C8-5 Transducer Probe recallZ-2359-2025 · Philips Ultrasound, Inc · initiated Jul 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Ultrasound, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
To provide clarification and labelling to define the useful life of ultrasound transducers in the field.
FDA root cause category: Device Design
What the company is doing
An URGENT Medical Device Correction Notice, dated 7/3/25, was mailed to consignees. This notice instructs consignees to locate all Philips Ultrasound transducers in their facility and enter their serial numbers into https://www.transducer.philips.com to obtain the manufacturing certificate for each of their devices. This certificate should be retained with its system documentation. Using the manufacturing certificate consignees are to calculate the useful life of their device. The expected useful life for Transesophageal transducers, X-Matrix transducers, and Lumify transducers is 5 years; The expected useful life for Endocavity transducers and Transcutaneous transducers is 7 years. If a transducer has exceeded its useful life, consignees are instructed to conduct transducer performance tests or contact Philips for guidance. Consignees are asked to complete and return the provided response form and circulate the recall notification to device users for their awareness. Consignees with any questions can contact Philips Customer Care Solutions Center at 1-800-722-9377.
Recall details
- Recall number
- Z-2359-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 3, 2025
- Posted by the FDA
- Aug 22, 2025
- Recalling firm
- Philips Ultrasound, Inc
- Quantity
- 171,322 units
- Product code
- IYN
- FDA submissions
- K030455, K043535, K123754, K132304, K081661
- Distribution
- US Nationwide distribution.
Product description
C8-5 Transducer Probe
Code information (models, lots, serial numbers)
Show the full code information (50k characters)
Model No. 989605391922, 989605438441, 989605440822, 989605387852, 989605387281, 989605391092, 8500-1664-01, 989605352343, 989605352342; UDI: (01)00884838061514(21)B1Q41K, (01)00884838061514(21)B1QB9Z, (01)00884838061514(21)B2859G, (01)00884838061514(21)B2YTR3, (01)00884838061514(21)B2YTTV, (01)00884838061514(21)B3F4JN, (01)00884838061514(21)F08DZC, 010088483806151421F01W6L, (01)00884838061514(21)F04XX6, (01)00884838061514(21)F0FCL7, (01)00884838061514(21)B3B4RD, (01)00884838061514(21)B288HM, (01)00884838061514(21)B2FXYD, (01)00884838061514(21)B2FXZB, (01)00884838061514(21)B1P4JM, (01)00884838061514(21)B1P9ZG, (01)00884838061514(21)B338L3, (01)00884838061514(21)F0C7VZ, (01)00884838061514(21)F0FHNV, (01)00884838061514(21)F0KM0Y, (01)00884838061514(21)B22GYC, (01)00884838061514(21)B21T06, (01)00884838067691(21)B26NBY, (01)00884838061514(21)B34J00, (01)00884838061514(21)B2FXJK, (01)00884838067691(21)B2882D, (01)00884838067691(21)B288VM, (01)00884838067691(21)B2FY2G, (01)00884838061514(21)B2YH6Z, (01)00884838061514(21)B30Q8X, 010088483806151421F0G4ZF, (01)00884838061514(21)F011BM, (01)00884838061514(21)B1QBW2, (01)00884838061514(21)B1QTN7, (01)00884838061514(21)B1QTVX, (01)00884838061514(21)B2YGHZ, (01)00884838061514(21)F018FW, (01)00884838061514(21)B37W7W, (01)00884838061514(21)B22NLJ, (01)00884838067691(21)B338F0, (01)00884838061514(21)B216LD, (01)00884838061514(21)F0B6W5, (01)00884838061514(21)B287ZV, (01)00884838061514(21)B288BD, (01)00884838061514(21)B29XXK, (01)00884838061514(21)B2FXQB, (01)00884838061514(21)B2FXV3, (01)00884838061514(21)B3CW78, (01)00884838061514(21)F08CCW, (01)00884838061514(21)F0FMW0, (01)00884838061514(21)F0L128, (01)00884838061514(21)B1P76L, (01)00884838061514(21)B1NYWZ, (01)00884838061514(21)B1Q340, (01)00884838061514(21)B1Q3CC, (01)00884838061514(21)B338NR, (01)00884838061514(21)B338RN, (01)00884838061514(21)B3J3FV, (01)00884838061514(21)B3J3G8, (01)00884838061514(21)B3M1RQ, (01)00884838061514(21)B3M1RR, (01)00884838061514(21)B3MPP0, (01)00884838061514(21)B3MZLX, (01)00884838061514(21)F089LG, (01)00884838061514(21)B351HN, (01)00884838061514(21)B351XC, (01)00884838061514(21)F09Z11, (01)00884838061514(21)B22GD9, (01)00884838061514(21)B26N7F, (01)00884838061491(21)B2JTQ2, (01)00884838061514(21)B1VKMR, (01)00884838061514(21)B2888T, (01)00884838061514(21)B1P9PT, (01)00884838061514(21)B1Q42P, (01)00884838061514(21)B28BRF, (01)00884838061514(21)B2W90F, (01)00884838067691(21)B2T82Z, (01)00884838061514(21)B1ZQHL, (01)00884838061514(21)B28549, (01)00884838061514(21)B285PT, (01)00884838067691(21)F090YB, (01)00884838067691(21)B2WXYT, (01)00884838061514(21)B28515, (01)00884838061514(21)F014Q9, (01)00884838061514(21)B28N4F, (01)00884838061514(21)F0FXK7, (01)00884838061514(21)B2X93W, (01)00884838061514(21)B1W3RV, (01)00884838061514(21)B3CW8Y, (01)00884838061514(21)B3DGJZ, (01)00884838061514(21)B22GNP, (01)00884838061514(21)B22H1D, (01)00884838061514(21)B1PJ5F, (01)00884838061514(21)B3G4B9, (01)00884838061514(21)B1YCVY, (01)00884838061514(21)F011BR, (01)00884838061514(21)B2P49P, (01)00884838061514(21)B2692H, (01)00884838067691(21)B1ZRKQ, (01)00884838067691(21)B288TC, (01)00884838061514(21)B268YQ, (01)00884838061514(21)B28ZHL, (01)00884838061514(21)B1Q5L1, (01)00884838061514(21)B1Q39L, (01)00884838061514(21)B2P47R, (01)00884838061514(21)B25GQ3, (01)00884838061514(21)F03W7G, (01)00884838061514(21)B1TJK3, (01)00884838061514(21)F09XT2, (01)00884838061514(21)F0GH4Y, (01)00884838061514(21)F091C0, (01)00884838067691(21)B26PDD, (01)00884838067691(21)B1P76K, (01)00884838067691(21)B1ZR2J, (01)00884838067691(21)B2FY19, (01)00884838067691(21)B338JZ, (01)00884838061514(21)B2FXXT, (01)00884838061514(21)B1P4DY, (01)00884838061514(21)B2274H, (01)00884838061514(21)B2274K, (01)00884838061514(21)B28606, (01)00884838061514(21)F032DL, (01)00884838061514(21)B3F6BW, (01)00884838061514(21)B34MNN, (01)00884838061514(21)B22LRZ, (01)00884838061514(21)B29XV9, (01)00884838061514(21)B29XVD, (01)00884838061514(21)B3H14J, (01)00884838061514(21)F02M3R, (01)00884838061514(21)F03W7D, (01)00884838061514(21)F07LXR, (01)00884838061514(21)F0GFF7, (01)00884838067691(21)B1QBPX, (01)00884838061514(21)B25226, (01)00884838061514(21)B288LW, (01)00884838061514(21)B22GG1, (01)00884838061514(21)B3VDGB, (01)00884838061514(21)B269GQ, (01)00884838061514(21)B1PL08, (01)00884838061514(21)F036NG, (01)00884838061514(21)F02CV3, (01)00884838061514(21)F03ZGJ, (01)00884838061514(21)F03ZGL, (01)00884838061514(21)B26Q62, (01)00884838061514(21)B34LN8, (01)00884838061514(21)F01ZTQ, (01)00884838061514(21)B2BGFB, (01)00884838061514(21)B2DVVQ, (01)00884838061514(21)B351DF, (01)00884838061514(21)B35ZV4, (01)00884838061514(21)B2P44Q, (01)00884838061514(21)B38D9V, (01)00884838061514(21)B3JGVY, (01)00884838067691(21)B1PL03, (01)00884838061514(21)B1VJVY, (01)00884838061514(21)B1XP0J, (01)00884838061514(21)B288LT, (01)00884838061514(21)B288LY, (01)00884838067691(21)F003HC, (01)00884838061514(21)B1YCFV, (01)00884838061514(21)B1ZR38, (01)00884838061514(21)B38X0B, (01)00884838061514(21)B28M7W, (01)00884838061514(21)B284Z1, (01)00884838061514(21)B1PJZM, (01)00884838061514(21)B28547, (01)00884838061514(21)B28N99, (01)00884838061514(21)B1VPC9, 