(21)B2TP12, (01)00884838068032(21)B3XJYT, (01)00884838067981(21)B24HN5, (01)00884838068032(21)B1RPT4, (01)00884838068032(21)B20JLJ, (01)00884838068032(21)B2D9PK, (01)00884838067981(21)B39T2Z, (01)00884838068032(21)B2GZQQ, (01)00884838068032(21)B1WRYT, (01)00884838067981(21)B3FT1T, (01)00884838068032(21)B1N2VP, (01)00884838067981(21)B40GWZ, (01)00884838068032(21)B3HWGD, (01)00884838068032(21)B2TTGT, (01)00884838067981(21)B1WQN8, (01)00884838068032(21)B24MWT, (01)00884838068032(21)B3HMTQ, (01)00884838068032(21)B2D98V, (01)00884838068032(21)B2GZZ5, (01)00884838067981(21)B32MZ4, (01)00884838068032(21)B20FKT, (01)00884838067981(21)B2PNJG, (01)00884838068032(21)B1MXVL, (01)00884838067981(21)B2ZH7N, (01)00884838068032(21)B2TP6X, (01)00884838067981(21)B1RKT6, (01)00884838067981(21)B2D8ZC, (01)00884838067981(21)B1MY9T, (01)00884838067981(21)B20HVC, (01)00884838067981(21)B3LDLB, (01)00884838067981(21)B2TNK0, (01)00884838067981(21)B32MYW, (01)00884838067981(21)B2TMFZ, (01)00884838068032(21)B1RRG9, (01)00884838068032(21)B24NHQ, (01)00884838068032(21)B20K53, (01)00884838068032(21)B24M5Z, (01)00884838068032(21)B27HC2, (01)00884838067981(21)B2ZHDB, (01)00884838067981(21)B35DKC, (01)00884838067981(21)B32N47, (01)00884838067981(21)B3L43W, (01)00884838067981(21)B1JZKN, (01)00884838067981(21)B1MZ36, (01)00884838067981(21)B2PVXW, (01)00884838067745(21)B1NM5B, (01)00884838067981(21)B1WT5K, (01)00884838067981(21)B2TP63, (01)00884838068032(21)B1MZ6W, (01)00884838067981(21)B1MYDG, (01)00884838067981(21)B3FGPF, (01)00884838067981(21)B2TTD7, (01)00884838067981(21)B35BZC, (01)00884838064577(21)B1F3NT, (01)00884838068032(21)B20FJ6, (01)00884838068032(21)B2TP10, (01)00884838067981(21)B35CPW, (01)00884838067981(21)B3FGGW, (01)00884838067981(21)B3FHHW, (01)00884838067981(21)B1RKTD, (01)00884838067745(21)B2P4RJ, (01)00884838068032(21)B2D9P6, (01)00884838067981(21)B3TB30, (01)00884838067981(21)B3ZCR7, (01)00884838067981(21)B3LFN1, (01)00884838067981(21)B2H864, (01)00884838067981(21)B3HWN9, (01)00884838067981(21)B2D9CL, (01)00884838067981(21)B2ZPCQ, (01)00884838067981(21)B35C6G, (01)00884838067981(21)B35H6C, (01)00884838068032(21)B3FGZR, (01)00884838064577(21)B287QN, (01)00884838067981(21)B2PWX6, (01)00884838067981(21)B2PL7P, (01)00884838067981(21)B2PNJJ, (01)00884838064577(21)B1F3Q6, (01)00884838068032(21)B32M6Q, (01)00884838068032(21)B3XJYV, (01)00884838064577(21)B1NM51, (01)00884838067745(21)B2P4DD, (01)00884838068032(21)B20K50, (01)00884838068032(21)B24MJP, (01)00884838067981(21)B1MZQD, (01)00884838067745(21)B2YV60, (01)00884838067745(21)B287YF, (01)00884838067936(21)B24QHR