Skip to content

i-STAT CG4+ cartridge (white). List Number: 03P85-25 recallZ-2582-2025 · Abbott Point Of Care Inc. · initiated Aug 21, 2025

Class II Open, Classified All recalls posted in September 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Point Of Care Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.

FDA root cause category: No Marketing Application

What the company is doing

Abbott notified consignees via email and letter on about 08/21/2025. Consignees were instructed to download, review and implement the updated i-STAT CG4+ cartridge Instructions for Use (IFU), discontinue use of the i-STAT CG4+ cartridges for testing of capillary whole blood samples for lactate and instead use venous or arterial whole blood or an alternate method for testing capillary whole blood for lactate, and update the pH and PO2 reportable ranges on the i-STAT 1 analyzers running i-STAT CG4+ cartridges. This feature is not customizable through the i-STAT 1 handheld keypad, please see the i-STAT DE Quick Reference Guide - Customizing Reportable Ranges (ART 770547-00 Revision A). To use the Custom Reportable Range feature, you must have the i-STAT 1 handheld and i-STAT DE version 2.3 or higher. Changes to the reportable range may require a change to the Laboratory Information System (LIS) interface. Additionally, consignees were requested to confirm receipt and understanding of this communication by responding to the business reply card included with the letter. If consignees have forwarded any i-STAT CG4+ cartridges to another facility, they were requested to notify them.

Recall details

Recall number
Z-2582-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Aug 21, 2025
Posted by the FDA
Sep 17, 2025
Recalling firm
Abbott Point Of Care Inc.
Quantity
9,067,050 units
Product code
CHL
FDA submissions
K244014, K940918, K982071
Distribution
US Nationwide distribution.

Product description

i-STAT CG4+ cartridge (white). List Number: 03P85-25.

Code information (models, lots, serial numbers)

List Number: 03P85-25. All lot numbers.

Other devices under the same FDA submission

5 devices cleared under K244014, K940918, K982071 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page