IntelliSpace Cardiovascular recallZ-2543-2025 · Philips Medical Systems Nederland B.V. · initiated Aug 8, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software issue that results in the display of outdated information.
FDA root cause category: Software design
What the company is doing
An URGENT Medical Device Correction notice dated 8/7/25 was mailed to consignees. The notification includes instructions for continued safe use of devices, which includes: 1. Ensure all finding codes with variables/modifiers are populated correctly on the report screen. Do not leave any FitB Modifier Code Fields Blank. 2. Verify report statements match the selections in the work area before finalizing report. Systems may continue to be used in accordance with the intended use and per the additional instructions. The provided recall notification is to be shared with all users of the device and retained with affected systems until the corrected software is installed. The provided recall response form should be returned within 30 days to [email protected]. A Philips representative will call consignees to schedule a time to install the software solution (FCO83000218). Consignees with any questions can contact Philips Informatics Customer Support at 1-800-996-1328 (x1).
Recall details
- Recall number
- Z-2543-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 8, 2025
- Posted by the FDA
- Sep 9, 2025
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Quantity
- 4 units
- Product code
- LLZ
- FDA submissions
- K153022
- Distribution
- US Nationwide distribution in the states of GA, NC & TX.
Product description
IntelliSpace Cardiovascular, Software 8.0.0.4.
Code information (models, lots, serial numbers)
Model No. 830089; UDI: (01)00884838122000(11)240325(10)8.0.0.0; Serial No. 2210, 2050, 463, 193.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- IntelliSpace Cardiovascular 8Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls132 reports
Product lines
Other devices under the same FDA submission
8 devices cleared under K153022 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- IntelliSpace Cardiovascular 7Philips Medical Systems Nederland B.V. · System, Image Processing, Radiological2 recalls99 reports
- IntelliSpace Cardiovascular 6Philips Medical Systems Nederland B.V. · System, image processing, radiological1 recall118 reports
- IntelliSpace Cardiovascular 5Philips Medical Systems Nederland B.V. · System, image processing, radiological52 reports
- IntelliSpace Cardiovascular 4.2Philips Medical Systems Nederland B.V. · System, image processing, radiological17 reports
- IntelliSpace Cardiovascular IntelliSpace Cardiovascular 4.1Philips Medical Systems Nederland B.V. · System, image processing, radiological8 reports
- IntelliSpace Cardiovascular IntelliSpace Cardiovascular 3.1Philips Medical Systems Nederland B.V. · System, image processing, radiological5 reports
- IntelliSpace Cardiovascular 3.2Philips Medical Systems Nederland B.V. · System, image processing, radiological5 reports
- IntelliSpace Cardiovascular 2.3Philips Medical Systems Nederland B.V. · System, image processing, radiological1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page