Pyxis Pro MedStation recallZ-0369-2026 · CareFusion 303, Inc. · initiated Sep 30, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
FDA root cause category: Under Investigation by firm
What the company is doing
On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which provides warnings, cautions and recommendations regarding fluid ingress. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material.
Recall details
- Recall number
- Z-0369-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 30, 2025
- Posted by the FDA
- Oct 27, 2025
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 32 units
- Product code
- BRY
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, United Arab Emirates, Australia.
Product description
Various models of BD Pyxis Pro 7-Drawer Auxiliary, Reference numbers: 1149-00 and 1152-00
Code information (models, lots, serial numbers)
All Serial Numbers/Ref #/UDI: 1149-00 10885403517822 and 1152-00 10885403517839
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- Pyxis Pro MedStation 1149-00Carefusion 303, Inc. · Cabinet, table and tray, anesthesia2 recalls4 reports
- Pyxis Pro MedStation 1152-00Carefusion 303, Inc. · Cabinet, table and tray, anesthesia2 recalls2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97566).
- Z-0367-2026 · Open, ClassifiedVarious models of BD Pyxis Medbank: Reference numbers: 1137-00 1145-00 1146-00 1147-00 1148-00 138902-01 138903-01 138905-01 138906-01 138907-01 138908-01 138909-01 138910-01 138…
- Z-0368-2026 · Open, ClassifiedVarious models of BD Pyxis MedFlex, Reference numbers: 1119-00 1139-00 139038-01 139039-01 139040-01 139041-01 139043-01 139044-01 139045-01 139046-01 139049-01 139051-01 139052-0…
- Z-0370-2026 · Open, ClassifiedBD Pyxis Pro MedStation Main, REF: 1155-00
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page