Brand Name: Atlan recallZ-2617-2025 · Draeger, Inc. · initiated Sep 9, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Draeger, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The device either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
FDA root cause category: Process control
What the company is doing
On 09/09/2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. Actions to be taken: - The corrective action is for a Drger Service Representative to inspect, and, if necessary, replace the ventilator motor assembly. You will be contacted by your local Drger Service Representative to arrange a date for the ventilator motor assembly inspection and, if necessary, replacement. - Please ensure that all users of the Drger Atlan as well as other persons concerned within your organization are made aware of this Urgent Medical Device Correction notice. - Please complete and return the attached Medical Device Correction Return Response Acknowledgment and Receipt Form to confirm that you have received this information. - If you have made the products available to third parties, please forward this Urgent Medical Device Correction notice to them. If you have any questions regarding the operation of your Atlan A350/A350XL anesthesia workstation, please contact Drger Service Technical Support between the hours of 8:00 AM 8:00 PM EST at 1-800-437-2437 (press 2 at the prompt, then 2, then 2 again). If you have any questions regarding this Urgent Medical Device Correction notice, please contact Michael Kelhart between the hours of 8:00 AM 4:30 PM EST at 267-664-1131 or via email at [email protected]. Adverse events or quality problems experienced with the use of this product may be reported to FDA s MedWatch Adverse Event Reporting program online at www.fda.gov/medwatch/report.htm or by phone at 1-800-FDA-1088.
Recall details
- Recall number
- Z-2617-2025
- Status
- Open, Classified
- Initiated by the company
- Sep 9, 2025
- Posted by the FDA
- Oct 1, 2025
- Recalling firm
- Draeger, Inc.
- Quantity
- 223
- Product code
- BSZ
- FDA submissions
- K230931
- Distribution
- Devices were imported from Draegerwerk Ag and Co. KGaA and distributed in the following states: AL, AZ, CA, CT, FL, GA, ID, IL, IN, KS, KY, MA, MN, MO, NC, NE, NM, NV, NY, OH, PA, TX, UT, VA, WI. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Belgium, Brazil, Chili, China, Croatia, Denmark, Ecuador, France, Georgia, Germany, Greece, Hungary, India, Indonesia, Italy, Japan, Jordan, Kuwait, Lithuania, Mexico, Netherlands, Peru, Poland, Romania, Russian Federation, Saudi Arabia, Serbia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom, Vietnam.
Product description
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Product Description: The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates. The device can be used for mechanical ventilation, manual ventilation, pressure-supported spontaneous breathing, and spontaneous breathing. The device is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions Component: Not applicable
