Stent recallZ-0108-2026 · Microvention Inc. · initiated Sep 5, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Microvention Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Flow diverter may have a manufacturing issue related to the tantalum length and/or tantalum attachment pattern, which may lead the device to be incompletely open or not properly apposed to the vessel wall, which may necessitate device recapture and withdrawal of the delivery system and microcatheter together. Insufficient apposition could lead to long-term health consequences.
FDA root cause category: Device Design
What the company is doing
On 9/5/2025, recall notices were mailed to customers who were asked to do the following: 1) Stop using and quarantine affected devices. 2) Provide the notice to medical facilities or users to whom you have distributed recalled devices. 3) Return effected devices. 4) Complete and return the customer acknowledgement form via email to [email protected] Direct questions to the firm at [email protected] On 10/29/2025, additional recall notices were distributed identifying additional distributed devices.
Recall details
- Recall number
- Z-0108-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 5, 2025
- Posted by the FDA
- Oct 10, 2025
- Recalling firm
- Microvention Inc.
- Quantity
- 467
- Product code
- OUT
- FDA submissions
- P180027, P180027S002
- Distribution
- Worldwide - US Nationwide distribution in the states of SD, CA, SC, TX, FL, CO, NE, NY and the countries of DE, ES, PL, GB, IT, IN, AU, KZ, JP, AT, SI, RS, GR, TR, AR, CH, SG, KR, SK, MX, DK, IR, SA, BY, PA, VN, ZA, BR, TW, CN, BE, FR, VE, MY.
Product description
Flow Re-Direction Endoluminal Devices: FRED 27, REF: FRED5019, FRED5029, FRED4528, FRED4026, FRED5526, FRED4508, FRED5514, FRED3536, FRED3511, FRED5014, FRED3516, MV-F451827; FRED X 27, REF: XFRED4017-MVE, XFRED4528-MVE, MV-F501427X, XFRED4518-MVE, XFRED5526-MVE, XFRED4539-MVE, FREDX5522-PMA, FREDX3522-PMA, FREDX4525-PMA, FREDX5015-PMA, FREDX5532-PMA, FREDX4013-PMA, FREDX3513-PMA, FREDX4515-PMA, FREDX3517-PMA, MV-F352427X
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
REF/UDI-DI(Lot): FRED5019/00811425020388(0000235839), 00842429117163(0000544498); FRED5029/00811425020395(0000256343), 00842429117170(0000507209); FRED4528/00811425020425(0000265666); FRED4026/00811425021033(0000333125, 0000576973); FRED5526/00842429117194(0000361117); FRED4508/00842429117095(0000372088); FRED5514/00842429117187(0000429786, 0000478385); FRED3536/00842429117033(0000460268, 0000445555); FRED3511/00842429117002(0000452923); FRED5014/00842429117156(0000563591, 0000464401); FRED3516/00842429117019(0000513146, 0000531733, 0000536217); MV-F451827/04987892121906(0000565094); XFRED4017-MVE/00842429107508(0000478838); XFRED4528-MVE/00842429107560(0000518323, 0000663751); MV-F501427X/04987892151231(0000579031); XFRED4518-MVE/00842429107553(0000663074, 0000663750); XFRED5526-MVE/00842429107638(0000704203); XFRED4539-MVE/00842429107577(00007311620); FREDX5522-PMA/00842429114483(0000265641, 0000469609, 0000731155); FREDX3522-PMA/00842429114315(0000469598); FREDX4525-PMA/00842429114414(0000469603); FREDX5015-PMA/00842429114445(0000469606); FREDX5532-PMA/00842429114490(0000469610); FREDX4013-PMA/00842429114346(0000469613, 0000663758, 0000668189); FREDX3513-PMA/00842429114292(0000478834); FREDX4515-PMA/00842429114391(0000478840); FREDX3517-PMA/00842429114308(0000731304); MV-F352427X/04987892151101(0000760963)
Devices named in this recall
9 devices in the FDA device registry whose identifier appears in the recall's code information.
- Stent FREDX5522-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall9 reports
- Stent FREDX5015-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall8 reports
- Stent FREDX3513-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall5 reports
- Stent FREDX4515-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall5 reports
- Stent FREDX4525-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall5 reports
- Stent FREDX4013-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall1 death3 reports
- Stent FREDX3517-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall2 reports
- Stent FREDX3522-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall2 reports
- Stent FREDX5532-PMAMicrovention, Inc. · Intracranial aneurysm flow diverter1 recall2 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page