UniCel DxH 900 COULTER Cellular Analysis System, configured… recallZ-2572-2025 · Beckman Coulter Inc. · initiated Aug 13, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.
FDA root cause category: Component change control
What the company is doing
On August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System (LIS), Middleware, or on affected instruments. Consult your Medical Director to determine if a retrospective review of results is warranted. Beckman Coulter recommends posting this letter on or near the affected analyzers Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. Acknowledge receipt of the notification.
Recall details
- Recall number
- Z-2572-2025
- Status
- Open, Classified
- Initiated by the company
- Aug 13, 2025
- Posted by the FDA
- Sep 17, 2025
- Recalling firm
- Beckman Coulter Inc.
- Product code
- GKZ
- FDA submissions
- K240252
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.
Product description
UniCel DxH 900 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C11478
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI: 15099590701574/ Serial Numbers: BB33238, BB33239, BB42317, BC05150, BC08198, BC10226, BC16320, BC21392, BC23453, BC25457, BC25465, BC30527, BC32542, BC42721, BC42726, BC49882, BD13219, BD41570, BD46639, BD46643, BD46646, BE01022, BE02044, BE02052, BE04080, BE21222, BE24254, BE24257, BE29334, BE31336, BE33366, BE42455, BE42459, BE46517, BE49539, BE49541, BF10071, BF12126, BF12129, BF18194, BF33377, BF40462, BG04041, BG04042, BG04053, BG05080, BG11182, BG11184, BG11185, BG14195, BG14200, BG14204, BG14208, BG16220, BG16231, BG17238, BG17250, BG17260, BG19283, BG20292, BG20294, BG25395, BG25412, BG25413, BG26435, BG40563, BG45631, BG46635, BH04028, BH04029, BH05051, BH19220, BH24260, BH31328, BD46642, BG22347, BG20307, BD46640, BC32545
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DxH 900 Hematology System C11478Beckman Coulter, Inc. · Counter, Differential Cell1 recall17 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97464).
- Z-2573-2025 · Open, ClassifiedUniCel DxH 690T COULTER Cellular Analysis System, configured with the Led HGB Photometer REF: C34520
- Z-2574-2025 · Open, ClassifiedUniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322
- Z-2575-2025 · Open, ClassifiedUniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page