010088483806151421F0252G, (01)00884838061514(21)B22GG3, (01)00884838061514(21)B25FHY, (01)00884838061514(21)B26N6L, (01)00884838061514(21)B26N6M, (01)00884838061514(21)B39DGQ, (01)00884838061514(21)B1VKXR, (01)00884838067691(21)B2WXZQ, (01)00884838067691(21)B22M7N, (01)00884838061514(21)B1QT10, (01)00884838061514(21)B2FXND, (01)00884838061514(21)F0BZJQ, (01)00884838061514(21)B22G2X, (01)00884838061514(21)B2C47H, (01)00884838067691(21)B2YGQW, (01)00884838067691(21)B31XC3, (01)00884838061514(21)B1YD1Y, (01)00884838061514(21)B288KC, (01)00884838061514(21)B2YGRP, (01)00884838061514(21)B37W7Y, (01)00884838061514(21)B3GLJ7, (01)00884838061514(21)B3KQX6, (01)00884838061514(21)F003L4, (01)00884838061514(21)F04359, (01)00884838061514(21)B1ZR39, (01)00884838061514(21)B23MG9, (01)00884838061514(21)B3J3FM, (01)00884838061514(21)F04T7X, (01)00884838067691(21)B1ZT5M, (01)00884838067691(21)B31XBB, (01)00884838067691(21)F086DH, (01)00884838061514(21)B31XFD, (01)00884838061514(21)B38D6K, (01)00884838061514(21)B3H11L, (01)00884838061514(21)B3H154, (01)00884838061514(21)B3H17W, (01)00884838061514(21)B1P9TL, (01)00884838061514(21)B269LY, (01)00884838061514(21)B29XG3, (01)00884838061514(21)B1VPC8, (01)00884838061514(21)B2YGJB, (01)00884838061514(21)F0B1XY, (01)00884838067691(21)B338FM, (01)00884838061514(21)B1VJY2, (01)00884838061514(21)B34MJR, (01)00884838061514(21)B34MPM, (01)00884838067691(21)B289L0, (01)00884838061514(21)B22LQB, (01)00884838067691(21)B26Q76, (01)00884838067691(21)B2Y0VM, (01)00884838067691(21)B3RQFB, (01)00884838061514(21)B2P4QR, (01)00884838061514(21)B1NZ1J, (01)00884838061514(21)B29KBH, (01)00884838067691(21)B1WH3H, (01)00884838067691(21)B288NY, (01)00884838067691(21)B288P1, (01)00884838061514(21)B284TH, (01)00884838061514(21)B31L86, (01)00884838061514(21)F02W8N, (01)00884838061514(21)F02WYN, (01)00884838061514(21)B3JGXH, (01)00884838061514(21)B1XD4N, (01)00884838061514(21)B2YHFV, (01)00884838067691(21)B2WXQK, (01)00884838067691(21)B3NBLV, (01)00884838061514(21)B28MBB, (01)00884838061514(21)B29JNC, (01)00884838061514(21)B29D7R, (01)00884838061514(21)B285WY, (01)00884838061514(21)B28N69, (01)00884838061514(21)F040Z3, (01)00884838061514(21)B22G2Y, (01)00884838061514(21)B29DGR, (01)00884838061514(21)B2FY3N, (01)00884838061514(21)B2P4C2, (01)00884838061514(21)B2P4CP, (01)00884838061514(21)B2P4CX, (01)00884838061514(21)B2P4CZ, (01)00884838061514(21)B2P4F2, (01)00884838061514(21)B2P4F6, (01)00884838061514(21)B2P4FL, (01)00884838061514(21)B3G477, (01)00884838067691(21)B289KT, (01)00884838067691(21)B227NR, (01)00884838061514(21)B25G9Z, (01)00884838061514(21)F044ZC, (01)00884838067691(21)B26PJ9, (01)00884838061514(21)B3J3G7, (01)00884838067691(21)B288GZ, (01)00884838061514(21)B22H8Z, (01)00884838061514(21)B2Y0VT, (01)00884838061514(21)B2YGHV, (01)00884838061514(21)F02PZC, (01)00884838061514(21)B1W3XN, (01)00884838061514(21)B22H7H, (01)00884838067691(21)B25FXP, (01)00884838061514(21)B34JK0, (01)00884838061514(21)B1XDH3, (01)00884838061514(21)B2WXRJ, (01)00884838061514(21)B1RTC0, (01)00884838061514(21)B3RBBN, (01)00884838061514(21)B3H17X, (01)00884838067691(21)B3MH53, (01)00884838061514(21)B2FXR0, (01)00884838061514(21)B25H3Q, (01)00884838067691(21)B1VK37, (01)00884838061514(21)B2FXMF, (01)00884838061514(21)B25GTZ, (01)00884838061514(21)B25H9C, (01)00884838061514(21)B26N3D, (01)00884838067691(21)B2FY1M, (01)00884838061514(21)B3JH03, (01)00884838061514(21)B22GYF, (01)00884838061514(21)B269PG, (01)00884838061514(21)B26NQP, (01)00884838067691(21)B1YD1W, (01)00884838061514(21)B211WN, (01)00884838061514(21)B2YTQ0, (01)00884838061514(21)B1PK82, (01)00884838061514(21)B2FXNW, (01)00884838061514(21)B2FXXP, (01)00884838061514(21)B268YW, (01)00884838067691(21)B1PJTF, (01)00884838061514(21)B251P5, (01)00884838061514(21)B3606Z, (01)00884838061514(21)B22N84, (01)00884838061514(21)B1VKZ0, (01)00884838061514(21)B1VPQ7, (01)00884838061514(21)B3GLVJ, (01)00884838061514(21)B3NT58, (01)00884838061514(21)F00Y9W, (01)00884838061514(21)F06HWZ, 010088483806151421F027PG, (01)00884838061514(21)B2277X, (01)00884838061514(21)B2P499, (01)00884838061514(21)F0BZJC, (01)00884838061514(21)B3KM20, (01)00884838067691(21)B1VKDD, (01)00884838067691(21)B2883K, (01)00884838061514(21)B284YZ, (01)00884838061514(21)B28N64, (01)00884838061514(21)B28517, (01)00884838067691(21)B29K8Z, (01)00884838067691(21)B2P458, (01)00884838067691(21)B31L61, (01)00884838061514(21)B34MW3, (01)00884838061514(21)B3J3F4, (01)00884838061514(21)B26PL0, (01)00884838061514(21)B3HDPZ, (01)00884838061514(21)F08DZB, (01)00884838067691(21)F062B5, (01)00884838067691(21)F086DG, (01)00884838061514(21)B26NDP, (01)00884838061514(21)B2P44T, (01)00884838061514(21)B1XNCK, (01)00884838061514(21)F078BG, (01)00884838061514(21)B38X0D, (01)00884838061514(21)B1PK6L, (01)00884838061514(21)B25H66, (01)00884838061514(21)B39DKZ, (01)00884838061514(21)F04PYB, (01)00884838061514(21)B2FXK0, (01)00884838061514(21)B1XDVD, (01)00884838061514(21)B2FY0J, (01)00884838061514(21)B2WXN5, (01)00884838061514(21)B1X975, (01)00884838061514(21)B2YH5N, (01)00884838061514(21)B30Q8B, (01)00884838061514(21)B3C6CX, (01)00884838061514(21)B3C6D4, (01)00884838061514(21)F05XDJ, (01)00884838061514(21)B28896, (01)00884838061514(21)B288BC, (01)00884838061514(21)B2Y0VZ, (01)00884838061514(21)B20Y2M, (01)00884838061514(21)B3606B, (01)00884838061514(21)B2FXRY, (01)00884838061514(21)B34LG1, (01)00884838061514(21)B3CWCV, (01)00884838061514(21)F0651H, (01)00884838061514(21)B2698W, (01)00884838061514(21)B1VJM6, (01)00884838061514(21)B3J3GV, (01)00884838061514(21)B3JGZW, (01)00884838061514(21)B3JGY3, (01)00884838061514(21)B3H16M, (01)00884838067691(21)B1ZRX1, (01)00884838061514(21)B1VKNP, (01)00884838061514(21)B288P6, (01)00884838061514(21)F05JHH, (01)00884838067691(21)B22P8K, (01)00884838061514(21)B31LJ9, (01)00884838061514(21)B31X46, (01)00884838061514(21)B31X6V, (01)00884838061514(21)B34MXC, (01)00884838061514(21)B3KM4P, (01)00884838061514(21)B1RY9F, (01)00884838061514(21)B1VKX7, (01)00884838061514(21)B22G83, (01)00884838061514(21)B39DNL, (01)00884838061514(21)B2YV6X, (01)00884838061514(21)B26B6T, (01)00884838061514(21)F0GF2W, (01)00884838061514(21)B2P4J9, (01)00884838067691(21)F04429, (01)00884838061514(21)B1TVMZ, (01)00884838061514(21)B34MQ1, (01)00884838061514(21)B288DL, (01)00884838061514(21)B288NP, (01)00884838061514(21)B29KB3, (01)00884838061514(21)B2FXZC, (01)00884838061514(21)B2FY04, (01)00884838061514(21)B2FY0H, (01)00884838061514(21)B2FY15, (01)00884838061514(21)B1PL00, (01)00884838061514(21)B1QBPW, (01)00884838067691(21)F08J79, (01)00884838061514(21)B20Y2N, (01)00884838061514(21)F004L1, (01)00884838061514(21)B2WXPT, (01)00884838061514(21)B2Y0TD, (01)00884838061514(21)B1XDQ9, (01)00884838061514(21)B22N0Q, (01)00884838061514(21)B288BH, (01)00884838061514(21)B39DMJ, (01)00884838061514(21)B1VQBL, (01)00884838061514(21)B20Y7R, (01)00884838061514(21)F04TNQ, (01)00884838061514(21)B39YPL, (01)00884838061514(21)B3B4TW, (01)00884838061514(21)B1VK4L, (01)00884838061514(21)B1VP2R, (01)00884838061514(21)B3520K, (01)00884838061514(21)B3CHRJ, (01)00884838061514(21)B30QMR, (01)00884838061514(21)B31X1P, (01)00884838067691(21)B3VJ65, (01)00884838061514(21)B1NY2F, (01)00884838067691(21)B25GZF, (01)00884838061514(21)B1Q5N0, (01)00884838061514(21)B1RT94, (01)00884838061514(21)B3MP5G, (01)00884838067691(21)F062B6, (01)00884838061514(21)B2FY31, (01)00884838067691(21)B3MPRD, (01)00884838061514(21)B3B4VD, (01)00884838067691(21)F0CFH8, (01)00884838061514(21)F03ZVZ, (01)00884838061514(21)F0406Q, (01)00884838061514(21)F0628W, (01)00884838061514(21)F06290, (01)00884838061514(21)B3GLDN, (01)00884838061514(21)F05LBD, (01)00884838061514(21)F0B639, (01)00884838061514(21)F0B87F, (01)00884838061514(21)F0C9D7, (01)00884838061514(21)F0FCL3, (01)00884838061514(21)F0FCL6, (01)00884838061514(21)F0FK9F, (01)00884838067691(21)F091D3, (01)00884838061514(21)B23LVG, (01)00884838061514(21)F02ZVZ, (01)00884838061514(21)B2YV5V, (01)00884838061514(21)B26B5F, (01)00884838061514(21)F09WK7, (01)00884838061514(21)B22G65, (01)00884838061514(21)B29KHQ, (01)00884838061514(21)B389FZ, (01)00884838061514(21)F03ZVT, (01)00884838061514(21)F041BW, (01)00884838061514(21)F041BX, (01)00884838061514(21)B29D14, (01)00884838061514(21)B2B9DY, 010088483806151421F01TC1, 010088483806151421F01X6V, (01)00884838061514(21)B284Z3, (01)00884838061514(21)B34J1N, (01)00884838067691(21)B1XRFB, (01)00884838061514(21)B1P4CH, (01)00884838061514(21)B1PKZG, (01)00884838061514(21)B3C6CK, (01)00884838061514(21)B20XMK, (01)00884838061514(21)B2FXR1, (01)00884838061514(21)B3RJ93, (01)00884838061514(21)F00P44, (01)00884838061514(21)B2T816, (01)00884838061514(21)B22MK0, (01)00884838061514(21)B2P454, (01)00884838061514(21)F0628Q, (01)00884838061514(21)F0FCL1, (01)00884838061514(21)F0FDL9, (01)00884838061514(21)B2Y0VV, (01)00884838061514(21)B2YV83, (01)00884838061514(21)B34LCN, 010088483806151421F01WFQ, (01)00884838061514(21)B3H6WJ, (01)00884838061514(21)B2YVDY, (01)00884838061514(21)B227F4, (01)00884838061514(21)B227NT, (01)00884838061514(21)B1VKDY, (01)00884838061514(21)B23MW2, (01)00884838067691(21)B1V3VX, (01)00884838061514(21)B31LDV, (01)00884838061514(21)B1NYLN, (01)00884838061514(21)B2FXN2, (01)00884838061514(21)B3J2L9, (01)00884838061514(21)F0853L, (01)00884838067691(21)B2FY1N, (01)00884838067691(21)B26Q75, (01)00884838061514(21)B28N4H, (01)00884838061514(21)B2YTZW, (01)00884838061514(21)F03XLJ, (01)00884838061514(21)B28519, (01)00884838067691(21)B338HF, (01)00884838061514(21)B2FXZ9, (01)00884838061514(21)B22M6J, (01)00884838061514(21)B269K3, (01)00884838061514(21)B3LP67, (01)00884838061514(21)B1QT31, (01)00884838061507(21)B2KPHY, (01)00884838061514(21)B29LHY, (01)00884838061514(21)F09B6P, (01)00884838067691(21)B31XC4, (01)00884838067691(21)B31XD1, (01)00884838067691(21)F08FC7, (01)00884838061514(21)B287ZJ, (01)00884838067691(21)F04428, (01)00884838061514(21)F00XXB, (01)00884838067691(21)B1QTVZ, (01)00884838061514(21)B284TN, (01)00884838061514(21)B22G2W, (01)00884838061514(21)B3B4Z3, (01)00884838061514(21)B1X8Y3, (01)00884838067691(21)B1ZRXN, (01)00884838061514(21)B22MLC, (01)00884838061514(21)B1XN8T, (01)00884838061514(21)B211WP, (01)00884838061514(21)B3606H, (01)00884838061514(21)B1ZR37, (01)00884838061514(21)B2880K, (01)00884838061514(21)B2X92N, (01)00884838067691(21)B1ZR5D, (01)00884838061514(21)B1XN4T, (01)00884838061514(21)B1XNQN, (01)00884838061514(21)B28BVD, (01)00884838061514(21)B28FK9, (01)00884838061514(21)B288C4, (01)00884838061514(21)B30Q7Q, (01)00884838061514(21)B2YV51, (01)00884838061514(21)B2FXLG, (01)00884838061514(21)F06291, (01)00884838061514(21)B269DK, (01)00884838061514(21)B288B5, (01)00884838061514(21)B288BV, (01)00884838061514(21)B2YGJX, (01)00884838061514(21)F0GFF9, (01)00884838067691(21)B1VQBT, (01)00884838067691(21)B252CR, (01)00884838061514(21)B2FXQF, (01)00884838061514(21)B2X92Q, (01)00884838061514(21)B2X93Z, (01)00884838067691(21)B289NB, (01)00884838061514(21)B1VJX3, (01)00884838061514(21)B2FXVV, (01)00884838061514(21)B3DGC6, (01)00884838061514(21)B2T9KT, (01)00884838061514(21)B35ZTJ, (01)00884838067691(21)B22LRY, (01)00884838067691(21)B1PKZH, (01)00884838061514(21)B29XRM, (01)00884838061514(21)B1X99P, (01)00884838061514(21)B268YT, (01)00884838061514(21)F03QC5, (01)00884838061514(21)B2YV0H, (01)00884838061514(21)F09JBD, (01)00884838061514(21)B2FXYC, (01)00884838061514(21)B29XWV, (01)00884838067691(21)B3QL3N, (01)00884838061514(21)B288NQ, (01)00884838061514(21)B2YGRJ, (01)00884838061514(21)B29XRY, (01)00884838061514(21)B2880L, (01)00884838061514(21)B2FXQV, (01)00884838061514(21)B2FXR2, (01)00884838061514(21)B3MZLZ, (01)00884838061514(21)B3N1QH, (01)00884838061514(21)F0G6DM, (01)00884838061514(21)F0K3KY, 010088483806151421F0G4ZC, (01)00884838061514(21)F05K63, (01)00884838061514(21)B1VJM4, (01)00884838061514(21)B1VKTH, (01)00884838061514(21)B1WFX3, (01)00884838061514(21)B1P9PR, (01)00884838061514(21)B3PTL0, (01)00884838061514(21)B3QL32, (01)00884838061514(21)B3R4H9, (01)00884838067691(21)B2226T, (01)00884838061514(21)B31XBZ, (01)00884838061514(21)B1PK17, (01)00884838061514(21)B1TW4L, (01)00884838061514(21)B2694T, (01)00884838061514(21)B1VKF0, (01)00884838061514(21)B3HBVT, (01)00884838067691(21)B1VQBP, (01)00884838061514(21)F00XVN, (01)00884838061514(21)F03VGM, (01)00884838061514(21)B20XMD, (01)00884838061514(21)F0FPV3, (01)00884838061514(21)F04PHC, (01)00884838061514(21)B2YGLH, (01)00884838061514(21)B31L1T, (01)00884838061514(21)B1VJM2, (01)00884838061514(21)B2YV9X, (01)00884838061514(21)F05M6F, (01)00884838061514(21)B28829, (01)00884838061514(21)B2889V, (01)00884838061514(21)B30QJ8, (01)00884838061514(21)F027Y4, (01)00884838067691(21)B39DGJ, (01)00884838061514(21)B2P4Q2, (01)00884838061514(21)B374PP, (01)00884838061514(21)B35ZR9, (01)00884838061514(21)B28M9W, (01)00884838061514(21)B28N66, (01)00884838061514(21)B2YH7D, (01)00884838061514(21)B31L2F, (01)00884838061514(21)B31L75, (01)00884838067691(21)B1PK3K, (01)00884838061514(21)B3LMQQ, (01)00884838061514(21)B2Y0VP, (01)00884838061514(21)B3HBVW, (01)00884838061514(21)B288PN, (01)00884838061514(21)B227W6, (01)00884838061514(21)B3F3LB, (01)00884838061514(21)B288JM, (01)00884838061514(21)B1P4JP, (01)00884838061514(21)B3MPQY, (01)00884838