Skip to content

Pyxis Pro MedStation 1149-00Recalls & reported problems

Carefusion 303, Inc. · Pyxis Pro MedStation Dispensing System cabinet, table and tray, anesthesia · Class 1

All models of Pyxis Pro MedStation (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
2

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10885403517822), received between Aug 20, 2026 and Sep 26, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (2 reports).

2 FDA recall records name this device's identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 3 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Nov 20, 2025Class IIOpen, Classified

    Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

    Z-0962-2026 · Root cause: Device Design · Recall details · FDA record ↗

  2. Sep 30, 2025Class IIOpen, Classified

    Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

    Z-0369-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Application Program Problem2
  • Break1
  • Electrical /Electronic Property Problem1
  • Mechanical Jam1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 26, 2026Mechanical Jam

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50908098 ↗
  • MalfunctionSep 18, 2026Break; Electrical /Electronic Property Problem; Mechanical Problem

    It was reported that when using the BD Pyxis¿ Pro MedStation Aux, the drawer 2 and 3 had experienced a recurring failure issue. The customer stated that this malfunction occurred while dispensing the medication which caused a delay to the patient care. There were no adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50814726 ↗
  • MalfunctionSep 9, 2026Application Program Problem

    It was reported that when using the BD Pyxis¿ Pro MedStation Aux, the drawer was broken. The customer stated that the malfunction occurred when a user tried to issue medication to the patient and caused a delay to patient care. However, there were no adverse events or injuries reported based on this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50688726 ↗
  • MalfunctionAug 20, 2026Application Program Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50358824 ↗

Other Cabinet, table and tray, anesthesia devices

All Cabinet, table and tray, anesthesia devices →

Sources and updates