UniCel DxH recallZ-2575-2025 · Beckman Coulter Inc. · initiated Aug 13, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

FDA root cause category: Component change control

What the company is doing

On August 13, 2025, Beckman Coulter initiated a "Urgent Medical Device Recall" notification via E-Mail. Beckman Coulter expanded the recall to additional consignees September 25, 2025 Beckman Coulter asked consignees to take the following actions: For specimens with a WBC count > 30 x 103 and = 90 x 103 cells/L, follow the actions listed within the customer letter. For specimens with a WBC count > 90 x 103 cells/L, follow the actions listed within the customer letter. To identify WBC counts > 30 x 103 cells/L, consider implementing decision rules within your Laboratory Information System (LIS), Middleware, or on affected instruments. Consult your Medical Director to determine if a retrospective review of results is warranted. Beckman Coulter recommends posting this letter on or near the affected analyzers Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected products listed above to another laboratory, please provide them a copy of this letter. Acknowledge receipt of the notification.

Recall details

Recall number
Z-2575-2025
Status
Open, Classified
Initiated by the company
Aug 13, 2025
Posted by the FDA
Sep 17, 2025
Recalling firm
Beckman Coulter Inc.
Product code
GKZ
FDA submissions
K120771
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Egypt, France, Germany, Ghana, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Italy, Japan, Jordan, Kenya, Korea Republic of, Kuwait, Kyrgyzstan, Libya, Lithuania, Mexico, Nepal, Netherlands, New Zealand, North Macedonia, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Spain, Switzerland, Taiwan, Province of China, Thailand, Trinidad and Tobago, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, United States of America, Uruguay, Vietnam.

Product description

UniCel DxH 600 COULTER Cellular Analysis System, configured with the Led HGB Photometer REF:B23858

Code information (models, lots, serial numbers)

UDI: 15099590192242/ Serial Numbers: AV52143, AZ02628, AZ06676, AZ24836, BA07048, BA08667, BA17727, BA31855, BB10648, BB49881, BC24713

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 97464).