Skip to content

Brand Name: Atlan recallZ-2616-2025 · Draeger, Inc. · initiated Sep 9, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Draeger, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

FDA root cause category: Process control

What the company is doing

On 09/09/2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. Actions to be taken: - The corrective action is for a Drger Service Representative to inspect, and, if necessary, replace the ventilator motor assembly. You will be contacted by your local Drger Service Representative to arrange a date for the ventilator motor assembly inspection and, if necessary, replacement. - Please ensure that all users of the Drger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. - Please complete and return the attached Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. - If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them. If you have any questions regarding the operation of your Atlan A350/A350XL anesthesia workstation, please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 2 again). If you have any questions regarding this Urgent Medical Device Correction notice, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 267-664-1131 or via email at [email protected]. Adverse events or quality problems experienced with the use of this product may be reported to FDA s MedWatch Adverse Event Reporting program online at www.fda.gov/medwatch/report.htm or by phone at 1-800-FDA-1088.

Recall details

Recall number
Z-2616-2025
Status
Open, Classified
Initiated by the company
Sep 9, 2025
Posted by the FDA
Oct 1, 2025
Recalling firm
Draeger, Inc.
Quantity
530
Product code
BSZ
FDA submissions
K230931
Distribution
Devices were imported from Draegerwerk Ag and Co. KGaA and distributed in the following states: AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, MA, MN, MO, NC, NE, NM, NV, NY, OH, PA, TX, UT, VA, WI. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Belgium, Brazil, Chili, China, Croatia, Denmark, Ecuador, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Jordan, Kuwait, Lithuania, Mexico, Netherlands, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam.

Product description

Brand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Product Description: The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates. The device can be used for mechanical ventilation, manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. The device is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions Component: Not applicable

Code information (models, lots, serial numbers)

Show the full code information (5k characters)

