t:slim X2 Insulin Pump with Control-IQ+ Technology 1014073Recalls & reported problems

Tandem Diabetes Care, Inc. · Control-IQ+ 7.x, Refurbished · Class 2

All models of t:slim X2 Insulin Pump (45) · t:slim X2 Insulin Pump with Control-IQ Technology (11)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Tandem t:slim X2 insulin pump has 15,187 adverse event reports, including 2 deaths and 2,306 injuries. Many describe the pump stopping, failing to charge or blocking flow, and reports often mention high or low blood sugar. There is 1 open recall, from August 2025, over a software issue that could cause an unexpected automatic insulin dose. Reports rose from 4,438 in the prior 12 months to 10,749 in the latest 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
14,723
Deaths reported
3
Injuries reported
2,322
Recalls naming it
1

Summary

The FDA's MAUDE database contains 14,723 adverse event reports that list this exact device (DI 00389152407319), received between Mar 27, 2025 and Sep 30, 2026 — classified as 3 deaths, 2,322 injuries, 12,398 malfunctions.

The most frequently reported specific device problem is “Pumping Stopped” (3,336 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 9 reportsApr 2025: 148 reportsMay 2025: 341 reportsJun 2025: 354 reportsJul 2025: 980 reportsAug 2025: 1,079 reportsSep 2025: 790 reportsOct 2025: 737 reportsNov 2025: 916 reportsDec 2025: 1,262 reportsJan 2026: 1,307 reportsFeb 2026: 1,234 reportsMar 2026: 1,247 reportsApr 2026: 1,259 reportsMay 2026: 1,038 reportsJun 2026: 1,154 reportsJul 2026: 477 reportsAug 2026: 306 reportsSep 2026: 85 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520261,307
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Aug 5, 2025Class IIOpen, Classified

    A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an unexpected automatic insulin correction bolus (autobolus ) which could result in hypoglycemia.

    Z-0232-2026 · Root cause: Software change control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Pumping Stopped3,336
  • Failure to Charge1,904
  • Incorrect Measurement1,365
  • Charging Problem1,264
  • Obstruction of Flow1,124
  • Adverse Event Without Identified Device or Use Problem1,001
  • Key or Button Unresponsive/not Working661
  • Premature Discharge of Battery632

Patient problems

  • No Clinical Signs, Symptoms or Conditions12,392
  • Hyperglycemia1,636
  • Hypoglycemia679
  • Elevated Ketones/ Diabetic Ketoacidosis92
  • Loss of consciousness35
  • Insufficient Information24
  • Fall7
  • Vomiting7

Recent adverse event reports

The 25 most recent of 14,723 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026False Alarm

    It was reported that ongoing altitude alarms occurred. Reportedly, the pump was not outside the labeled altitude operating range. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer reloaded the cartridge and delivered a correction bolus via the pump to address the BG. Customer reverted to an alternate method of insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50954870 ↗
  • InjurySep 30, 2026False Alarm

    It was reported that an altitude alarm occurred while the customer was not outside of labeled operating altitude range. Reportedly, an obstruction was blocking the speaker holes. The customer experienced an elevated blood glucose (BG) level of 586 mg/dL. Customer took no action to address the BG. After removing the obstruction from the speaker holes, the alarm cleared.

    Patient problems: Hyperglycemia

    MAUDE report 50954747 ↗
  • InjurySep 30, 2026Pumping Stopped

    It was reported that pump experienced malfunction alarm, with no events occurring before issue. Customer¿s blood glucose level was reported as ¿High,¿ with specific value unknown. Customer gave correction injection using multiple daily injections and did not require emergency help, switched to multiple daily injections as backup insulin therapy, and Technical Support arranged replacement pump under warranty, confirmed shipping and storage instructions, and instructed customer to return malfunctioning pump and set up replacement pump and connected glucose sensor; no additional information was received regarding the final outcome of event.

    Patient problems: Hyperglycemia

    MAUDE report 50952677 ↗
  • InjurySep 30, 2026Unexpected Shutdown

    It was reported that the pump shut down unexpectedly on an unspecified date in (b)(6) 2026. The customer experienced an elevated blood glucose (BG) level that was over 600 mg/dL with an unspecified level of ketones that was not identified as life threatening by the healthcare provider. The customer first went to the Emergency Room and was subsequently hospitalized in the Intensive Care Unit (ICU) where they were treated with intravenous fluids of saline and insulin. Treatment resolved the issue with no permanent damage to the customer. Customer was released from the hospital after 2 days. Customer did not use the pump until the time of troubleshooting and Tandem Technical Support assisted customer with turning the pump back on and resuming insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50951426 ↗
  • InjurySep 29, 2026Pumping Stopped

    It was reported that an altitude alarm occurred. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer delivered a correction bolus via the pump to address the BG. Reportedly a new cartridge was loaded, and insulin delivery was resumed.