061514(21)B3H9LN, (01)00884838061514(21)B3H9LX, (01)00884838061514(21)B2P46G, (01)00884838061514(21)B3H9GB, (01)00884838061514(21)F04W6G, (01)00884838061514(21)B29KB5, (01)00884838061514(21)B2FXN1, (01)00884838061514(21)B288B9, (01)00884838061514(21)B227YN, (01)00884838061514(21)B22G67, (01)00884838061514(21)B22NMD, (01)00884838061514(21)B374X4, (01)00884838061514(21)B1PK8Y, (01)00884838061514(21)B211WM, (01)00884838061514(21)B22311, (01)00884838061514(21)B2FXJF, (01)00884838061514(21)B1ZRM4, (01)00884838061514(21)F0607D, (01)00884838061514(21)F0651Q, (01)00884838061514(21)B286J9, (01)00884838061514(21)B29D4Y, (01)00884838061514(21)B29D7T, (01)00884838061514(21)B29DCR, (01)00884838061514(21)B29XDN, (01)00884838061514(21)B3D85Y, (01)00884838061514(21)B2P46D, (01)00884838061514(21)B2P4CB, (01)00884838061514(21)B1W47V, (01)00884838061514(21)B287WZ, (01)00884838061514(21)B287XL, (01)00884838061514(21)B2FXKF, (01)00884838061514(21)B351KB, (01)00884838061514(21)F048R3, (01)00884838061514(21)B216L6, (01)00884838061514(21)B23LWH, (01)00884838061514(21)B2T802, (01)00884838061514(21)B338Q0, (01)00884838067691(21)B2WXQM, (01)00884838061514(21)B29XXH, (01)00884838061514(21)F0GFF8, (01)00884838061514(21)B288NR, (01)00884838061514(21)B2P47T, (01)00884838061514(21)B287ZH, (01)00884838061514(21)B1QBTX, (01)00884838061514(21)B22NHT, (01)00884838061514(21)B2851B, (01)00884838061514(21)B28585, (01)00884838061514(21)B2859H, (01)00884838061514(21)B2859L, (01)00884838061514(21)B285FN, (01)00884838061514(21)B285FV, (01)00884838061514(21)B285M2, (01)00884838061514(21)B285WZ, (01)00884838061514(21)B286CM, (01)00884838061514(21)B286CR, (01)00884838061514(21)B29XG1, (01)00884838061514(21)B1XD4P, (01)00884838061514(21)F01346, (01)00884838067691(21)B20Y7N, (01)00884838061514(21)F0B1DB, (01)00884838061514(21)B1TKXX, (01)00884838061514(21)B34J5F, (01)00884838061514(21)F05X5Z, (01)00884838061514(21)B2WXQN, (01)00884838061514(21)B1PL04, (01)00884838061514(21)B1PJXL, (01)00884838061514(21)B1X8Y5, (01)00884838061514(21)F04Z5P, (01)00884838067691(21)B3RQRD, (01)00884838067691(21)F0003V, (01)00884838061514(21)B22LM4, (01)00884838061514(21)F0BZJ0, (01)00884838067691(21)B29JND, (01)00884838061514(21)B1TVZN, (01)00884838061514(21)B2FXXV, (01)00884838061514(21)B1P76Q, (01)00884838061514(21)B338QJ, (01)00884838061514(21)F05X5X, (01)00884838061514(21)F0K3L0, (01)00884838067691(21)B1ZRHX, (01)00884838061514(21)B2FXMM, (01)00884838061514(21)F09J7Q, (01)00884838061514(21)B1QBPK, (01)00884838061514(21)B37V5G, (01)00884838061514(21)B3H10B, (01)00884838061514(21)F03MY0, (01)00884838061514(21)B29XXM, (01)00884838061514(21)B351YX, (01)00884838061514(21)B22H71, (01)00884838061514(21)F04HPR, (01)00884838067691(21)B28BFM, (01)00884838067691(21)B338HG, (01)00884838067691(21)F003HB, (01)00884838061514(21)F0BZK0, (01)00884838061514(21)B1Q5HH, (01)00884838061514(21)F04HDW, (01)00884838061514(21)F04R5C, (01)00884838061514(21)F07C5N, (01)00884838067691(21)B1QB9W, (01)00884838067691(21)B2P47V, (01)00884838061514(21)F01Q62, (01)00884838061514(21)F04PY6, (01)00884838061514(21)B2FXLV, (01)00884838067691(21)B2FXVK, (01)00884838061514(21)B1QBB0, (01)00884838061514(21)B1TVN1, (01)00884838061514(21)B1VKMW, (01)00884838061514(21)B252PW, (01)00884838061514(21)B3MYLM, (01)00884838061514(21)B3MZQK, (01)00884838061514(21)F012LR, (01)00884838061514(21)F0FFMP, (01)00884838061514(21)B22MM5, (01)00884838061514(21)B1TKZ0, (01)00884838061514(21)B28B51, (01)00884838061514(21)B1Z3LZ, (01)00884838061514(21)B288BJ, (01)00884838061514(21)B288PW, (01)00884838061514(21)B289HR, (01)00884838061514(21)B2FXNZ, (01)00884838067691(21)B288H2, (01)00884838061514(21)B39YQM, (01)00884838061514(21)B288JQ, (01)00884838061514(21)B2YTR5, (01)00884838061514(21)F069R4, (01)00884838061514(21)F0B6W2, (01)00884838061514(21)F0L2K6, (01)00884838061514(21)B228KN, (01)00884838061514(21)B26B89, (01)00884838067691(21)B31L9M, (01)00884838061514(21)B29XMZ, (01)00884838067691(21)B269YK, (01)00884838061514(21)B34MPK, (01)00884838061514(21)B2YTYX, (01)00884838061750(21)B3KM4N, (01)00884838061514(21)B37420, (01)00884838061514(21)B26Q3C, (01)00884838061514(21)B3QL1R, (01)00884838067691(21)B1Z3N9, (01)00884838061514(21)B2221F, (01)00884838061514(21)B1PJZN, (01)00884838061514(21)B338PX, (01)00884838061514(21)B1Q5KX, (01)00884838061514(21)F04LB4, (01)00884838061514(21)F0G420, (01)00884838061514(21)B374PJ, (01)00884838067691(21)B1YCD1, (01)00884838061514(21)B287ZF, (01)00884838067691(21)F05VT8, (01)00884838061514(21)B37V8L, (01)00884838061514(21)B2BGJN, (01)00884838067691(21)B288V9, (01)00884838061514(21)B288L0, (01)00884838061514(21)B1PK6X, (01)00884838061514(21)B1RTC2, (01)00884838061750(21)B1QBF0, (01)00884838061514(21)B3H10L, (01)00884838061514(21)B1QT3X, (01)00884838061514(21)B3KR00, (01)00884838061514(21)B3C6HL, (01)00884838061514(21)B20YQJ, (01)00884838061514(21)B2FXLY, (01)00884838061514(21)B2DWK1, (01)00884838061750(21)B3N1RQ, (01)00884838067691(21)B1XP1L, (01)00884838067691(21)B1NYL4, (01)00884838061514(21)F011KM, 010088483806151421F0G4ZD, (01)00884838061514(21)F05XDH, (01)00884838061514(21)B2FXVT, (01)00884838061750(21)B1ZQ8Q, (01)00884838067691(21)B1VPGZ, (01)00884838061514(21)B22G45, (01)00884838067691(21)B39DGG, (01)00884838067691(21)B22LRW, (01)00884838067691(21)B2120H, (01)00884838067691(21)B288J1, 010088483806151421F01WFV, 010088483806151421F01WFW, (01)00884838061514(21)B1X8ZM, (01)00884838067691(21)B2YGLB, (01)00884838061514(21)B1PK6M, (01)00884838061514(21)B22MTG, (01)00884838061514(21)F078BH, (01)00884838061514(21)F0BZHK, (01)00884838061514(21)B34J3N, (01)00884838061514(21)B1VQGB, (01)00884838061514(21)B1X8Y1, (01)00884838061514(21)B39DCT, (01)00884838061514(21)F0FLTN, (01)00884838061514(21)B28827, (01)00884838061514(21)B3MP6R, (01)00884838067691(21)F003HD, (01)00884838061514(21)B1PJTZ, (01)00884838061514(21)B1PJWR, (01)00884838061514(21)B1QT4X, (01)00884838061514(21)B1QT50, (01)00884838061514(21)B1QTVV, (01)00884838061514(21)B2854F, (01)00884838061514(21)B2888Y, (01)00884838061514(21)B351KT, (01)00884838061514(21)B3HDPY, (01)00884838061514(21)B269NK, (01)00884838061514(21)B287XV, (01)00884838061750(21)B20XTG, (01)00884838067691(21)B252PQ, (01)00884838061514(21)B2225F, (01)00884838061514(21)F0176W, (01)00884838061514(21)B22N4L, (01)00884838067691(21)B26NQL, (01)00884838061514(21)B3J3FR, (01)00884838061514(21)B288JF, (01)00884838067691(21)B2P4B4, (01)00884838061514(21)B1TVPC, (01)00884838061514(21)F01HWW, (01)00884838061514(21)F03RHK, (01)00884838061514(21)F0GH4X, (01)00884838061514(21)B31X4V, (01)00884838061514(21)B29JNG, (01)00884838061514(21)B1TJD1, (01)00884838067691(21)B25FCP, (01)00884838061514(21)B287ZX, (01)00884838061514(21)B28M78, (01)00884838061514(21)B28M7D, (01)00884838061514(21)B2P4FQ, (01)00884838061514(21)B39