Lot Code: UDI-DI Number: 04048675556176 Serial Numbers Distributed in the US: ASTB-0181 ASTC-0005,ASTC-0006,ASTC-0007 ASTD-0244,ASTD-0259 ASTE-0180,ASTE-0254,ASTE-0255 ASTH-0083,ASTH-0464 ASUA-0479,ASUA-0482 ASUB-0012,ASUB-0013,ASUB-0015,ASUB-0017,ASUB-0018,ASUB-0292 ASUC-0065,ASUC-0084,ASUC-0085,ASUC-0086,ASUC-0087,ASUC-0088,ASUC-0090,ASUC-0091, ASUC-0092,ASUC-0096,ASUC-0097,ASUC-0099,ASUC-0100,ASUC-0101,ASUC-0102,ASUC-0103, ASUC-0104,ASUC-0105,ASUC-0107,ASUC-0108,ASUC-0109,ASUC-0110,ASUC-0111,ASUC-0112, ASUC-0113,ASUC-0116,ASUC-0117,ASUC-0118,ASUC-0119,ASUC-0120,ASUC-0121,ASUC-0122, ASUC-0123,ASUC-0125,ASUC-0126,ASUC-0127,ASUC-0128,ASUC-0129,ASUC-0130,ASUC-0131, ASUC-0132,ASUC-0133,ASUC-0134,ASUC-0135,ASUC-0136,ASUC-0137,ASUC-0138,ASUC-0139, ASUC-0140,ASUC-0141,ASUC-0142,ASUC-0143,ASUC-0143,ASUC-0144,ASUC-0145,ASUC-0162, ASUC-0163,ASUC-0164,ASUC-0165,ASUC-0166,ASUC-0167,ASUC-0168,ASUC-0169,ASUC-0170, ASUC-0171,ASUC-0172,ASUC-0173,ASUC-0203,ASUC-0204,ASUC-0205,ASUC-0206,ASUC-0207, ASUC-0208,ASUC-0209,ASUC-0210,ASUC-0218,ASUC-0219,ASUC-0220,ASUC-0221,ASUC-0222, ASUC-0223,ASUC-0224,ASUC-0225,ASUC-0235,ASUC-0236,ASUC-0237,ASUC-0340,ASUC-0341 ASUD-0244,ASUD-0261,ASUD-0262,ASUD-0263,ASUD-0267,ASUD-0281,ASUD-0282,ASUD-0283 Serial Numbers Distributed OUS ASUA-0353,ASUA-0441,ASUA-0447,ASUA-0472,ASUA-0478,ASUA-0496,,SUA-0497,ASUA-0500, ASUB-0004,ASUB-0024,ASUB-0025,ASUB-0030,ASUB-0033,ASUB-0034,ASUB-0035,ASUB-0038, ASUB-0039,ASUB-0040,ASUB-0041,ASUB-0043,ASUB-0044,ASUB-0046,ASUB-0048,ASUB-0063, ASUB-0064,ASUB-0066,ASUB-0067,ASUB-0069,ASUB-0073,ASUB-0074,ASUB-0075,ASUB-0076, ASUB-0077,ASUB-0095,ASUB-0096,ASUB-0097,ASUB-0098,ASUB-0099,ASUB-0100,ASUB-0101, ASUB-0102,ASUB-0104,ASUB-0105,ASUB-0107,ASUB-0108,ASUB-0109,ASUB-0112,ASUB-0113, ASUB-0114,ASUB-0115,ASUB-0116,ASUB-0118,ASUB-0120,ASUB-0121,ASUB-0122,ASUB-0123, ASUB-0124,ASUB-0128,ASUB-0129,ASUB-0130,ASUB-0136,ASUB-0137,ASUB-0139,ASUB-0140, ASUB-0141,ASUB-0143,ASUB-0144,ASUB-0145,ASUB-0146,ASUB-0147,ASUB-0148,ASUB-0149, ASUB-0150,ASUB-0151,ASUB-0152,ASUB-0153,ASUB-0154,ASUB-0155,ASUB-0156,ASUB-0157, ASUB-0158,ASUB-0159,ASUB-0160,ASUB-0161,ASUB-0162,ASUB-0163,ASUB-0177,ASUB-0178, ASUB-0188,ASUB-0189,ASUB-0190,ASUB-0191,ASUB-0192,ASUB-0193,ASUB-0194,ASUB-0195, ASUB-0196,ASUB-0197,ASUB-0198,ASUB-0199,ASUB-0201,ASUB-0202,ASUB-0203,ASUB-0204, ASUB-0205,ASUB-0206,ASUB-0207,ASUB-0208,ASUB-0209,ASUB-0210,ASUB-0211,ASUB-0223, ASUB-0224,ASUB-0225,ASUB-0226,ASUB-0227,ASUB-0229,ASUB-0230,ASUB-0231,ASUB-0232, ASUB-0233,ASUB-0234,ASUB-0235,ASUB-0236,ASUB-0237,ASUB-0238,ASUB-0240,ASUB-0243, ASUB-0245,ASUB-0252,ASUB-0253,ASUB-0254,ASUB-0255,ASUB-0256,ASUB-0257,ASUB-0258, ASUB-0260,ASUB-0261,ASUB-0262,ASUB-0263,ASUB-0264,ASUB-0265,ASUB-0271,ASUB-0272, ASUB-0273,ASUB-0276,ASUB-0278,ASUC-0001ASUC-0004,ASUC-0016,ASUC-0017,ASUC-0018, ASUC-0019,ASUC-0020,ASUC-0022,ASUC-0024,ASUC-0026,,ASUC-0028,,ASUC-0030,ASUC-0031, ASUC-0032,ASUC-0033,ASUC-0034,ASUC-0035,ASUC-0036,ASUC-0039,ASUC-0061,ASUC-0062, ASUC-0063,ASUC-0226,ASUC-0227,ASUC-0228,ASUC-0229,ASUC-0230,ASUC-0232,ASUC-0233, ASUC-0234,ASUC-0247,ASUC-0248,ASUC-0249,ASUC-0250,ASUC-0251,ASUC-0252,ASUC-0253, ASUC-0254,ASUC-0255,ASUC-0256,ASUC-0257,ASUC-0262,ASUC-0263,ASUC-0264,ASUC-0265, ASUC-0266,ASUC-0267,ASUC-0269,ASUC-0272,ASUC-0273,ASUC-0274,ASUC-0275,ASUC-0276, ASUC-0277,ASUC-0278,ASUC-0279,ASUC-0284,ASUC-0286,ASUC-0287,ASUC-0288,ASUC-0289, ASUC-0290,ASUC-0291,ASUC-0292,ASUC-0293,ASUC-0294,ASUC-0295,ASUC-0296,ASUC-0300, ASUC-0301,ASUC-0302,ASUC-0303,ASUC-0304,ASUC-0305,ASUC-0306ASUC-0307,ASUC-0308, ASUC-0309,ASUC-0310,ASUC-0311,ASUC-0314,ASUC-0315,ASUC-0316,ASUC-0319,ASUC-0320, ASUC-0321,ASUC-0322,ASUC-0323,ASUC-0339,ASUC-0342,ASUC-0343,ASUC-0344,ASUC-0345, ASUC-0346,ASUC-0348,ASUC-0349,ASUC-0350,ASUC-0351,ASUC-0352,ASUC-0353,ASUC-0354, ASUC-0355,ASUC-0356,ASUC-0357,ASUC-0358,ASUC-0359,ASUC-0360,ASUC-0361,ASUC-0362, ASUC-0363,ASUC-0364,ASUC-0365,ASUC-0366,ASUC-0367,ASUC-0380,ASUC-0381,ASUC-0382, ASUC-0383,ASUC-0384,ASUC-0385,ASUC-0386,ASUC-0387,ASUC-0391,ASUC-0392,ASUC-0393, ASUC-0394,ASUC-0395,ASUC-0396,ASUC-0397,ASUC-0398,ASUC-0399,ASUC-0400,ASUC-0401, ASUC-0402,ASUC-0403,ASUC-0404,ASUC-0405,ASUC-0406,ASUC-0407,ASUC-0408,ASUC-0409, ASUC-0410,ASUC-0416,ASUD-0001,ASUD-0002,ASUD-0004,ASUD-0005,ASUD-0006,ASUD-0007, ASUD-0008,ASUD-0009,ASUD-0010,ASUD-0011,ASUD-0012,ASUD-0013,ASUD-0014,ASUD-0015, ASUD-0016,ASUD-0018,ASUD-0019,ASUD-0020,ASUD-0022,ASUD-0023,ASUD-0041,ASUD-0042, ASUD-0043,ASUD-0044,ASUD-0046,ASUD-0047,ASUD-0048,ASUD-0049,ASUD-0050,ASUD-0051, ASUD-0057,ASUD-0059,ASUD-0060,ASUD-0061,ASUD-0062,ASUD-0063,ASUD-0064,ASUD-0065, ASUD-0066,ASUD-0074,ASUD-0075,ASUD-0077,ASUD-0078,ASUD-0079,ASUD-0080,ASUD-0081, ASUD-0098,ASUD-0099,ASUD-0100,ASUD-0101,ASUD-0102,ASUD-0103,ASUD-0104,ASUD-0105, ASUD-0106,ASUD-0107,ASUD-0108,ASUD-0109,ASUD-0126,ASUD-0127,ASUD-0128,ASUD-0129, ASUD-0130,ASUD-0131,ASUD-0133,ASUD-0135,ASUD-0136,ASUD-0140,ASUD-0141,ASUD-0150, ASUD-0168,ASUD-0173,ASUD-0174,ASUD-0175,ASUD-0181,ASUD-0182,ASUD-0183,ASUD-0184, ASUD-0185,ASUD-0186,ASUD-0196,ASUD-0198,ASUD-0199,ASUD-0200,ASUD-0201,ASUD-0202, ASUD-0203,ASUD-0204,ASUD-0205,ASUD-0206,ASUD-0219,ASUD-0220,ASUD-0221,ASUD-0222, ASUD-0223,ASUD-0224,ASUD-0225,ASUD-0226,ASUD-0227,ASUD-0228,ASUD-0238,ASUD-0239, ASUD-0245,ASUD-0246,ASUD-0247,ASUD-0248,ASUD-0249,ASUD-0250,ASUD-0251,ASUD-0252, ASUD-0253,ASUD-0258,ASUD-0259,ASUD-0260,ASUD-0273,ASUD-0274

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K230931 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97525).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page