    Patient problems: Hyperglycemia

    MAUDE report 50941313 ↗
  • InjurySep 29, 2026Pumping Stopped

    It was reported that an altitude alarm occurred. The customer experienced an elevated blood glucose (BG) level displayed as \"High\"; although a specific BG value was not provided. Customer administered a correction injection to address the BG. Reportedly a new cartridge was loaded, and insulin delivery was resumed.

    Patient problems: Hyperglycemia

    MAUDE report 50941298 ↗
  • MalfunctionSep 29, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50940988 ↗
  • InjurySep 29, 2026Obstruction of Flow

    It was reported that customer experienced multiple occlusion alarms, including Alarm 2 followed by Alarm 26, while using infusion set and cartridge. Customer¿s blood glucose (BG) level was displayed as \"High\"; however, specific value was not provided. Customer gave correction insulin by a manual injection, changed infusion set and cartridge, resumed insulin delivery, and Technical Support provided assistance and later closed case when no further help was needed.

    Patient problems: Hyperglycemia

    MAUDE report 50939685 ↗
  • MalfunctionSep 29, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50939016 ↗
  • InjurySep 28, 2026Obstruction of Flow

    It was reported that customer experienced occlusion alarms, and Technical Support confirmed alarm indicating delivery blockage over several days. Customer¿s blood glucose level was reported as ¿High,¿ with specific value unknown. Customer gave correction dose using pump, resumed insulin use, did not need help from any third party, and Technical Support closed case with no additional information received regarding further outcome of event.

    Patient problems: Hyperglycemia

    MAUDE report 50925052 ↗
  • InjurySep 28, 2026Pumping Stopped

    It was reported that pump experienced malfunction with speaker alarm and displayed malfunction code that could not be reset, with no notable events occurring before the issue. Customer¿s blood glucose level was reported as ¿High¿ and exceeded 500 mg/dL, but specific value was unknown. Customer administered correction injection using multiple daily injections, planned to use multiple daily injections as backup insulin therapy, did not require emergency assistance, and Technical Support arranged pump replacement and discussed out-of-warranty loaner pump option.

    Patient problems: Hyperglycemia

    MAUDE report 50924164 ↗
  • MalfunctionSep 28, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50924141 ↗
  • InjurySep 28, 2026Pumping Stopped

    It was reported that pump experienced a malfunction alarm related to speaker while no notable events occurred prior to the issue. The customer¿s blood glucose (BG) level was displayed as ¿High¿; however, a specific value was not provided. At the time of the report the customer had not yet taken action to address the BG level. Customer planned to use multiple daily injections as backup therapy, CTS arranged replacement pump and confirmed shipping details, and customer was instructed to place malfunctioning pump in storage mode and return it within 10 days, then program pump settings and reconnect CGM to replacement pump, which customer understood and acknowledged.

    Patient problems: Hyperglycemia

    MAUDE report 50923611 ↗
  • MalfunctionSep 28, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50923109 ↗
  • MalfunctionSep 25, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904812 ↗
  • MalfunctionSep 25, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50904769 ↗
  • InjurySep 25, 2026Insufficient Device Problem Information

    It was reported that the pump was requiring customer to repeat the load fill tubing during the load sequence. Customer¿s blood glucose (BG) level was displayed as \"High\"; however, specific value was not provided. Elevated BG was addressed by administering a correction bolus. The customer reloaded the cartridge and resumed insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50903536 ↗
  • InjurySep 25, 2026Pumping Stopped

    It was reported that pump displayed a malfunction alarm related to touchscreen and showed a malfunction code, and customer¿s blood glucose reading displayed ¿High¿ with a value above 500 mg/dL. Customer¿s blood glucose level exceeded 500 mg/dL, but specific value was not known. Customer gave correction insulin by injection and did not need help from emergency services, and Technical Support provided pump reset information, assisted customer with resetting malfunction alarm, confirmed alarm did not recur, advised customer to continue using pump, and instructed customer to call back if malfunction appeared again.

    Patient problems: Hyperglycemia

    MAUDE report 50902829 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer experienced a blood glucose (BG) level that was displayed as ¿High¿; however, a specific value was not provided. Reportedly, the customer alleged the pump software did not accurately adjust the amount of insulin delivered resulting in the high BG. Customer gave a correction bolus via the pump to address BG level. Tandem Technical Support (CTS) performed a system check and performed a review of the pump data to determine a possible cause. CTS determined that the pump functioned as intended and the cause of the high BG was unknown. Multiple contact attempts were made by Tandem Technical Support to obtain additional information regarding the issue; however, the customer did not respond. No additional patient or event information was available.

