CLEARLINK/CONTINU-FLO recallZ-2594-2025 · Baxter Healthcare Corporation · initiated Aug 8, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Baxter Healthcare Corporation is issuing an Urgent Medical Device Recall for one lot of Continu-Flo solution sets listed below due to customer reports of tubing separation.
FDA root cause category: Under Investigation by firm
What the company is doing
Baxter issued an Urgent Medical Device Recall notice was issued its consignees on 8/8/2025 via U.S.P.S., first class mail. The notice explained the issue, potential hazard involved and requested the following: Actions to be Taken by Customers 1. Immediately locate, isolate, and cease all use of the affected lot number of the product. The product code and lot number can be found on the individual product package and shipping carton. 2. Contact Baxter Healthcare Center for Service to arrange for return and replacement product at 888-229-0001 between the hours of 7:00 am and 6:00 pm Central Time, Monday through Friday. Please have your Baxter 8-digit ship-to account number, product code, lot number, and quantity of product to be returned ready when calling. 3. If you received this communication directly from Baxter, please acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please contact them to arrange for return of the affected product. Please note that responding to Baxter is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to the director of nursing, director of pharmacy, director of purchasing, facility risk manager, and any other departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of the affected product that you distribu
Recall details
- Recall number
- Z-2594-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 8, 2025
- Posted by the FDA
- Sep 19, 2025
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 14400 units
- Product code
- FPA
- FDA submissions
- K203609
- Distribution
- US Nationwide distribution.
Product description
Baster Continu-Flo Solution Set, Non-Vented, 2 Clearlink Luer Activated Valves, Backcheck Valve. 10 drops/mL. 112" (2.8 m), Product Code REF 2C8519; intravascular administration set
Code information (models, lots, serial numbers)
UDI/DI 00085412048994, Lot Number R24I20079, Exp. 9/21/2026
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CLEARLINK/CONTINU-FLO 2C8519Baxter Healthcare Corporation · Set, administration, intravascular3 recalls1 death221 reports
Product lines
Other devices under the same FDA submission
18 devices cleared under K203609 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- CLEARLINK/CONTINU-FLO/DUO-VENT 2C8541Baxter Healthcare Corporation · Set, administration, intravascular2 recalls797 reports
- CLEARLINK/CONTINU-FLO 2C8537Baxter International Inc. · Set, administration, intravascular2 recalls490 reports
- CLEARLINK/CONTINU-FLO/DUO-VENT UC8519Baxter International Inc. · Set, administration, intravascular1 recall66 reports
- CLEARLINK/DUO-VENT 2C7462Baxter Healthcare Corporation · Set, administration, intravascular57 reports
- DUO-VENT 1C8507Baxter Healthcare Corporation · Set, administration, intravascular1 recall4 reports
- CLEARLINK 2C7461Baxter Healthcare Corporation · Set, administration, intravascular49 reports
- CLEARLINK/DUO-VENT 2H7462Baxter International Inc. · Set, administration, intravascular47 reports
- NA 1C8109Baxter Healthcare Corporation · Set, administration, intravascular32 reports
- CLEARLINK/CONTINU-FLO 2C8546Baxter Healthcare Corporation · Set, administration, intravascular22 reports
- INTERLINK 2C6425Baxter Healthcare Corporation · Set, administration, intravascular11 reports
- NA 1C8160Baxter Healthcare Corporation · Set, administration, intravascular7 reports
- CLEARLINK/CONTINU-FLO/DUO-VENT ALT2403Baxter Healthcare Corporation · Set, administration, intravascular6 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page