DRN, (01)00884838061514(21)B22GYD, (01)00884838061514(21)B1Z4N8, (01)00884838061514(21)B39819, (01)00884838061514(21)B3R4J9, (01)00884838061514(21)B3RJ62, (01)00884838061514(21)B1YCG4, (01)00884838061514(21)F05KJZ, (01)00884838061514(21)F09559, (01)00884838061514(21)B3H6XC, (01)00884838061514(21)B38D7M, (01)00884838067691(21)B39DHG, (01)00884838061514(21)B39DN4, (01)00884838061514(21)B37W9J, (01)00884838061514(21)B1Q5L3, (01)00884838061514(21)B2YTQ8, (01)00884838061514(21)B269JZ, (01)00884838061514(21)B26NDK, (01)00884838061514(21)B26Q66, (01)00884838061514(21)F03ZGN, (01)00884838061514(21)B1P4D1, (01)00884838061514(21)B1PB1N, (01)00884838061514(21)B20VTN, (01)00884838061514(21)F0FLTP, (01)00884838061514(21)B2DWLQ, (01)00884838061514(21)B2P467, (01)00884838061514(21)B28839, (01)00884838061514(21)B2FY0M, (01)00884838061514(21)B2P4LL, (01)00884838067691(21)F06LQT, (01)00884838061514(21)B22H4W, (01)00884838067691(21)B288LD, (01)00884838061514(21)B3CHPN, (01)00884838061514(21)B3JYHX, (01)00884838061514(21)F07KCM, (01)00884838061514(21)B3NBLK, (01)00884838061514(21)B222VB, (01)00884838061514(21)B1XDV9, (01)00884838061514(21)B22G49, (01)00884838061514(21)B20Y9F, (01)00884838061514(21)B3H9KF, (01)00884838061514(21)B3HBWK, (01)00884838061514(21)F0G422, (01)00884838061514(21)B2FXJD, (01)00884838061514(21)F09NZ3, (01)00884838061514(21)B288ND, (01)00884838061514(21)F051M1, (01)00884838061514(21)B25H2M, (01)00884838061514(21)B2FXL5, (01)00884838061514(21)F08TXJ, (01)00884838061514(21)B3CW9R, (01)00884838061514(21)F01B45, (01)00884838061514(21)B284XK, (01)00884838061514(21)B3M1YV, (01)00884838061514(21)B3JH26, (01)00884838061514(21)F05Z8Z, (01)00884838067691(21)B1PJXP, (01)00884838061514(21)F0B8RX, (01)00884838067691(21)B25H3M, (01)00884838061514(21)B3KLYL, (01)00884838061514(21)B1QT40, (01)00884838061514(21)F0B6DV, (01)00884838061514(21)B338JV, (01)00884838061514(21)B28586, (01)00884838061514(21)F0FK3L, (01)00884838061514(21)B269WX, (01)00884838061514(21)B3J79V, (01)00884838061514(21)B1P9TM, (01)00884838061514(21)B3QL6T, (01)00884838061514(21)F078DD, (01)00884838061514(21)F0BDF4, (01)00884838061514(21)F0BFRY, (01)00884838061514(21)F0BZGK, (01)00884838061514(21)B3JV7D, (01)00884838067691(21)B22H74, (01)00884838061514(21)B31L8X, (01)00884838061514(21)F01964, (01)00884838061514(21)B338HB, (01)00884838061514(21)F05W4T, (01)00884838061514(21)F01BKX, (01)00884838061514(21)F01CMJ, (01)00884838061514(21)B3741M, (01)00884838061514(21)B3VJ43, (01)00884838061514(21)F03V0M, (01)00884838061514(21)F085XT, (01)00884838061514(21)B34J0R, (01)00884838061514(21)B29D3H, (01)00884838067691(21)B2P46M, (01)00884838061514(21)B2FY1J, (01)00884838061514(21)B288HL, (01)00884838061514(21)B3HDNY, (01)00884838061514(21)F07QWH, (01)00884838067691(21)B3N1H4, (01)00884838061514(21)B2FXVL, (01)00884838061514(21)B338JJ, (01)00884838061514(21)B284TJ, (01)00884838061514(21)B22GBM, (01)00884838061514(21)B28N9D, (01)00884838061514(21)B2YHBG, (01)00884838061514(21)B288LL, (01)00884838061514(21)B26B8C, 010088483806151421F0G8V3, (01)00884838061514(21)B288T9, (01)00884838067691(21)B20Y7Y, (01)00884838061514(21)B2862M, (01)00884838061514(21)B3JRGQ, (01)00884838061514(21)B26NYL, (01)00884838061514(21)B22G98, (01)00884838061514(21)B3R4K6, (01)00884838061514(21)B1YCW5, (01)00884838061514(21)B22GYB, (01)00884838061514(21)B23M3D, (01)00884838061514(21)F01TW1, (01)00884838067691(21)B25FHG, (01)00884838061514(21)B31LH5, (01)00884838061514(21)F09MW6, (01)00884838061514(21)B3QKJY, (01)00884838067691(21)B31X8T, (01)00884838061514(21)F0B1XR, (01)00884838061514(21)F05K65, (01)00884838061514(21)B2527R, (01)00884838061514(21)B26NX1, (01)00884838061514(21)B20XMC, (01)00884838061514(21)B20XMM, (01)00884838061514(21)B2FY03, 010088483806151421F0G2GM, (01)00884838061514(21)B28M8L, (01)00884838061514(21)B351LV, (01)00884838061514(21)F0GF2V, (01)00884838061514(21)B30Q8L, (01)00884838061514(21)F01DD9, (01)00884838061514(21)B25GL6, (01)00884838061514(21)B3PTWG, (01)00884838061514(21)F08M46, (01)00884838061514(21)F010NR, (01)00884838061514(21)B288PY, (01)00884838061514(21)B22G2R, (01)00884838061514(21)B22GD8, (01)00884838061514(21)B22N87, (01)00884838061514(21)F02Q7Q, (01)00884838061514(21)F0BG4R, (01)00884838061514(21)B1VK35, (01)00884838061514(21)B3JGR9, (01)00884838061514(21)B1TVZM, (01)00884838061514(21)B29XTJ, (01)00884838061514(21)B288GN, (01)00884838061514(21)B1TVR0, (01)00884838061514(21)B22G69, (01)00884838061514(21)B31LJG, (01)00884838061514(21)B21TDJ, (01)00884838061514(21)B288M7, (01)00884838061514(21)B1VPFX, (01)00884838061514(21)B20XMH, (01)00884838061514(21)B20XXF, (01)00884838061514(21)B20Y2V, (01)00884838061514(21)F0BBT3, (01)00884838061514(21)B1Z4N2, (01)00884838061514(21)B1WFNR, (01)00884838061514(21)B1VK36, (01)00884838061514(21)B2T9KF, (01)00884838061514(21)B1VQBN, (01)00884838061514(21)B1C236, (01)00884838061514(21)B1P9RK, (01)00884838061514(21)B1P9Y6, (01)00884838067691(21)B2532Y, (01)00884838061514(21)B1P9Y4, (01)00884838061514(21)B29XT9, (01)00884838061514(21)B2DWDP, (01)00884838061514(21)B34JDQ, (01)00884838061514(21)B1Q3CF, (01)00884838061514(21)B1QBF2, (01)00884838061514(21)B1QBFH, (01)00884838061514(21)B251P3, (01)00884838061514(21)B287ZY, (01)00884838061514(21)B3GLL2, (01)00884838061514(21)B3KQZ1, (01)00884838067691(21)F05Z2J, (01)00884838061514(21)B1PL01, (01)00884838061514(21)B1RYK9, (01)00884838061514(21)B34JF3, (01)00884838061514(21)B3GLN6, (01)00884838061514(21)B26Q3D, (01)00884838061514(21)B1TJHL, (01)00884838061514(21)B1TJHH, (01)00884838061514(21)B2285Q, (01)00884838061514(21)B2YV8X, (01)00884838061514(21)B3HBRR, (01)00884838061514(21)B1VPFT, (01)00884838061514(21)B3D0VN, (01)00884838061514(21)B3KLZB, (01)00884838061514(21)B1V436, (01)00884838061514(21)B1QBL8, (01)00884838061514(21)B2FY37, (01)00884838067691(21)B1TKYW, (01)00884838061514(21)F0GFF6, (01)00884838061514(21)B1VK1W, (01)00884838061514(21)B253F0, (01)00884838061514(21)B1XD5K, (01)00884838061514(21)B1YCK6, (01)00884838061514(21)B31XC2, (01)00884838061514(21)B2888R, (01)00884838061514(21)B1TVWM, (01)00884838061514(21)B1YCKG, (01)00884838061514(21)B22MJV, (01)00884838061514(21)B2851C, (01)00884838061514(21)B285M4, (01)00884838061514(21)B20XMQ, (01)00884838061514(21)B20W0R, (01)00884838061514(21)B216LF, (01)00884838061514(21)B22GX6, (01)00884838061514(21)B2FXV6, (01)00884838061514(21)B30QF6, (01)00884838061514(21)B22MTB, (01)00884838061514(21)B227HT, (01)00884838061514(21)B1XN8W, (01)00884838067691(21)B216LK, (01)00884838067691(21)B22230, (01)00884838067691(21)B22M7P, (01)00884838061514(21)B20Y2T, (01)00884838061514(21)B22MQ7, (01)00884838061514(21)B227W2, (01)00884838061514(21)B20VTM, (01)00884838061514(21)B3GLRV, (01)00884838061514(21)B3H6TC, (01)00884838061514(21)B251P2, (01)00884838061514(21)F05VJX, (01)00884838061514(21)B25GG2, (01)00884838061514(21)B2YTQF, (01)00884838061514(21)B25FHW, (01)00884838061514(21)B25FX1, (01)00884838061514(21)B29XVL, (01)00884838061514(21)B288HR, (01)00884838061514(21)B269DG, (01)00884838061514(21)B28MR9, (01)00884838067691(21)B3VJ5Y, (01)00884838061514(21)B269RN, (01)00884838061514(21)B26B6V, (01)00884838061514(21)F07RRZ, 