Spectrum IQ Infusion Pump, Product Code 3570009 recallZ-2415-2025 · Baxter Healthcare Corporation · initiated Jul 14, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Customers were contacted via telephone beginning 7/14/25. The included information about the product, the issue and the action to be taken by the customers. Baxter Healthcare is requesting the return of the impacted product to perform all tests required and confirm that the product perform as expected. If you have any questions, phone (Toll Free): 800-422-9837 or 847-948-4770; hours of operation: Monday through Friday from 8:00 am to 5:00 pm CST

Recall details

Recall number
Z-2415-2025
Status
Open, Classified
Initiated by the company
Jul 14, 2025
Posted by the FDA
Aug 22, 2025
Recalling firm
Baxter Healthcare Corporation
Quantity
6 units
Product code
FRN
FDA submissions
K230041
Distribution
US distribution to states of: AZ, FL

Product description

Spectrum IQ Infusion Pump, Product Code 3570009

Code information (models, lots, serial numbers)

UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates