Tandem Mobi, pump, Control-IQ 7.9 1014801Recalls & reported problems

Tandem Diabetes Care, Inc. · Tandem Mobi, pump, Control-IQ 7.x · Class 2

All models of Tandem Mobi (7)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Tandem Mobi insulin pump is worn to deliver insulin. Reports mostly describe pumps that stop, blocked flow and occlusion alarms, air in the device, and charging failures. There are 7,413 reports, including 1 death and 1,389 injuries. Two recalls are open, covering software defects that could cause unexpected insulin doses and incorrect glucose trend readings. Reports rose to 5,843 in the last 12 months from 1,570 in the 12 months before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
7,327
Deaths reported
1
Injuries reported
1,405
Recalls naming it
2

Summary

The FDA's MAUDE database contains 7,327 adverse event reports that list this exact device (DI 00389152480114), received between May 5, 2025 and Sep 30, 2026 — classified as 1 death, 1,405 injuries, 5,921 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (3,028 reports).

2 FDA recall records name this device's identifier, 2 still open.

1 other recall cover devices cleared under the same FDA submission (K241078); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 84 reportsJun 2025: 157 reportsJul 2025: 333 reportsAug 2025: 351 reportsSep 2025: 268 reportsOct 2025: 377 reportsNov 2025: 457 reportsDec 2025: 661 reportsJan 2026: 688 reportsFeb 2026: 678 reportsMar 2026: 868 reportsApr 2026: 791 reportsMay 2026: 673 reportsJun 2026: 484 reportsJul 2026: 179 reportsAug 2026: 122 reportsSep 2026: 156 reportsOct 2026: 0 reports202420252026868
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

    Z-0232-2026 · Root cause: Software change control · Recall details · FDA record ↗

  2. Feb 27, 2025Class IOpen, Classified

    A software defect in Version 7.9 of the pump software for Tandem t:slim X2 and Tandem Mobi pumps, when used with Control IQ+ technology, will cause the pump to incorrectly interpolate glucose trends when the Estimated Glucose Value (EGV) is above 255 mg/dL at the start or end of a gap in data collection due to a lapse in connection from a paired continuous glucose monitor (CGM) sensor, which can lead to under-delivery or over-delivery of insulin based on inaccurate result leading to severe cases of hypoglycemia or hyperglycemia.

    Z-1520-2025 · Root cause: Software Design Change · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped3,028
  • Obstruction of Flow1,644
  • Adverse Event Without Identified Device or Use Problem609
  • Mechanical Problem386
  • Air/Gas in Device296
  • Failure to Charge269
  • Priming Problem132
  • Use of Device Problem131

Patient problems

  • No Clinical Signs, Symptoms or Conditions5,926
  • Hyperglycemia1,107
  • Hypoglycemia282
  • Elevated Ketones/ Diabetic Ketoacidosis79
  • Loss of consciousness15
  • Vomiting8
  • Insufficient Information6
  • Convulsion/Seizure5

Recent adverse event reports

The 25 most recent of 7,327 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50954998 ↗
  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Hypoglycemia

    MAUDE report 50954604 ↗
  • MalfunctionSep 30, 2026Mechanical Problem

    It was reported that a cartridge alarm occurred during the load sequence. Customer reverted to an alternate method of insulin therapy. There was no adverse impact to the customer¿s blood glucose level.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951724 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951128 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951120 ↗
  • MalfunctionSep 30, 2026Failure to Charge

    It was reported that Tandem Mobi battery would charge up to 95% but would not reach 100%. There was no adverse impact to the customer¿s blood glucose level. Multiple attempts were made by Tandem Technical Support to follow-up with the customer on the reported issue, but no response was received.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950707 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. The customer¿s blood glucose (BG) level was 212 mg/dL. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50950704 ↗
  • MalfunctionSep 30, 2026Break

    It was reported that cartridge was damaged on two occasions, both occurring on same day. There was no adverse impact to the customer¿s blood glucose level. Customer¿s mother planned to call back after son left school to continue troubleshooting, and no additional information was received regarding the outcome of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949865 ↗
  • MalfunctionSep 30, 2026Data Problem