010088483806151421F0G6DN, (01)00884838061514(21)B289HP, (01)00884838061514(21)B28BLL, (01)00884838061514(21)B28518, (01)00884838061514(21)B26BM4, (01)00884838061514(21)B39DM7, (01)00884838061514(21)B2FXQR, (01)00884838061514(21)F0FCL9, (01)00884838061514(21)B2854H, (01)00884838061514(21)F01B41, (01)00884838061514(21)F03NWY, (01)00884838061514(21)B2WXZM, (01)00884838061514(21)B285ZB, (01)00884838061514(21)B29XPN, (01)00884838067691(21)B26PD9, (01)00884838061514(21)B2859M, (01)00884838061514(21)B26B0J, (01)00884838061514(21)B29XP2, (01)00884838061514(21)B288H6, (01)00884838061514(21)B2YV70, (01)00884838061514(21)B2P4CK, (01)00884838061514(21)B29XRN, (01)00884838061514(21)B287XK, (01)00884838061514(21)B288GW, (01)00884838061514(21)B28825, (01)00884838061514(21)B39DFZ, (01)00884838061514(21)B3C671, (01)00884838061514(21)B1ZRFW, (01)00884838061514(21)B31X66, (01)00884838061514(21)F03V0K, (01)00884838061514(21)B2FXMZ, (01)00884838061514(21)B288FH, (01)00884838061514(21)B30QMX, (01)00884838067691(21)F08WF2, (01)00884838061514(21)B287ZZ, (01)00884838061514(21)B288BW, (01)00884838061514(21)F04PHH, (01)00884838061514(21)F04RJQ, (01)00884838061514(21)B351LT, (01)00884838061514(21)B288LP, (01)00884838061514(21)B288BL, (01)00884838061514(21)B3QKK3, (01)00884838061514(21)B2P44X, (01)00884838061514(21)B1P4CW, (01)00884838061514(21)B1P4P5, (01)00884838061514(21)B288PP, (01)00884838061514(21)B374T2, (01)00884838061514(21)B2W90J, (01)00884838061514(21)B2T83K, (01)00884838061514(21)B2P4PY, (01)00884838061514(21)B39YMW, (01)00884838061514(21)B2Y0QW, (01)00884838061514(21)B29XT8, (01)00884838061514(21)B30Q9F, (01)00884838061514(21)F0B3VD, (01)00884838061514(21)B2YTXK, (01)00884838061514(21)B3H17M, (01)00884838061514(21)B2YH5F, (01)00884838061514(21)B30QB2, (01)00884838061514(21)B338J1, (01)00884838061514(21)B2YGJ9, (01)00884838061514(21)B1RYK6, (01)00884838061514(21)B2YH31, (01)00884838061514(21)B2YV97, (01)00884838061514(21)B30Q8N, (01)00884838061514(21)B2YGYR, (01)00884838061514(21)B31XF5, (01)00884838061514(21)F0FJJN, (01)00884838061514(21)B3DGJV, (01)00884838061514(21)B30QF4, (01)00884838061514(21)B30QFB, (01)00884838061514(21)B2YGWY, (01)00884838061514(21)B2880N, (01)00884838061514(21)B2YHHJ, (01)00884838061514(21)B31LG0, (01)00884838061514(21)F01NYP, (01)00884838061514(21)F01TBZ, (01)00884838061514(21)B1P4FT, (01)00884838061514(21)B3D86W, (01)00884838061514(21)B3F3NV, (01)00884838061514(21)B29XT0, (01)00884838061514(21)F01QRM, (01)00884838061514(21)F01QX9, 010088483806151421F01WRF, 010088483806151421F02766, (01)00884838061514(21)B28N25, (01)00884838061514(21)F0K3KX, (01)00884838061514(21)B31X7W, (01)00884838061514(21)F041BV, (01)00884838061514(21)B2FXJW, (01)00884838061514(21)B3DGHY, (01)00884838061514(21)B37WF1, (01)00884838061514(21)B39DQT, (01)00884838061514(21)F09XWB, (01)00884838061514(21)B3B4RL, (01)00884838061514(21)B3B4V1, (01)00884838067691(21)B26B6L, (01)00884838061514(21)B3GLG8, (01)00884838061514(21)F04W5K, (01)00884838061514(21)B3PKBT, (01)00884838061514(21)B3G4BH, (01)00884838061514(21)B3H9DN, (01)00884838061514(21)B3H9DX, (01)00884838061514(21)B3HBW8, (01)00884838061514(21)B3HBWC, (01)00884838061514(21)B3MPP3, (01)00884838061514(21)B1PL0C, (01)00884838061514(21)F08132, (01)00884838061514(21)F02Y0N, (01)00884838061514(21)F09DW6, (01)00884838061514(21)B3D83X, (01)00884838061514(21)F0043R, (01)00884838061514(21)B3VDHX, (01)00884838061514(21)B3QKJ0, (01)00884838061514(21)F01471, 010088483806151421F0236N, (01)00884838061514(21)F00PJQ, (01)00884838061514(21)B2YHFY, (01)00884838061514(21)F00LQB, (01)00884838061514(21)F00LQC, (01)00884838061514(21)B338K1, (01)00884838061514(21)F00KXF, (01)00884838061514(21)F06L9K, (01)00884838061514(21)B20Y7W, (01)00884838061514(21)F00WPF, (01)00884838061514(21)B31X6R, (01)00884838061514(21)F00YWG, (01)00884838061514(21)F00ZLC, (01)00884838061514(21)F010NT, (01)00884838061514(21)F01H35, (01)00884838061514(21)B1QBBD, 010088483806151421F027PD, (01)00884838061514(21)F0FH3M, (01)00884838067691(21)B288KQ, (01)00884838061514(21)F0BB7G, (01)00884838061514(21)F03MXT, 010088483806151421F0G7BD, (01)00884838061514(21)B23MXR, (01)00884838061514(21)F04QMK, (01)00884838061514(21)B29XNN, (01)00884838061514(21)B20Y9N, (01)00884838061514(21)F03Y1Z, (01)00884838061514(21)B3JWXK, (01)00884838061514(21)B1XD5G, (01)00884838061514(21)F09YL4, (01)00884838061514(21)F0BK4T, (01)00884838061514(21)B2FXZZ, (01)00884838061514(21)B22P8H, (01)00884838061514(21)F09D5Y, (01)00884838067691(21)B1RY9P, (01)00884838061514(21)B1PJ5J, (01)00884838067691(21)B1NYL7, (01)00884838061514(21)B30Q7D, (01)00884838061514(21)B3H5ZX, (01)00884838061514(21)F00P43, (01)00884838061514(21)F0B52T, (01)00884838061514(21)B1WGC7, (01)00884838061514(21)F0F304, (01)00884838061514(21)B3MYL5, (01)00884838067691(21)B28MRC, (01)00884838067691(21)B338DW, (01)00884838061514(21)B1QTP3, (01)00884838061514(21)B2YH1Q, (01)00884838061514(21)B30Q8P, (01)00884838061514(21)B34J72, (01)00884838061514(21)B22P8F, (01)00884838061514(21)B20VTW, (01)00884838061514(21)B2YHHH, (01)00884838061514(21)B31X41, (01)00884838061514(21)B1P4H9, (01)00884838061514(21)B1RY9T, 010088483806151421F0FVXL, (01)00884838061514(21)B3KYV4, (01)00884838061514(21)F029X6, (01)00884838061514(21)B29XRX, (01)00884838061514(21)F0FJJR, (01)00884838061514(21)F0831N, (01)00884838061514(21)F095YL, (01)00884838061514(21)B3VJ10, (01)00884838061514(21)F0BHDM, (01)00884838061514(21)F0BZGT, (01)00884838061514(21)F07Z1M, (01)00884838061514(21)F08HRR, (01)00884838061514(21)F04L4H, (01)00884838061514(21)F092R2, (01)00884838061514(21)B2P47J; Serial