, (01)00884838068032(21)B20HWJ, (01)00884838067981(21)B27P9H, (01)00884838068032(21)B1MZG3, (01)00884838068032(21)F04XFW, (01)00884838068018(21)B20HJ6, (01)00884838068032(21)B20JYC, (01)00884838067981(21)B3LD7Q, (01)00884838067981(21)B2ZH7J, (01)00884838068032(21)B1RL0P, (01)00884838068032(21)B2ZHDK, (01)00884838067981(21)B1N223, (01)00884838068032(21)B3HMTM, (01)00884838068032(21)B27NKJ, (01)00884838068032(21)F07TW1, (01)00884838067981(21)B1JZBP, (01)00884838068032(21)B1RRG2, (01)00884838068032(21)B1X2X6, (01)00884838068032(21)B24HR6, (01)00884838067981(21)B24NL4, (01)00884838067981(21)B35D1F, (01)00884838067981(21)B2TN3X, (01)00884838067981(21)B3HWND, (01)00884838064577(21)B1YRQY, (01)00884838068032(21)B1WY7H, (01)00884838067981(21)B27NDN, (01)00884838068032(21)B24MX3, (01)00884838067981(21)B3HMHR, (01)00884838067981(21)B3TB38, (01)00884838067981(21)B3FGP4, (01)00884838067981(21)B40H6C, (01)00884838068032(21)B3YRMT, (01)00884838068032(21)B3ZBMH, (01)00884838067745(21)B288R8, (01)00884838067981(21)B1JYQR, (01)00884838067981(21)B3HXWN, (01)00884838067981(21)B27NDP, (01)00884838067981(21)B3TFFV, (01)00884838067981(21)B1MZ1G, (01)00884838067981(21)F07L4T, (01)00884838067981(21)B40H6K, (01)00884838067981(21)B2ZPCY, (01)00884838067981(21)B1RQ1Y, (01)00884838067981(21)B2TL95, (01)00884838068032(21)B20FL9, (01)00884838067981(21)B24HHD, (01)00884838067981(21)B3FG7H, (01)00884838067981(21)F09JPG, (01)00884838067981(21)B1MYCY, (01)00884838067981(21)B20NKH, (01)00884838067981(21)B39T30, (01)00884838067981(21)B1MYDC, (01)00884838067981(21)B2PKDD, (01)00884838067745(21)B288M8, (01)00884838068032(21)B24HNV, (01)00884838067981(21)B1X2TB, (01)00884838067981(21)B2TP30, (01)00884838067981(21)B24HHG, (01)00884838067981(21)B3ZCR9, (01)00884838067745(21)B2FXXX, (01)00884838068032(21)B1WT35, (01)00884838067981(21)B2PL7N, (01)00884838068032(21)B1N2F0, (01)00884838068032(21)B20FYC, (01)00884838067981(21)B20GF1, (01)00884838067981(21)B20P1R, (01)00884838067981(21)B1RKTB, (01)00884838068032(21)B1RRG8, (01)00884838068032(21)B2TL44, (01)00884838067981(21)B1X2YP, (01)00884838067981(21)B24PC3, (01)00884838067981(21)B2D920, (01)00884838068032(21)B24M60, (01)00884838067981(21)F08W6K, (01)00884838067981(21)B2H57M, (01)00884838067981(21)B2H7TL, (01)00884838067981(21)B2PKT9, (01)00884838068032(21)B27N2H, (01)00884838068032(21)B27N53, (01)00884838067981(21)B32N4C, (01)00884838068032(21)B2TP04, (01)00884838067981(21)B24NX8, (01)00884838068032(21)B2TLM1, (01)00884838067745(21)B2P4H2, (01)00884838067981(21)B1MYD9, (01)00884838068032(21)B20FL3, (01)00884838068032(21)B1RPZT, (01)00884