    It was reported that graph in mobile app previously did not show readings in upper left corner after app reconnected with pump, although issue was not occurring at time of call and no error messages were seen. There was no adverse impact to the customer¿s blood glucose level. Customer was informed that app could take several minutes to reload pump information after connection was restored and was instructed to call back if pump information did not load within one hour, and no additional information was received regarding the outcome of the event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949843 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949791 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Technical Support provided information to reset the pump, assisted the caller with the reset, and confirmed that the malfunction code did not recur. The customer was advised to continue using the pump for insulin therapy.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50949478 ↗
  • MalfunctionSep 30, 2026Pumping Stopped

    A malfunction was reported on the pump. There was no reported adverse impact to the customer's blood glucose level. Multiple contact attempts were made by Tandem Technical Support to obtain additional information regarding the issue; however, the customer did not respond. No additional patient or event information was available.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50948011 ↗
  • MalfunctionSep 30, 2026Failure to Charge

    It was reported that Tandem Mobi pump battery would not charge despite use of charging pad, cable, wall adapter, and confirmed working outlet, and pump battery had already been depleted. There was no adverse impact to the customer¿s blood glucose level, as customer relied on multiple daily injections for insulin since issue occurred. Customer tried different charging cable and charging pad without success, CTS arranged replacement charging supplies via two-day shipping, attempted multiple follow-up calls to confirm resolution, and ultimately closed case issue when no further contact was received from customer.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50947700 ↗
  • MalfunctionSep 30, 2026Failure to Charge

    It was reported that Tandem Mobi pump battery would not charge even though customer used standard charging pad, cable, wall adapter, and confirmed wall outlet was working, and initial attempts including leaving pump on charging pad for 30 minutes did not resolve issue. There was no adverse impact to the customer¿s blood glucose level because pump was not actively set up. Customer tried different charging cable without success, then tried different charging pad which allowed pump to begin charging; customer ultimately continued using original charging supplies, and no further assistance was required.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946938 ↗
  • MalfunctionSep 30, 2026Failure to Charge

    It was reported that the pump battery could not be charged. There was no adverse impact to the customer's blood glucose. Reportedly, the pump successfully began charging after leaving the pump plugged into the power source.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946683 ↗
  • MalfunctionSep 30, 2026Obstruction of Flow

    It was reported that customer experienced intermittent occlusion alarms. There was no adverse impact to the customer¿s blood glucose level, and reported values did not exceed 500 mg/dL. Customer changed soft cannula infusion set and cartridge, resumed insulin therapy, and no issues were observed with infusion set; no additional information was received regarding any further outcome of event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946646 ↗
  • MalfunctionSep 30, 2026Obstruction of Flow

    It was reported that customer experienced an occlusion alarm. There was no adverse impact to the customer¿s blood glucose level. Customer resolved issue by changing infusion set and cartridge, resuming insulin, and no additional assistance was needed, so Technical Support planned to close case.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50946644 ↗
  • InjurySep 30, 2026Pumping Stopped

    It was reported that pump displayed malfunction alarm Malf 24. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. Customer gave correction insulin injection using manual injections, used manual injections as backup therapy, and did not require help from others. Technical Support provided instructions for storage mode and pump return, advised customer to enter pump settings and reconnect continuous glucose monitor to replacement pump, and arranged shipment of replacement pump under warranty.

    Patient problems: Hyperglycemia

    MAUDE report 50945064 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941260 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941259 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941249 ↗
  • MalfunctionSep 29, 2026Mechanical Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941135 ↗
  • InjurySep 29, 2026Pumping Stopped

    Patient problems: Hyperglycemia

    MAUDE report 50941125 ↗
  • MalfunctionSep 29, 2026Obstruction of Flow

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941120 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50941114 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K241078. One submission often covers many models, so check the product and lot details in each record.

  • Z-0427-2026 · Oct 6, 2025 · Open, Classified
    Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a mo

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →