Why it was recalled

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

FDA root cause category: Under Investigation by firm

What the company is doing

On July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.

Recall details

Recall number
Z-2611-2025
Status
Open, Classified
Initiated by the company
Jul 31, 2025
Posted by the FDA
Sep 26, 2025
Recalling firm
Medline Industries, LP
Quantity
7913 eaches
Product code
NLH
FDA submissions
K241224
Distribution
Domestic US distribution nationwide. International distribution pending. No international distribution.

Product description

Reprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of cardiac structures. Medline Item Numbers BD710DF282CRH BD710DF282RRH BD710FJ282CRH BD710FJ282RRH

Code information (models, lots, serial numbers)

Item Number/UDI-DI BD710DF282CRH 10197344044023; BD710DF282RRH 10197344044030; BD710FJ282CRH 10197344044047; BD710FJ282RRH 10197344044054; LOTS EP241216 EP241218 EP241226 EP241231 EP250110 EP250121 EP250210

Devices named in this recall

4 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 97496).