838068032(21)B24HB5, (01)00884838068032(21)B2ZQ6T, (01)00884838068032(21)F07RGF, (01)00884838067745(21)B1F416, (01)00884838068032(21)B3ZBML, (01)00884838067981(21)B1WRXQ, (01)00884838067981(21)B3FHHL, (01)00884838067981(21)B39MMW, (01)00884838067981(21)B3LFN6, (01)00884838067981(21)B2D8ZB, (01)00884838067981(21)F07L4Q, (01)00884838067981(21)B35DKF, (01)00884838067981(21)B2PL7Y, (01)00884838067981(21)B2PW0Y, (01)00884838067981(21)B1RQF2, (01)00884838067981(21)B24NXX, (01)00884838067981(21)B2H093, (01)00884838067981(21)B40H69, (01)00884838067981(21)B24NL6, (01)00884838067981(21)B1MZQ8, (01)00884838067981(21)B2H7X0, (01)00884838067981(21)B27NDQ, (01)00884838067981(21)B2TKTY, (01)00884838068032(21)B20K52, (01)00884838067981(21)B3HM3D, (01)00884838067981(21)B3HWNH, (01)00884838067981(21)B3TBM2, (01)00884838067981(21)B2TNT9, (01)00884838067981(21)B2PL7T, (01)00884838068032(21)B20JLQ, (01)00884838068032(21)F04XFH, (01)00884838068032(21)B1WTC3, (01)00884838068032(21)B20JQQ, (01)00884838067745(21)B1F42F; Serial No. B1YRNV, B1JZV4, B35BZF, B24PBY, B27T4Z, B35H6L, B1RPRM, B2DGF2, B1MZ1D, B1X19N, B1MYCZ, B1RKT8, B3PYXD, B1MX6P, B2TL3C, B19PX5, B20NKJ, B19PJ2, B1X2X3, B2TN85, B2TTB8, B2ZHBM, B2ZHDD, B1JX1Z, B2TKV5, B40GX1, B1JY2Y, B1RRJL, B3ZCR0, F09JPD, B20G80, B2PKDF, B2TL8Y, B1JX8H, B20NHL, B1YRVW, B1RQDW, B24FWM, B2PKT2, B35P3D, F08W6H, B35C6P, B1MXNG, B20HV9, B1JX04, B19PX7, B1F3XP, B1K09Q, B1F3RL, B1MXCF, B1RPML, B19T8P, B39MMX, B39RBC, B19PNH.B1JX8L, B19PWX, B1RPJP, B19J0Z, B27FQF, B1DFF6, B1MXC2, B1D3Z5, B2TL3B, B1DFT9, B20J83, B3LDL6, B35P3H, B19PR6, B2PKDH, B2D91Y, B1N2Z7, B2TP2K, B1JQD8, B24P7D, B24P7F, B2PVXV, B2H8BR, B39T34, B3HMHL, B1JYQQ, B1X1LF, B3FGP7, B1JPP7, B20J4G, B2D91Z, B35BZJ, B1JZKT, B19TGM, B19T3L, B1JZH3, B1MZW1, B2D6LW, B24NLZ, B19PJ3, B2PNJB, B1JX6B, B2ZQDK, B24M85, B2H07F, B35P3C, B2TTFV, B3HM3G, B1JYG7, B27G10, B19PNN, B2TTFX, B35HLV, B3FG7D, B3FGP8, B3HM3J, B3LD7N, B3LDL3, B3ZCM0, B2PKGD, B1MYB6, B1RQ18, B39RBD, B2TP71, B1N2GT, B20P5V, B24NZN, B19T65, B1N2PB, B1DD06, B1JY7B, B3FGGM, B3HLRQ, B3HMHT, B19PX3, B2D6M2, B3FGGT, B2ZPCX, B20J4P, B2H02V, B2H052, B2H3B2, B19JBH, B1F3WX, B1MZ1H, B19JBL, B1JXMQ, B2TN88, B32T3N, B2D923, B35C6R, B3FHHK, B20PRK, B35DBL, B1015C, B2ZH7P, B1MYYK, B1X1LC, B24HZ1, B3LFN4, B2ZPYN, B1X1LB, B1JYZD, B2TNT0, B39RB9, B3L442, B1MXCD, B1MZQG, B1RRJM, B20HV6, B1RM25, B19TYZ, B35D8Z, B24P7P, B2PP8B, B27PKV, B3TFFT, B1MXTD, B1K0PN, B27PKW, B2TKV2, B2TP2M, B27P92, B1JZXC, B3ZCR8, B2H7TD, B2ZQDT, B19TDT, B2PWGJ, B2PWWZ, B2ZPCT, B1JQV9, B1JZKQ, B20PQD, B2ZHDC, B2H07H, B2PWX1, F07DCT, B1DFZR, B1MYPP, B1JZH5, B1