NephroMax Kit recallZ-2535-2025 · Boston Scientific Corporation · initiated Aug 5, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Encore 26 Inflation Device is used with balloon dilation catheters to create and monitor pressure in the balloon and to deflate the balloon. An internal investigation determined that during use, foreign material particles could migrate from affected Encore 26 Inflation Devices into a balloon dilatation catheter.
FDA root cause category: Under Investigation by firm
What the company is doing
Boston Scientific issued an Urgent Medical Device Removal notice to it consignees on 8/5/2025 via letter delivered using a traceable means. The notice explained the issue. potential adverse health consequences associated with use in each device and requested the following instructions be followed: 1. Further distribution or use of any remaining devices affected by this removal should cease immediately. Do NOT use affected devices and remove those devices from your facility s inventory. Segregate the units in a secure place until they can be returned to Boston Scientific. 2. Immediately post this information in a visible location near the affected devices to ensure it is readily accessible to all handlers and users of the device. 3. Forward this notice to any healthcare professional from your organization and to any facilities where affected devices have been transferred, including hospitals or sites within your network. If you are a distributor, this notice must be forwarded to your customers to ensure notification of this device removal is carried out to the end-user level. 4. Complete and return the enclosed Reply Verification Tracking Form per the enclosed instructions on page four of the notice 5. Return affected devices. The entire Encore 26 Advantage Kit, NephroMax Kit, or UroMax Ultra Kit must be returned in order to be reimbursed. For questions, please contact your local sales representative.
Recall details
- Recall number
- Z-2535-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 5, 2025
- Posted by the FDA
- Sep 5, 2025
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 458 units
- Product code
- LJE
- FDA submissions
- K121614
- Distribution
- Worldwide distribution.
Product description
Boston Scientific NephroMax Kit, Material Numbers (UPN): 1. M0062101180; 2. M0062101600. (Urology)
Code information (models, lots, serial numbers)
1. M0062101180, GTIN 08714729077589, Lot Numbers: 35275630, 35603649, 35618255, 35644223, 35768091; 2. M0062101600, GTIN 08714729834540, Lot Numbers: 35268077, 35294478, 35547681, 35576896, 35627459.
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- NephroMax Kit M0062101180Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall25 reports
- NephroMax Kit M0062101600Boston Scientific Corporation · CATHETER, NEPHROSTOMY1 recall11 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97396).
- Z-2533-2025 · Open, ClassifiedBoston Scientific Encore 26 Inflation Device, Material Numbers (UPN): 1. H74904526011; 2. H74904526052; 3. M0067101140; 4. M001151050; 5. M001151062; 6. M00566670. (Interve…
- Z-2534-2025 · Open, ClassifiedBoston Scientific Encore Advantage Kit, Material Numbers (UPN): 1. H74904527011; 2. H74904527052. (Interventional Cardiology and Peripheral Interventions)
- Z-2536-2025 · Open, ClassifiedBoston Scientific UroMax Ultra Kit, Material Numbers (UPN): 1. M0062251200, 2. M0062251210, 3. M0062251220, 4. M0062251230, 5. M0062251240, 6. M0062251260, 7. M0062251290, …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page