RefleXion X1 Model RXM1000 recallZ-2486-2025 · Reflexion Medical, Inc. · initiated Jul 22, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Reflexion Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
FDA root cause category: Software change control
What the company is doing
On 07/22/2025, the firm sent an "Urgent Medical Device Recall Safety Notification (Correction)" Letter to customer informing them that RefleXion is initiating a Recall (Correction) of the RefleXion X1 System (model RXM1000). An issue has been identified with a configuration setting for the maximum number of fractions that can be delivered within a specified time period. The system should be configured to allow the delivery of up to two (2) fractions, including completion of partial fractions, for each treatment plan within a 12-hour period. As a result of this issue there is a potential to unintentionally administer more than two treatments in a 12-hour period, potentially leading to serious toxicity. Customer are instructed to: 1) Acknowledge receipt of the email by signing and returning the attached Consignee Acknowledgment Form at their earliest convenience. 2) Until the configuration setting is corrected, RefleXion Medical advises users to employ current radiation therapy procedures within the clinic to perform daily checks of treatment session records for radiation therapy patients to prevent delivery of unintended fractions. 3) It is requested that customers post this notification on or near the RefleXion Medical RXM1000 system until this issue can be resolved. For questions or assistance - email [email protected]
Recall details
- Recall number
- Z-2486-2025
- Status
- Open, Classified
- Initiated by the company
- Jul 22, 2025
- Posted by the FDA
- Sep 2, 2025
- Recalling firm
- Reflexion Medical, Inc.
- Quantity
- 11
- Product code
- QVA
- Distribution
- U.S. Nationwide distribution in the states of CA, CT, LA, NJ, OR, PA, and TX.
Product description
RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.
Code information (models, lots, serial numbers)
Model Number: RXM1000 UDI Codes that includes Serial Numbers: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RefleXion Medical Radiotherapy System RXM1000Reflexion Medical, Inc. · Fludeoxyglucose F18-Guided Radiation Therapy System3 recalls36 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page