Skip to content

t:slim X2 insulin pump with Interoperable Technology… recallZ-0232-2026 · Tandem Diabetes Care, Inc. · initiated Aug 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Tandem Diabetes Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

FDA root cause category: Software change control

What the company is doing

On 8/5/2025, Field Safety Notices were sent to customers informing them of the following: Actions to avoid the issue: - Wait until your G7 sensor session ends before inserting a new G7 sensor. - However, if you do insert a new G7 sensor before your last sensor session ends, wait to pair until at least 20 minutes after the last session ends. Firm is working on a software update to resolve this issue. If you have questions or need help, please call firm's Customer Support 1-877-801-6901 or email [email protected]. Our team is available 24/7/365. On 9/16/2025, correction notices were emailed to customers who were informed the firm updated user guides to align with instructions in the correction notice. Also, the firm will notify impacted users via an updated Field Correction Notice when the software update is available. The notice will provide instructions on how to remotely update their software.

Recall details

Recall number
Z-0232-2026
Status
Open, Classified
Initiated by the company
Aug 5, 2025
Posted by the FDA
Oct 23, 2025
Recalling firm
Tandem Diabetes Care, Inc.
Quantity
122,958
Product code
QFG
FDA submissions
K201214, K203234, K240309, K241078
Distribution
US nationwide distribution including Puerto Rico.

Product description

t:slim X2 insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014070, 1014071, 1014072, 1014073, and User Guide when using Dexcom G7 sensor; Tandem Mobi insulin pump with Interoperable Technology, Control-IQ+ technology, REF: 1014801, 1014811, 1014813, and User Guide when using Dexcom G7 sensor

Code information (models, lots, serial numbers)

t:slim X2, UDI-DI: 00389152407012, 00389152407111, 00389152407210, 00389152407319 with software versions 7.9.0.1 and 7.10.1, User Guide versions not including and prior to: AW-1014562 Revision D; Tandem Mobi, UDI-DI: 00389152480114, 00389152481111, 00389152481319 with software version 7.9.0.1, User Guide versions not including and prior to: AW-1014938 Revision C

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K201214, K203234, K240309, K241078 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.