JZZG, B1N2P8, B1RQF1, B1WTD3, B2TMFY, B35DKD, B35HLZ, B19K8H, B2TTG1, B3HXWF, B19J0R, B2ZQDQ, B2PVXL, B19PNM, B1JYV3, B1MXR6, B1MXRC, B2D9CW, B32MV8, B3FT1R, F07L4R, F09CVW, B1X31N, B24M5T, B3ZCR1, B19PHZ, B1JQVC, B2PNJD, B1MYB1, B1DFPZ, B24NZB, B1JYNL, B2TNFG, B0X78J, B39T37, B3FHHM, B1JY7G, B20G9L, B2PW0Q, B2PWGG, B2PWWX, B2PWX3, B2TL91, B32N43, B24HTL, B32T3L, B39M3C, B3FGZQ, B3HWG3, B3HWG9, B2H4TF, B20G9J, B1MY6N, B1JZTX, B2TNK2, B24NG3, B1DDHT, B20P1X, B1DFTC, B2ZHH4, B35JQJ, B39J3K, B2TL90, B2PR5B, B20J9L, B20HGT, B20NKL, B2TL8W, B3HM3L, B1F3N1, B1F3N8, B1X19V, B1X2YV, B27GFD, B35DBH, B1JZZK, B3HXWJ, B1DFQN, B1DFV5, B1JXMK, B20HQH, B2TNT4, B19T39, B1MZBP, B24FWT, B27FKK, B32M94, B39RBH, B3HY3G, B2PNJ9, B24P8W, B2D8ZF, B1WTCY, B20J2H, B2H0BX, B1MXXV, B1D8ZM, B32MYV, B1N2GL, B35C6H, B32M99, B32MYX, B2H077, B3FG77, B2YV0R, B1F33C, B3HM3H, B3L440, B20J9K, B3FHHX, B3HLRM, B1X2YD, B20P4T, B2H57H, B3FGPD, B1MXCJ, B24NX6, B2H090, B20GCQ, B19PLD, B2GZXG, B2PKDG, B288RJ, B1JZBN, B3FTTH, B2D8Z8, B20NHK, B20PQF, B24HYW, B24NL5, B3LFMY, B3PYKZ, B20J4T, B2TNT6, B3FGP9, B1N2Z3, B2H4TM, B1JQDC, B19TDW, B1JQD7, B19T1G, B1MZ1L, B0X79H, B1N23C, B1DFK4, B2PQNL, B1N2GK, B1X2YL, B1X2YY, B2D922, B27FKG, B29XQ2, B1MZ33, B1MRVX, B1DF3J, B105MR, B3FTTG, B1JXVV, B1JPF1, B2ZQDL, B1MYT4, B1X1Q1, B1MYPT, B2PW0R, B1JZZT, B1N21X, B19TGN, B2PQT2, B1RHYC, B1K0NL, B1WQNB, B2TL93, F09JPB, B1RQ88, B39MB2, B1RPTV, B3HY3K, B27PKQ, B2TP62, B3XVYJ, B287WR, B19PHX, B1WTC8, F054X7, B1MXJM, B1DFBM, B20J9F, B19PNL, B1F3XR, B1RQ6D, B2TTDC, B3HXWL, B1DFFV, B1DF84, B1X1LK, B3FGP5, B3FT1N, B1MXN4, B35DK8, B2TKV1, B1JY2W, B1JYXF, B24NZ7, B27MZ2, B2PKDJ, B2D6LZ, B1DFXD, B1N2RF, B1X19W, B1WRWG, B1WT3B, B24HVR, B1DDGZ, B1D4HY, B1MR96, B1D4CC, B2TNK1, F09CVV, B19JMM, B35H6G, B19TLM, B1JQDF, B1DFV4, B1D678, B2ZHD3, B3Y5T7, B1F3QV, B287QM, B24NLV, B19TV9, B20P4P, B287YQ, B1YRZH, B1N10P, B2PVXM, B1JXKC, B3FGPG, B1JZDT, B1N2HF, B1JP28, B32N48, B19T70, B19TVW, B1JZDV, B1JZZL, B1F3NB, B24HP5, B1MXG1, B20FJ2, B27FNP, B288VF, B2P4CL, B2H7X3, B1MZ1J, B1N1Z6, B1WT5N, B20NHQ, B2H051, B1YT29, B3FT1M, B20G9M, B20NKN, B35C6M, B1X1N6, B2TNT7, B1F404, F07RGD, B3HXWP, B19PJ0, B2GZZ7, B3L43R, B19T1J, B2D6KR, B2TMRW, B2ZHGY, B1MZW2, B1N21Z, B2ZJ42, B2PVXQ, B1X1JW, B1N2GM, B20FLB, B1JZ80, B2PWGM, B1MZ71, B2TTD9, B19T3G, B1MXN8, B1MY9V, B1DFXG, B2PP8C, B1JX44, B24NZ6, B1JZ5K, B2TMQM, B32N45, B3HLRH, B1D4NV, B1D8G7, B1MYT3, B1DFF7, B32MV7, B24HR8, B3FHHN, B1JYV4, B1MXQ6, B1N224, B1N242, B20HV5, B20P20, B19PW3, B1JXW0, B338JL, B2TN62, B1DD05, B1YRQ6, B20JQL, B1K0X6, B19TW0, B19T66, B2TTD2, B1RKN7, B1JP6C, B1MZQ9, B24M3C, B27NRL, F05LWR, B20J9J, B1MXQ5, B2PW0W, B1JQXC, B2ZQDP, B19TGG, B1DFBP, B1N244, B28MCK, B20P1Y, B2ZJ41, B1JNZ1, B3FGZT, B1F32K, B1RQD0, B39HXV, B20J2F, B2TKTZ, B1JYQN, F07RGG, B1LTFF, B24NR2, B3HXWK, B2PR8R, B2ZHH0, B1DFP3, B3ZBG2, B24F21, B35D19, B0X76Y, B1MYYJ, B1K0ZG.B35D2L, B2TT7B, B1MRCL, B24HB7, B2FXHG, B1DF0N, B2P4GY, B1MZG8, B3HM3F, B1MZQF, B2D6LX, B1DFQ0, B1RKNC, B20JHZ, B32T8M, B1WRXY, B1DFK7, B1MZQB, B2PWWY, B1JZZW, B1JY2Q, B2TP12, B3XJYT, B24HN5, B1RPT4, B1JQ14, B20JLJ, B2D9PK, B39T2Z, B2GZQQ, B1WRYT, B1DDH5, B0X78Z, B0X79F, B0X77F, B3FT1T, B1N2VP, B1JX03, B40GWZ, B19TH3, B3HWGD, B2TTGT, B1WQN8, B1K0VH, B24MWT, B1D43Z, B1JY9K, B3HMTQ, B2D98V, B2GZZ5, B1JZV7, B32MZ4, B20FKT, B1JX22, B1DFQR, B2PNJG, B1MXVL, B2ZH7N, B2TP6X, B1RKT6, B2D8ZC, B1JQVB, B1MY9T, B20HVC, B3LDLB, B1JQVD, B2TNK0, B32MYW, B1K036, B2TMFZ, B1RRG9, B1FP9D, B24NHQ, B20K53, B24M5Z, B1K0PM, B27HC2, B2ZHDB, B35DKC, B19L93, B32N47, B3L43W, B1JZKN, B1MZ36, B2PVXW, B1NM5B, B19T9C, B1MXVF, B1WT5K, B2TP63, B1DFV7, B1DD8Z, B1MZ6W, B1MYDG, B3FGPF, B1D4G7, B2TTD7, B35BZC, B1F3NT, B20FJ6, B1K0DH, B2TP10, B1JXXR, B1NMFW, B35CPW, B3FGGW, B3FHHW, B1RKTD, B2P4RJ, B2D9P6, B3TB30, B3ZCR7, B3LFN1, B2H864, B3HWN9, B2D9CL, B2ZPCQ, B35C6G, B35H6C, B3FGZR, B287QN, B2PWX6, B2PL7P, B2PNJJ, B1F3Q6, B32M6Q, B3XJYV, B1MX8Q, B1D67B, B1NM51, B2P4DD, B20K50, B24MJP, B1MZQD, B2YV60, B287YF, B24QHR, B1DDTG, B0J5L8, B0X791, B1FP95, B1DF8B, B1JZNM, B20HWJ, B27P9H, B1MZG3, F04XFW, B20JHY, B20JYC, B3LD7Q, B2ZH7J, B1RL0P, B2ZHDK, B1MX8N, B19T3C, B1N223, B1JXRQ, B3HMTM, B27NKJ, B1DF5P, F07TW1, B1JXKH, B1JZBP, B1RRG2, B1X2X6, B24HR6, B1D3R1, B1DD08, B24NL4, B35D1F, B2TN3X, B3HWND, B1YRQY, B1WY7H, B1NM63, B27NDN, B24MX3, B3HMHR, B3TB38, B3FGP4, B40H6C, B1DDPV, B3YRMT, B3ZBMH, B288R8, B1JYNH, B1JYQR, B3HXWN, B27NDP, B1JX23, B1JQV7, B3TFFV, B1MZ1G, B19T9J, F07L4T, B40H6K, B2ZPCY, B1RQ1Y, B2TL95, B20FL9, B24HHD, B3FG7H, B19TVB, F09JPG, B0PF85, B1MYCY, B20NKH, B39T30, B1MYDC, B2PKDD, B1D6C4, B288M8, B24HNV, B1X2TB, B2TP30, B1JQXG, B24HHG, B3ZCR9, B1JY2R, B19TQ6, B1DDWK, B2FXXX, B1WT35, B2PL7N, B1N2F0, B20FYC, B20GF1, B20P1R, B1RKTB, B1JQVG, B1RRG8, B1MRNX, B1DFK3, B1DFPW, B2TL44, B1X2YP, B10127, B24PC3, B2D920, B24M60, F08W6K, B2H57M, B2H7TL, B2PKT9, B27N2H, B27N53, B32N4C, B2TP04, B24NX8, B2TLM1, B2P4H2, B1MYD9, B1DF85, B1JYCZ, B1D610, B20FL3, B1MXJH, B1RPZT, B24HB5, B2ZQ6T, F07RGF, B1F416, B3ZBML, B1WRXQ, B3FHHL, B39MMW, B3LFN6, B2D8ZB, F07L4Q, B1JXVX, B00YXP, B35DKF, B2PL7Y, B2PW0Y, B1DF89, B1RQF2, B24NXX, B2H093, B40H69, B24NL6, B1MZQ8, B2H7X0, B27NDQ, B2TKTY, B20K52, B3HM3D, B3HWNH, B3TBM2, B2TNT9, B2PL7T, B20JLQ, F04XFH, B1WTC3, B20JQQ, B1F42F.
Devices named in this recall
6 devices in the FDA device registry whose identifier appears in the recall's code information.
- Transducer S12-4 989605361973Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic3 recalls0 reports
- Transducer C5-2 989605430242Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic3 recalls0 reports
- Transducer C8-4v 989605364943Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic2 recalls0 reports
- Transducer C8-4v 989803002685Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic2 recalls0 reports
- Transducer C8-4v 989605408153Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic2 recalls0 reports
- Transducer C8-4V 989605430232Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic2 recalls0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97217).
- Z-2347-2025 · Open, Classified3D9-3v Transducer Probe
- Z-2349-2025 · Open, ClassifiedBP10-5ec Transducer Probe
- Z-2350-2025 · Open, ClassifiedC10-3v Transducer Probe
- Z-2351-2025 · Open, ClassifiedC10-4ec Transducer Probe
- Z-2352-2025 · Open, Classifiedc5-1 Transducer Probe
- Z-2353-2025 · Open, ClassifiedC5-2 Transducer Probe
- Z-2354-2025 · Open, ClassifiedC5-2 Lumify Transducer Probe
- Z-2355-2025 · Open, ClassifiedC6-2 Transducer Probe
- Z-2356-2025 · Open, ClassifiedC6-3 Transducer Probe
- Z-2357-2025 · Open, ClassifiedC8-4v Transducer Probe
- Z-2359-2025 · Open, ClassifiedC8-5 Transducer Probe
- Z-2360-2025 · Open, ClassifiedC9-2 Transducer Probe
- Z-2361-2025 · Open, ClassifiedC9-3io Transducer Probe
- Z-2362-2025 · Open, ClassifiedC9-3v Transducer Probe
- Z-2363-2025 · Open, ClassifiedC9-4 Transducer Probe
- Z-2364-2025 · Open, ClassifiedC9-4v Transducer Probe
- Z-2365-2025 · Open, ClassifiedC9-5ec Transducer Probe
- Z-2366-2025 · Open, ClassifiedD2cwc Transducer Probe
- Z-2367-2025 · Open, ClassifiedD2TCD Transducer Probe
- Z-2368-2025 · Open, ClassifiedD5cwc Transducer Probe
- Z-2369-2025 · Open, ClassifiedeL18-4 Transducer Probe
- Z-2370-2025 · Open, ClassifiedL10-4LAP Transducer Probe
- Z-2371-2025 · Open, ClassifiedL11-3 Transducer Probe
- Z-2372-2025 · Open, ClassifiedL12-3 Transducer Probe
- Z-2373-2025 · Open, ClassifiedL12-4 Transducer Probe
- Z-2374-2025 · Open, ClassifiedL12-4 Lumify Transducer Probe
- Z-2375-2025 · Open, ClassifiedL12-5 38mm Transducer Probe
- Z-2376-2025 · Open, ClassifiedL12-5 50 Transducer Probe
- Z-2377-2025 · Open, ClassifiedL15-7io Transducer Probe
- Z-2379-2025 · Open, ClassifiedL17-5 Transducer Probe
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page