Skip to content

VITROS recallZ-2560-2025 · Ortho-Clinical Diagnostics, Inc. · initiated Jul 29, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

VITROS Chemistry Products LAC Slides on VITROS XT 3400 and VITROS XT 7600 Systems are experiencing an increased occurrence of calibration failure related to lots from coating 0130 and above with Condition Code TH4-63J during calibration, causing the delay of lac test results and potentially impact patient management. The issue was identified by QuidelOrtho through an increase in customer complaints compared to the historical baseline with average failure rate of 1% for coating 0130 in 2025 while average of failure rate for other coating was 0.06% in 2025 and average of failure rate for all coatings is 0.03% for 2024.

FDA root cause category: Device Design

What the company is doing

On 29 July 2025, QuidelOrtho mailed a hardcopy customer letter (Ref. CL2025-201) via FedEx overnight courier and/ or electronically sent the customer letter using ORTHO PLUS e-Communications to customers. Consignees were instructed to load ADD DRV 6339 (or above) on your VITROS XT 3400 and/or VITROS XT 7600 System(s), complete and return the provided Confirmation of Receipt, save the notification with your User Documentation or post this notification by each VITROS XT 3400/VITROS XT 7600 System in your laboratory until the issue has been resolved, and forward the information if affected units were further distributed.

Recall details

Recall number
Z-2560-2025
Status
Open, Classified
Initiated by the company
Jul 29, 2025
Posted by the FDA
Sep 12, 2025
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
2,320 units
Product code
KHP
FDA submissions
K955286
Distribution
US Nationwide distribution.

Product description

VITROS Chemistry Products LAC Slides. Catalog Number: 8150112. 1 Sales Unit = contains 5 cartridges of 18 slides, for a total of 90 tests.

Code information (models, lots, serial numbers)

Catalog number 8150112; UDI# 10758750004454; Coating 0130 and above, this includes Lot numbers 3533-0130-2947, 3533-0130-2972, 3533-0130-3001, 3533-0130-3002, 3533-0130-3048, 3533-0130-3084, 3533-0131-5887, 3533-0131-5958, 3533-0131-5962, 3533-0131-6058 ; Expiration date range 01-Oct-2026 to 01-Jan-2027

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page