Zap-X Radiosurgery System 300150Recalls & reported problems

Zap Surgical Systems, Inc. · The self-shielded ZAP-X® Radiosurgery System is a computer-controlled system for planning and delivering non-invasive stereotactic radiosurgery for tumors, lesions and conditions in the brain, head, and neck when radiation treatment is indicated. The system is self-shielded for ionizing radiation, allowing it to be installed in a non-bunker environment. A linear accelerator mounted on a dual-gantry provides a therapeutic radiation source, and a kV imaging system is used to locate and track the treatment target throughout the treatment. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00860183001504), received between Jun 16, 2022 and Jun 16, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Application Program Problem: Parameter Calculation Error” (1 report).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 25, 2025Class IIOpen, Classified

    If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

    Z-2639-2025 · Root cause: Software change control · Recall details · FDA record ↗

  2. Jul 5, 2022Class IIOpen, Classified

    Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

    Z-1742-2022 · Root cause: Software design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Problem: Parameter Calculation Error1

Patient problems

  • Insufficient Information1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 16, 2022Application Program Problem: Parameter Calculation Error; Application Program Problem: Parameter Calculation Error

    DURING TREATMENT, TRACKING ALGORITHM HAD DIFFICULTIES ALIGNING PATIENT.

    Patient problems: Insufficient Information

    MAUDE report 14716282 ↗