    Patient problems: Hyperglycemia

    MAUDE report 50890463 ↗
  • DeathSep 24, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Elevated Ketones/ Diabetic Ketoacidosis

    MAUDE report 50890170 ↗
  • MalfunctionSep 24, 2026Pumping Stopped

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50889177 ↗
  • InjurySep 24, 2026Use of Device Problem

    It was reported that the customer received an incomplete bolus alert as they had not completed a bolus request. The customer then experienced an elevated blood glucose level of 513 mg/dL. The customer administered a bolus via the pump to address the BG and continued to use the pump for insulin therapy. Tandem technical support educated the customer on the meaning of an incomplete bolus alert.

    Patient problems: Hyperglycemia

    MAUDE report 50888872 ↗
  • InjurySep 22, 2026Obstruction of Flow

    It was reported that customer experienced recurring occlusion alarms while using pump, including verified Alarm 2, which led to very high blood sugar. Customer¿s blood glucose level reached 529 mg/dL due to the issue. Customer delivered correction dose using pump, then changed infusion set and cartridge, resumed insulin, did not require assistance from any third party, and Customer Tech Support (CTS) closed case with no additional support needed.

    Patient problems: Hyperglycemia

    MAUDE report 50862642 ↗
  • InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the customer experienced a blood glucose (BG) level that was displayed as ¿High¿; however, a specific value was not provided, cause was unknown. Customer gave a correction bolus via the pump to address BG level. Reportedly, the customer continued to use the pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50862547 ↗
  • InjurySep 22, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that customer experienced a blood glucose (BG) level of 592 mg/dL, cause was unknown. Customer gave a correction bolus via the pump to address BG level. Reportedly, the customer continued to use the pump for insulin therapy.

    Patient problems: Hyperglycemia

    MAUDE report 50860363 ↗

FDA clearance and classification

The FDA regulates product code QFG (alternate controller enabled insulin infusion pump) under 21 CFR 880.5730 as a Class II device. 21 CFR 880.5730 ↗

Devices of this type normally need 510(k) clearance before sale.

FDA definition of this device type: An alternate controller enabled insulin infusion pump (ACE insulin pump) is a device intended for the infusion of insulin into a patient. The ACE insulin pump may include basal and bolus drug delivery at set or variable rates. ACE insulin pumps are designed to reliably and securely communicate with external devices, such as automated insulin dosing systems, to allow insulin delivery commands to be received, executed, and confirmed. ACE insulin pumps are intended to be used both alone and in conjunction with digitally connected medical devices for the purpose of insulin delivery.

Common questions about the t:slim X2 Insulin Pump with Control-IQ+ Technology 1014073

Has the t:slim X2 Insulin Pump with Control-IQ+ Technology 1014073 been recalled?

Yes. 1 FDA recall record names this device's identifier, and it is still open. The recall, Z-0232-2026 (Class II), began on Aug 5, 2025: A software defect in versions 7.9.0.1 and 7.10.1 of the pump software for Tandem t:slim X2 and version 7.9.0.1 of the Tandem Mobi pump with Control-IQ+ technology that are paired with a G7 sensor may result in an… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.

What problems have been reported with the t:slim X2 Insulin Pump with Control-IQ+ Technology 1014073?

The FDA's MAUDE database lists 14,723 adverse event reports for this device, received between Mar 27, 2025 and Sep 30, 2026. In those reports, the most common specific device problems are “Pumping Stopped” (3,336), “Failure to Charge” (1,904) and “Incorrect Measurement” (1,365); the most common patient problems are “Hyperglycemia” (1,636), “Hypoglycemia” (679) and “Elevated Ketones/ Diabetic Ketoacidosis” (92). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the t:slim X2 Insulin Pump with Control-IQ+ Technology 1014073?

3 reports are classified as deaths, 2,322 as injuries and 12,398 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the t:slim X2 Insulin Pump with Control-IQ+ Technology 1014073 MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the t:slim X2 Insulin Pump with Control-IQ+ Technology 1014073, and is it still sold?

The FDA's device database lists Tandem Diabetes Care, Inc. as the company responsible for it (brand t:slim X2 Insulin Pump with Control-IQ+ Technology, model 1014073). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Alternate Controller Enabled Insulin Infusion Pump devices

All Alternate Controller Enabled Insulin Infusion Pump devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page