No. B1Q41K, B1QB9Z, B2859G, B2YTR3, B2YTTV, B3F4JN, F08DZC, F01W6L, B1LH30, B1LNMB, F04XX6, F0FCL7, B3B4RD, B288HM, B2FXYD, B2FXZB, B1P4JM, B1P9ZG, B1LH4Z, B338L3, F0C7VZ, F0FHNV, F0KM0Y, B22GYC, B21T06, B26NBY, B34J00, B2FXJK, B2882D, B288VM, B2FY2G, B2YH6Z, B30Q8X, F0G4ZF, F011BM, B1QBW2, B1QTN7, B1QTVX, B1K2GR, B2YGHZ, F018FW, B37W7W, B22NLJ, B338F0, B216LD, F0B6W5, B1LHYX, B287ZV, B288BD, B29XXK, B2FXQB, B2FXV3, B3CW78, F08CCW, F0FMW0, F0L128, B1P76L, B1NYWZ, B1Q340, B1Q3CC, B338NR, B338RN, B1MJTF, B3J3FV, B3J3G8, B3M1RQ, B3M1RR, B3MPP0, B3MZLX, F089LG, B351HN, B351XC, F09Z11, B22GD9, B2FXM2, B26N7F, B2FXR5, B1VKMR, B1K54R, B1DWDF, B1DTXD, B15C90, B2888T, B1P9PT, B1Q42P, B28BRF, B2W90F, B2T82Z, B1ZQHL, B28549, B285PT, F090YB, B1LH6B, B2WXYT, B1FDVW, B28515, F014Q9, B1FRWV, B28N4F, F0FXK7, B1F2B5, B1DTRV, B2X93W, B1W3RV, B3CW8Y, B3DGJZ, B22GNP, B22H1D, B1PJ5F, B3G4B9, B1YCVY, F011BR, B2P49P, B2692H, B1ZRKQ, B288TC, B1F0TQ, B1K2CC, B268YQ, B28ZHL, B1Q5L1, B1K7WB, B1FT1J, B1Q39L, B2P47R, B25GQ3, F03W7G, B1TJK3, B1K6ZM, F09XT2, F0GH4Y, B1KX6M, F091C0, B26PDD, B1P76K, B1ZR2J, B1GJG5, B2FY19, B338JZ, B2FXXT, B1P4DY, B2274H, B2274K, B28606, F032DL, B1F5CH, B3F6BW, B34MNN, B22LRZ, B29XV9, B29XVD, B3H14J, F02M3R, F03W7D, F07LXR, F0GFF7, B1QBPX, B25226, B288LW, B22GG1, B3VDGB, B1KXDP, B1K4WC, B269GQ, B1PL08, F036NG, F02CV3, F03ZGJ, F03ZGL, B26Q62, B1MHNF, B34LN8, F01ZTQ, B2BGFB, B2DVVQ, B351DF, B35ZV4, B2P44Q, B38D9V, B3JGVY, B1PL03, B1VJVY, B1XP0J, B288LT, B288LY, F003HC, B1YCFV, B1ZR38, B1MHNJ, B38X0B, B28M7W, B284Z1, B1PJZM, B28547, B28N99, B1VPC9, F0252G, B22GG3, B25FHY, B26N6L, B26N6M, B39DGQ, F0433V, B1CT63, B1VKXR, B2WXZQ, B22M7N, B1QT10, B2FXND, F0BZJQ, B22G2X, B2C47H, B2YGQW, B31XC3, B1KZFD, B1GQ0P, B1YD1Y, B288KC, B2YGRP, B37W7Y, B3GLJ7, B3KQX6, F003L4, F04359, B1ZR39, B23MG9, B3J3FM, F04T7X, B1ZT5M, B31XBB, F086DH, B1FCKD, B1K82M, B1GPQ1, B1K2PK, B31XFD, B38D6K, B3H11L, B3H154, B3H17W, B1NVKB, B1K53K, B1P9TL, B269LY, B29XG3, B1VPC8, B2YGJB, F0B1XY, B338FM, B1VJY2, B34MJR, B34MPM, B289L0, B22LQB, B26Q76, B2Y0VM, B3RQFB, B1KZ2B, B1FZ7Z, B2P4QR, B1NZ1J, B1K2BK, B29KBH, B1WH3H, B288NY, B288P1, B284TH, B31L86, F02W8N, F02WYN, B1M9G5, B3JGXH, B1XD4N, B2YHFV, B2WXQK, B3NBLV, F041BY, B28MBB, B29JNC, B1FZ0Y, B29D7R.B285WY, B28N69, F040Z3, B22G2Y, B29DGR, B2FY3N, B2P4C2, B2P4CP, B2P4CX, B2P4CZ, B2P4F2, B2P4F6, B2P4FL, B3G477, B289KT, B227NR, B25G9Z, F044ZC, B1F0W7, B1F5KF, B26PJ9, B3J3G7, B288GZ, B22H8Z, B2Y0VT, B2YGHV, F02PZC, B1W3XN, B22H7H, B1DX8V, B25FXP, B34JK0, B1XDH3, B2WXRJ, B1F9Y5, B1RTC0, B3RBBN, B3H17X, B1MHQZ, B3MH53, B2FXR0, B25H3Q, B1VK37, B1GQ5M, B2FXMF, B25GTZ, B25H9C, B26N3D, B1DTXF, B2FY1M, B1KZ6V, B1CT0P, B3JH03, B22GYF, B269PG, B26NQP, B1YD1W, B1M2YC, B1CWWB, B211WN, B2YTQ0, B1PK82, B2FXNW, B2FXXP, B1K795, B268YW, B1PJTF, B251P5, B3606Z, B22N84, B1VKZ0, B1VPQ7, B3GLVJ, B3NT58, F00Y9W, F06HWZ, F027PG, B1FWNY, B2277X, B2P499, F0BZJC, B3KM20, B1VKDD, B2883K, B284YZ, B28N64, B28517, B29K8Z, B2P458, B31L61, B34MW3, B3J3F4, B1GFL6, B26PL0, B1F59Q, B3HDPZ, B1GK0P, B1GJCZ, F08DZB, F062B5, F086DG, F043ZR, B26NDP, B2P44T, B1LHYR, B1XNCK, F078BG, B38X0D, B1PK6L, B1NYG1, B25H66, B39DKZ, F04PYB, B2FXK0, B1ML4N, B1NVNL, B1XDVD, B2FY0J, B2WXN5, B1X975, B1M9NJ, B2YH5N, B30Q8B, B3C6CX, B3C6D4, F05XDJ, B28896, B288BC, B2Y0VZ, B1FW73, B1GJ8G, B20Y2M, B3606B, B2FXRY, B34LG1, B3CWCV, B1F4Y8, F0651H, B1K2FF, B2698W, B1VJM6, B3J3GV, B3JGZW, B288V4, B3JGY3, B3H16M, B1ZRX1, B1VKNP, B288P6, B1K7R3, B3NVMK, F05JHH, B22P8K, B31LJ9, B31X46, B31X6V, B34MXC, B3KM4P, B1RY9F, B1VKX7, B22G83, B39DNL, B2YV6X, B26B6T, F0GF2W, B2P4J9, F04429, B1TVMZ, B1F5FH, B34MQ1, B288DL, B288NP, B29KB3, B2FXZC, B2FY04, B2FY0H, B2FY15, B1PL00, B1QBPW, B1DX8Z, F08J79, B20Y2N, F004L1, B2WXPT, B2Y0TD, B1K2BD, B1F51T, B1XDQ9, B22N0Q, B288BH, B39DMJ, B1VQBL, B20Y7R, F04TNQ, B28M9L, B39YPL, B3B4TW, B1VK4L, B1VP2R, B3520K, B3CHRJ, B30QMR, B31X1P, B3VJ65, B1NY2F, B25GZF, B1Q5N0, B1RT94, B3MP5G, F062B6, B2FY31, B3MPRD, B3B4VD, F0CFH8, F03ZVZ, F0406Q, F0628W, F06290, B3GLDN, F05LBD, F0B639, F0B87F, F0C9D7, F0FCL3, F0FCL6, F0FK9F, F091D3, B23LVG, F02ZVZ, B2YV5V, B26B5F, F09WK7, B22G65, B29KHQ, B1MQJL, B389FZ, F03ZVT, F041BW, F041BX, B29D14, B2B9DY, B1CQZF, B1NYRT, F01TC1, F01X6V, B284Z3, B34J1N, B1XRFB, B28M9J, B1P4CH, B1PKZG, B1MKL6, B1MHQY, B3C6CK, B20XMK, B2FXR1, B3RJ93, F00P44, B1F51Y, B2T816, B22MK0, B2P454, F0628Q, F0FCL1, F0FDL9, B2Y0VV, B2YV83, B34LCN, F01WFQ, B3H6WJ, B2YVDY, B227F4, B227NT, B1VKDY, B23MW2, B1V3VX, B31LDV, B1NYLN, B2FXN2, B3J2L9, F0853L, B2FY1N, B1D0NY, B26Q75, B1K70X, B28N4H, B2YTZW, B1K85W, F03XLJ, B1DTRR, B28519, B338HF, B2FXZ9, B22M6J, B269K3, B3LP67, B1QT31, B2889N, B29LHY, F09B6P, B31XC4, B31XD1, F08FC7, B1KX1R, B287ZJ, F04428, F00XXB, B1NYNQ, B1K7CL, B1QTVZ, B284TN, B22G2W, B1GQ9Q, B1GJ8F, B3B4Z3, B1X8Y3, B1ZRXN, B22MLC, B1XN8T, B1CT0M, B211WP, B3606H, B1ZR37, B2880K, B2X92N, B1ZR5D, B1XN4T, B1XNQN, B28BVD, 02V1DY, B28FK9, B1DX90, B1F5R3, B1FW8T, B288C4, B1K7TW, B1K76M, B30Q7Q, B2YV51, B2FXLG, F06291, B269DK, B288B5, B288BV, B2YGJX, F0GFF9, B1VQBT, B252CR, B2FXQF, B2X92Q, B2X93Z, B289NB, B1VJX3, B2FXVV, B3DGC6, B2T9KT, B35ZTJ, B1FDN9, B1F29V, B22LRY, B1DW71, B1PKZH, B29XRM, B1X99P, B268YT, F03QC5, B2YV0H, F09JBD, B2FXYC, B29XWV, B3QL3N, B1CT74, B1CQX1, B288NQ, B2YGRJ, B29XRY, B1K7ZH, B1FT0M, B2880L, B2FXQV, B2FXR2, B3MZLZ, B3N1QH, F0G6DM, F0K3KY, F0G4ZC, F05K63, B1MK1D, B1LD9W, B1M9KV, B1VJM4, B1VKTH, B1WFX3, B1P9PR, B3PTL0, B3QL32, B3R4H9, B2226T, B1GFL8, B31XBZ, B1PK17, B1TW4L, B2694T, B1VKF0, B3HBVT, B1VQBP, F00XVN, F03VGM, B20XMD, F0FPV3, B12RND, F04PHC, B2YGLH, B31L1T, B1VJM2, B2YV9X, F05M6F, B28829, B2889V, B30QJ8, F027Y4, B39DGJ, B2P4Q2, B374PP, B35ZR9, B28M9W, B28N66, B2YH7D, B31L2F, B31L75, B1K6TH, B1K7WC, B1PK3K, B3LMQQ, B2Y0VP, B3HBVW, B288PN, B29XDP, B227W6, B3F3LB, B288JM, B1GFNF, B1P4JP, B3MPQY, B3H9LN, B3H9LX, B1DX5V, B2P46G, B3H9GB, F04W6G, B29KB5, B1CW42, B2FXN1, B288B9, B227YN, B1LG4R, B1KZ2M, B22G67, B22NMD, B374X4, B1PK8Y, B211WM, B22311, B2FXJF, B1ZRM4, B1CT73, B1GJ2N, F0607D, F0651Q, B286J9, B29D4Y, B29D7T, B29DCR, B29XDN, B3D85Y, B2P46D, B2P4CB, B1W47V, B287WZ, B287XL, B1FDRW, B2FXKF, B351KB, F048R3, B216L6, B23LWH, B2T802, B338Q0, B2WXQM, B1M9NK, B29XXH, F0GFF8, B288NR, B2P47T, B287ZH, B1QBTX, B22NHT, B2851B, B28585, B2859H.B2859L, B285FN, B285FV, B285M2, B285WZ, B286CM, B286CR, B29XG1, B1XD4P, F01346, B20Y7N, B0T4HT, F0B1DB, B1TKXX, B34J5F, F05X5Z, B2WXQN, B1PL04, B1PJXL, B1X8Y5, F04Z5P, B3RQRD, F0003V, B1M2TN, B1K77L, B22LM4, F0BZJ0, B29JND, B1K2MM, B1GFGB, B1TVZN, B2FXXV, B1P76Q, B338QJ, F05X5X, F0K3L0, B1K2JB, B1ZRHX, B2FXMM, F09J7Q, B1QBPK, B37V5G, B3H10B, F03MY0, B1F0Z5, B29XXM, B351YX, B22H71, F04HPR, B28BFM, B338HG, F003HB, F0BZK0, B1GFL7, B1FWRF, B29JRG, B1Q5HH, F04HDW, F04R5C, F07C5N, B1QB9W, B2P47V, F01Q62, F04PY6, B2FXLV, B2FXVK, B1QBB0, B1TVN1, B1VKMW, B252PW, B3MYLM, B3MZQK, F012LR, F0FFMP, B1DXH4, B22MM5, B1TKZ0, B28B51, B1Z3LZ, B288BJ, B288PW, B289HR, B2FXNZ, B1GPLX, B288H2, B39YQM, B288JQ, B2YTR5, F069R4, F0B6W2, F0L2K6, B228KN, B26B89, B31L9M, B29XMZ, B269YK, B1GFNH, B1LD82, B34MPK, B2YTYX, B3KM4N, B37420, B26Q3C, B3QL1R, B1Z3N9, B2221F, B1PJZN, B338PX, B38X36, B1Q5KX, F04LB4, F0G420, B374PJ, B1YCD1, B287ZF, B1FCK6, F05VT8, B37V8L, B2BGJN, B1KYZM, B1LH95, B288V9, B288L0, B1PK6X, B1RTC2, B1QBF0, B3H10L, B1QT3X, B1M9G4, B3KR00, B3C6HL, B20YQJ, B2FXLY, B2DWK1, B3N1RQ, B1MHFR, B1XP1L, B1NYL4, B1FWB7, F011KM, F0G4ZD, F05XDH, B2FXVT, B1ZQ8Q, B1VPGZ, B22G45, B39DGG, F05N5V, F05ML7, B22LRW, B2120H, B288J1, F01WFV, F01WFW, B1X8ZM, B2YGLB, B1PK6M, B22MTG, F078BH, F0BZHK, B1GPLT, B34J3N, B1MHG1, B1VQGB, B1X8Y1, B39DCT, F0FLTN, B1ML3N, B28827, B3MP6R, F003HD, B1PJTZ, B1PJWR, B1QT4X, B1QT50, B1QTVV, B2854F, B1GF28, B1MKKC, B2888Y, B351KT, B3HDPY, B269NK, B287XV, B20XTG, B1LQBN, B252PQ, B2225F, F0176W, B1GJZD, B22N4L, B26NQL, B3J3FR, B1MK1F, B288JF, B1F5R4, B2P4B4, B1TVPC, F01HWW, F03RHK, F0GH4X, B29KRC, B31X4V, B29JNG, B1TJD1, B25FCP, B287ZX, B28M78, B28M7D, B2P4FQ, B1GFGG, B39DRN, B22GYD, B1Z4N8, F043ZQ, B39819, B3R4J9, B3RJ62, B1YCG4, F05KJZ, F09559, B3H6XC, B38D7M, B39DHG, B1F4RW, B39DN4, B37W9J, B1GQCJ, B1Q5L3, B2YTQ8, B1DXHX, B269JZ, B26NDK, B26Q66, F03ZGN, B1P4D1, B1PB1N, B20VTN, F0FLTP, B2DWLQ, B2P467, B28839, B2FY0M, B1GQ5L, B1LH5W, B2P4LL, F06LQT, B22H4W, B288LD, B3CHPN, B3JYHX, F07KCM, B3NBLK, B222VB, B1XDV9, B22G49, B20Y9F, B1GJ4W, B3H9KF, B3HBWK, F0G422, B1K7HX, B2FXJD, F09NZ3, F05W4P, B288ND, F051M1, B25H2M, B2FXL5, F08TXJ, B3CW9R, F01B45, B284XK, B126ZL, B3M1YV, B3JH26, F05Z8Z, B1PJXP, F0B8RX, B25H3M, B3KLYL, B1QT40, F0B6DV, B338JV, B28586, F0FK3L, B1LH51, B269WX, B3J79V, B1P9TM, B3QL6T, F078DD, F0BDF4, F0BFRY, F0BZGK, B3JV7D, B22H74, B31L8X, F01964, B338HB, F05W4T, F01BKX, F01CMJ, B3741M, B3VJ43, F03V0M, F085XT, B34J0R, B29D3H, B2P46M, B2FY1J, B288HL, B3HDNY, F07QWH, B3N1H4, B2FXVL, B338JJ, B284TJ, B22GBM, B28N9D, B2YHBG, B126ZM, B288LL, B26B8C, F0G8V3, B288T9, B20Y7Y, B2862M, B3JRGQ, B26NYL, B1FWCW, B22G98, B3R4K6, B1YCW5, B22GYB, B23M3D, B1KXDN, F01TW1, B25FHG, B31LH5, B1NVNJ, F09MW6, B3QKJY, B31X8T, F0B1XR, F05K65, B2527R, B26NX1, B1NYPC, B1KZ10, B1LCV6, B1KZDZ, B1F54F, B20XMC, B20XMM, B1D086, B2FY03, F0G2GM, B28M8L, B351LV, F0GF2V, B1LBRB, B1LH96, B30Q8L, F01DD9, B25GL6, B1F22J, B3PTWG, F08M46, B1KXBH, B1GQG5, F010NR, B1K4YW, B1FWNF, B288PY, B22G2R, B22GD8, B22N87, F02Q7Q, F0BG4R, B1GJ26, B1GQ2W, B1K2BB, B1VK35, B3JGR9, B1FW8W, B1KWZ0, B1TVZM, B1K7ZT, B29XTJ, B1GQDJ, B288GN, B1GQCH, B1TVR0, B1M9P7, B22G69, B31LJG, B21TDJ, B288M7, B1VPFX, B20XMH, B20XXF, B20Y2V, F0BBT3, B1FB7W, B1Z4N2, B1WFNR, B1VK36, B2T9KF, B1VQBN, B1C236, B1P9RK, B1P9Y6, B2532Y, B1P9Y4, B29XT9, B2DWDP, B34JDQ, B1Q3CF, B1QBF2, B1QBFH, B251P3, B287ZY, B3GLL2, B3KQZ1, B1K6ZJ, F05Z2J, B1PL01, B1RYK9, B34JF3, B3GLN6, B1FWND, B1FRVN, B1FRTC, B26Q3D, B1TJHL, B1TJHH, B2285Q, B2YV8X, B3HBRR, B1VPFT, B3D0VN, B3KLZB, B1FVY4, B1V436, B1QBL8, B2FY37, B1TKYW, F0GFF6, B1LDG1, B1VK1W, B253F0, B1XD5K, B1YCK6, B31XC2, B2888R, B1TVWM, B1YCKG, B22MJV, B2851C, B285M4, B20XMQ, B20W0R, B216LF, B22GX6, B2FXV6, B30QF6, B22MTB, B227HT, B1XN8W, B216LK, B22230, B22M7P, B20Y2T, B22MQ7, B227W2, B20VTM, B3GLRV, B3H6TC, B251P2, F05VJX, B25GG2, B2YTQF, B25FHW, B1K2MH, B25FX1, B29XVL, B288HR, B269DG, B1KZHX, B28MR9, B3VJ5Y, B269RN, B26B6V, F07RRZ, F0G6DN, B1M9P4, B1MHFX, B1MK1K, B1LHBZ.B1M2TR, B289HP, B28BLL, B28518, B26BM4, B39DM7, B2FXQR, F0FCL9, B2854H, F01B41, F03NWY, B2WXZM, B285ZB, B29XPN, B26PD9, B1DX8X, B2859M, B26B0J, B29XP2, B288H6, B2YV70, B2P4CK, B29XRN, B287XK, B288GW, B28825, B39DFZ, B3C671, B1K2J9, B1ZRFW, B31X66, F03V0K, B2FXMZ, B288FH, B30QMX, F08WF2, B287ZZ, B29XPT, B288BW, F04PHH, F04RJQ, B351LT, B288LP, B288HF, B288BL, B3QKK3, B3PKBY, B2P44X, B1P4CW, B1P4P5, B288PP, B374T2, B1C239, B2W90J, B2T83K, B2P4PY, B39YMW, B2Y0QW, B29XT8, B30Q9F, F0B3VD, B2YTXK, B3H17M, B2YH5F, B30QB2, B338J1, B2YGJ9, B1RYK6, B2YH31, B2YV97, B30Q8N, B2YGYR, B31XF5, F0FJJN, B3DGJV, B30QF4, B30QFB, B2YGWY, B2880N, B2YHHJ, B31LG0, F01NYP, F01TBZ, B1P4FT, B3D86W, B3F3NV, B29XT0, F01QRM, F01QX9, F01WRF, F02766, B28N25, F0K3KX, B31X7W, F041BV, B2FXJW, B3DGHY, B37WF1, B39DQT, F09XWB, B3B4RL, B3B4V1, B26B6L, B1MKLM, B3GLG8, F04W5K, B3PKBT, B3G4BH, B3H9DN, B3H9DX, B3HBW8, B3HBWC, B3MPP3, B1PL0C, F08132, F02Y0N, F09DW6, B2C110, B3D83X, F0043R, B3VDHX, B3QKJ0, F01471, F0236N, F00PJQ, B2YHFY, F00LQB, F00LQC, B338K1, F00KXF, F06L9K, B20Y7W, F00WPF, B31X6R, F00YWG, F00ZLC, F010NT, B1MQGB, F01H35, B1GFL3, B1QBBD, F027PD, F0FH3M, B288KQ, F0BB7G, B1LD0P, F03MXT, F0G7BD, B1LZQ2, B23MXR, F04QMK, B29XNN, B20Y9N, F03Y1Z, B3JWXK, B1XD5G, F09YL4, F0BK4T, B2FXZZ, B22P8H, F09D5Y, B1RY9P, B1PJ5J, B1NYL7, B30Q7D, B3H5ZX, F0B52R, F00P43, F0B52T, B1WGC7, F0F304, B3MYL5, B1M9P5, B28MRC, B338DW, B1GPKV, B1FDW1, B1QTP3, B1KZJW, B2YH1Q, B30Q8P, B34J72, B22P8F, B20VTW, B2YHHH, B31X41, B1P4H9, B1RY9T, F0FVXL, B3KYV4, F029X6, B1LZX9, F0K3KZ, B29XRX, F0FJJR, F0831N, F095YL, B3VJ10, F0BHDM, F0BZGT, F07Z1M, F08HRR, F04L4H, F092R2, B2P47J.
Devices named in this recall
5 devices in the FDA device registry whose identifier appears in the recall's code information.
- Transducer L15-7io 989605387852Philips Ultrasound LLC · Transducer, ultrasonic, diagnostic5 recalls0 reports
- Transducer C9-2 989605391092Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic5 recalls0 reports
- Transducer C8-5 989605391922Philips Ultrasound LLC · System, imaging, pulsed doppler, ultrasonic2 recalls0 reports
- Transducer C8-5 989605352343Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic1 recall0 reports
- Transducer C8-5 989605440822Philips Ultrasound LLC · System, imaging, pulsed echo, ultrasonic1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97217).
- Z-2347-2025 · Open, Classified3D9-3v Transducer Probe
- Z-2349-2025 · Open, ClassifiedBP10-5ec Transducer Probe
- Z-2350-2025 · Open, ClassifiedC10-3v Transducer Probe
- Z-2351-2025 · Open, ClassifiedC10-4ec Transducer Probe
- Z-2352-2025 · Open, Classifiedc5-1 Transducer Probe
- Z-2353-2025 · Open, ClassifiedC5-2 Transducer Probe
- Z-2354-2025 · Open, ClassifiedC5-2 Lumify Transducer Probe
- Z-2355-2025 · Open, ClassifiedC6-2 Transducer Probe
- Z-2356-2025 · Open, ClassifiedC6-3 Transducer Probe
- Z-2357-2025 · Open, ClassifiedC8-4v Transducer Probe
- Z-2358-2025 · Open, ClassifiedC8-4v Transducer Probe
- Z-2360-2025 · Open, ClassifiedC9-2 Transducer Probe
- Z-2361-2025 · Open, ClassifiedC9-3io Transducer Probe
- Z-2362-2025 · Open, ClassifiedC9-3v Transducer Probe
- Z-2363-2025 · Open, ClassifiedC9-4 Transducer Probe
- Z-2364-2025 · Open, ClassifiedC9-4v Transducer Probe
- Z-2365-2025 · Open, ClassifiedC9-5ec Transducer Probe
- Z-2366-2025 · Open, ClassifiedD2cwc Transducer Probe
- Z-2367-2025 · Open, ClassifiedD2TCD Transducer Probe
- Z-2368-2025 · Open, ClassifiedD5cwc Transducer Probe
- Z-2369-2025 · Open, ClassifiedeL18-4 Transducer Probe
- Z-2370-2025 · Open, ClassifiedL10-4LAP Transducer Probe
- Z-2371-2025 · Open, ClassifiedL11-3 Transducer Probe
- Z-2372-2025 · Open, ClassifiedL12-3 Transducer Probe
- Z-2373-2025 · Open, ClassifiedL12-4 Transducer Probe
- Z-2374-2025 · Open, ClassifiedL12-4 Lumify Transducer Probe
- Z-2375-2025 · Open, ClassifiedL12-5 38mm Transducer Probe
- Z-2376-2025 · Open, ClassifiedL12-5 50 Transducer Probe
- Z-2377-2025 · Open, ClassifiedL15-7io Transducer Probe
- Z-2379-2025 · Open, ClassifiedL17-5 Transducer Probe
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page