Skip to content

ZAP-X Radiosurgery System, Model: 300150 recallZ-2639-2025 · Zap Surgical Systems · initiated Jul 25, 2025

Class II Open, Classified All recalls posted in September 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zap Surgical Systems or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

FDA root cause category: Software change control

What the company is doing

On 7/25/2025, correction notices were sent to customers who were asked to do the following: Firm will be revising the software to address this issue. Until we supply you with the revised software, the firm strongly advises that you follow these steps to prevent potential collisions. 1. In the rare event that the system triggers a proximity error message during a long gantry move (greater-than-180-degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, please manually extract the patient. 2. After the patient has been safely extracted, please return the gantry to its home position before resuming any treatment procedures. If you have any further questions related to this situation email [email protected]

Recall details

Recall number
Z-2639-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Jul 25, 2025
Posted by the FDA
Sep 26, 2025
Recalling firm
Zap Surgical Systems
Quantity
17
Product code
IYE
FDA submissions
K211663
Distribution
Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.

Product description

ZAP-X Radiosurgery System, Model: 300150

Code information (models, lots, serial numbers)

UDI-DI: 00860183001504, Serial Numbers: ZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181. Software version: TDS V1.10.1

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates