Skip to content

Pyxis Pro MedStationRecalls & reported problems across all models

Carefusion 303, Inc. · 6 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

3 open recalls name at least one Pyxis Pro MedStation model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
33
Deaths reported
0
Injuries reported
0
Recalls
3

We found 6 models of Pyxis Pro MedStation by Carefusion 303, Inc. in the FDA's device registry (GUDID). Together they are named in 33 adverse event reports in the FDA's MAUDE database, received between Oct 9, 2025 and Sep 26, 2026 — classified as 33 malfunctions.

The most frequently reported specific device problem is “Application Program Problem” (7 reports).

3 FDA recall records name at least one of these models by its device identifier, 3 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 4 reportsMay 2026: 5 reportsJun 2026: 3 reportsJul 2026: 4 reportsAug 2026: 6 reportsSep 2026: 8 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Pyxis Pro MedStation recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Nov 20, 2025Open, Classified

    Automated Dispensing Cabinets (ADC) received a Half Height CUBIE drawer firmware update, which caused a Cubie Insert event which prompts software to generate a duplicate address, which causes an error and the drawer to fail, which may lead to inability or delay in accessing stored items, delay in the replenishment of ADCs or patient specific medications stored outside of ADCs.

    Z-0962-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  2. Sep 30, 2025Open, Classified

    Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

    Z-0370-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Sep 30, 2025Open, Classified

    Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.

    Z-0369-2026 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

Most reported Pyxis Pro MedStation problems

Device problems

  • Application Program Problem7
  • Mechanical Problem7
  • Electrical /Electronic Property Problem5
  • No Apparent Adverse Event4
  • Activation Failure3
  • Break3
  • Failure to Align3
  • Labelling, Instructions for Use or Training Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions33

Pyxis Pro MedStation models

6 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Pyxis Pro MedStation adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Carefusion 303, Inc. products

All Carefusion 303, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page