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Lot Code: UDI-DI Number: 04048675556183 Serial Numbers Distributed in the US: ASTA-0307,ASTA-0309,ASTA-0311,ASTA-0349,ASTA-0350,ASTA-0409 ASTB-0048 ASTC-0244,ASTC-0245 ASTD-0008,ASTD-0009,ASTD-0010,ASTD-0011,ASTD-0013,ASTD-0014,ASTD-0212,ASTD-0239 ASTH-0307,ASTH-0308,ASTH-0309,ASTH-0310,ASTH-0311,ASTH-0332,ASTH-0333,ASTH-0334, ASTH-0335,ASTH-0336,ASTH-0383,ASTH-0384 ASTJ-0014 ASUA-0469,ASUA-0483,ASUA-0484,ASUA-0486,ASUA-0488 ASUB-0093,ASUB-0169,ASUB-0170,ASUB-0171,ASUB-0172,ASUB-0173,ASUB-0174,ASUB-0175, ASUB-0176,ASUB-0183,ASUB-0184,ASUB-0185,ASUB-0186,ASUB-0187,ASUB-0215,ASUB-0216, ASUB-0217,ASUB-0218,ASUB-0219,ASUB-0221,ASUB-0222,ASUB-0250,ASUB-0251,ASUB-0266, ASUB-0267, ASUB-0268,ASUB-0269,ASUB-0270 ASUC-0067,ASUC-0068,ASUC-0069,ASUC-0072,ASUC-0114,ASUC-0115,ASUC-0146,ASUC-0148, ASUC-0149,ASUC-0150,ASUC-0151,ASUC-0153,ASUC-0155,ASUC-0156,ASUC-0157,ASUC-0158, ASUC-0174,ASUC-0175,ASUC-0176,ASUC-0178,ASUC-0179,ASUC-0180,ASUC-0181,ASUC-0182, ASUC-0183,ASUC-0184,ASUC-0185,ASUC-0186,ASUC-0187,ASUC-0188,,ASUC-0189,ASUC-0190, ASUC-0191,ASUC-0192,ASUC-0193,ASUC-0194,ASUC-0196,ASUC-0197,ASUC-0198,ASUC-0199, ASUC-0200,ASUC-0201,ASUC-0202,ASUC-0211,ASUC-0212,ASUC-0213,ASUC-0214,ASUC-0215, ASUC-0216,ASUC-0217,ASUC-0238,ASUC-0239,ASUC-0285,ASUC-0368,ASUC-0369,ASUC-0370, ASUC-0371,ASUC-0372,ASUC-0373 ASUD-0142,ASUD-0142,ASUD-0143,ASUD-0143,ASUD-0144,ASUD-0145,ASUD-0146,ASUD-0147, ASUD-0164,ASUD-0165,ASUD-0165,ASUD-0165,ASUD-0165,ASUD-0187,ASUD-0188,ASUD-0189, ASUD-0190,ASUD-0191,ASUD-0192,ASUD-0193,ASUD-0207,ASUD-0208,ASUD-0209,ASUD-0210, ASUD-0211,ASUD-0212,ASUD-0212,ASUD-0275,ASUD-0276,ASUD-0277,ASUD-0294 Serial Numbers Distributed OUS (Direct shipped from Draegerwerk AG & Co., KGaA): ASUB-0094,ASUB-0279,ASUB-0280,ASUB-0281,ASUB-0282,ASUB-0284,ASUB-0285,ASUB-0286, ASUC-0005,ASUC-0006,ASUC-0007,ASUC-0008,ASUC-0010,ASUC-0011,ASUC-0012,ASUC-0013, ASUC-0014,ASU-0042,ASUC-0043,ASUC-0044,ASUC-0045,ASUC-0046,ASUC-0047,ASUC-0048, ASUC-0049,ASUC-0050,ASUC-0051,ASUC-0052,ASUC-0053,ASUC-0054,ASUC-0057,ASUC-0060, ASUC-0073,ASUC-0074,ASUC-0075,ASUC-0076,ASUC-0077,ASUC-0078,ASUC-0079,ASUC-0080, SUC-0081,ASUC-0082,ASUC-0331,ASUC-0377,ASUC-0378,ASUC-0379,ASUD-0067,ASUD-0068, ASUD-0088,ASUD-0089,ASUD-0091,ASUD-0092,ASUD-0093,ASUD-0094,ASUD-0095,ASUD-0096, ASUD-0097,ASUD-0112,ASUD-0113,ASUD-0114,ASUD-0115,ASUD-0116,ASUD-0117,ASUD-0118, ASUD-0119,ASUD-0120,ASUD-0121,ASUD-0122,ASUD-0160,ASUD-0166
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Atlan A350 XL 8621600Drägerwerk AG & Co. KGaA · Gas-Machine, Anesthesia4 recalls1 death14 reports
Other devices under the same FDA submission
1 device cleared under K230931 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Atlan A350 8621500Drägerwerk AG & Co. KGaA · Gas-Machine, Anesthesia4 recalls63 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97525).
- Z-2616-2025 · Open, ClassifiedBrand Name: Atlan Product Name: Atlan A350 Model/Catalog Number: 8211500 Software Version: Not applicable Product Description: The Atlan is an anesthesia workstation intended for u…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page